Assessing the dose values received by patients during conventional radiography X-ray examinations and the technical condition of the equipment used for this purpose

Background: X-ray examination is associated with patient exposure to ionizing radiation. Dose values depend on the type of medical procedure used, the X-ray unit technical condition and exposure conditions selected. The aim of this study was to determine the dose value received by patients during ce...

Full description

Bibliographic Details
Main Authors: Marcin Bekas, Krzysztof Pachocki, Elżbieta Waśniewska, Dagmara Bogucka, Andrzej Magiera
Format: Article
Language:English
Published: Nofer Institute of Occupational Medicine 2014-12-01
Series:Medycyna Pracy
Subjects:
Online Access:http://medpr.imp.lodz.pl/Ocena-wielkosci-dawek-otrzymywanych-przez-pacjentow-podczas-ogolnodiagnostycznych-badan-rentgenowskich-oraz-ocena-stanu-technicznego-stosowanego-sprzetu,545,0,2.html
_version_ 1819075384832950272
author Marcin Bekas
Krzysztof Pachocki
Elżbieta Waśniewska
Dagmara Bogucka
Andrzej Magiera
author_facet Marcin Bekas
Krzysztof Pachocki
Elżbieta Waśniewska
Dagmara Bogucka
Andrzej Magiera
author_sort Marcin Bekas
collection DOAJ
description Background: X-ray examination is associated with patient exposure to ionizing radiation. Dose values depend on the type of medical procedure used, the X-ray unit technical condition and exposure conditions selected. The aim of this study was to determine the dose value received by patients during certain conventional radiography X-ray examinations and to assess the technical condition of medical equipment used for this purpose. Material and Methods: The study covered the total number of 118 conventional diagnostic X-ray units located in the Masovian Voivodeship. The methodology used to assess the conventional diagnostic X-ray unit technical condition and the measurement of the radiation dose rate received by patients are based on test procedures developed by the Department of Radiation Protection and Radiobiology of the National Institute of Public Health – National Institute of Hygiene (Warszawa, Poland) accredited for compliance with PN-EN 17025 standard by the Polish Centre for Accreditation. Results: It was found that 84.7% of X-ray units fully meet the criteria set out in the Polish legislation regarding the safe use of ionizing radiation in medicine, while 15.3% of the units do not meet some of them. The broadest dose value range was recorded for adult patients. Particularly, during lateral (LAT) lumbar spine radiography the recorded entrance surface dose (ESD) values ranged from 283.5 to 7827 μGy (mean: 2183.3 μGy). Conclusions: It is absolutely necessary to constantly monitor the technical condition of all X-ray units, because it affects population exposure to ionizing radiation. Furthermore, it is essential to raise radiographers’ awareness of the effects that ionizing radiation exposure can have on the human body. Med Pr 2014;65(6):715–721
first_indexed 2024-12-21T18:24:34Z
format Article
id doaj.art-d172fbead9ab4e37a4ab1b6be6f3e9f7
institution Directory Open Access Journal
issn 0465-5893
2353-1339
language English
last_indexed 2024-12-21T18:24:34Z
publishDate 2014-12-01
publisher Nofer Institute of Occupational Medicine
record_format Article
series Medycyna Pracy
spelling doaj.art-d172fbead9ab4e37a4ab1b6be6f3e9f72022-12-21T18:54:28ZengNofer Institute of Occupational MedicineMedycyna Pracy0465-58932353-13392014-12-0165671572110.13075/mp.5893.00044Assessing the dose values received by patients during conventional radiography X-ray examinations and the technical condition of the equipment used for this purposeMarcin BekasKrzysztof PachockiElżbieta Waśniewska Dagmara BoguckaAndrzej MagieraBackground: X-ray examination is associated with patient exposure to ionizing radiation. Dose values depend on the type of medical procedure used, the X-ray unit technical condition and exposure conditions selected. The aim of this study was to determine the dose value received by patients during certain conventional radiography X-ray examinations and to assess the technical condition of medical equipment used for this purpose. Material and Methods: The study covered the total number of 118 conventional diagnostic X-ray units located in the Masovian Voivodeship. The methodology used to assess the conventional diagnostic X-ray unit technical condition and the measurement of the radiation dose rate received by patients are based on test procedures developed by the Department of Radiation Protection and Radiobiology of the National Institute of Public Health – National Institute of Hygiene (Warszawa, Poland) accredited for compliance with PN-EN 17025 standard by the Polish Centre for Accreditation. Results: It was found that 84.7% of X-ray units fully meet the criteria set out in the Polish legislation regarding the safe use of ionizing radiation in medicine, while 15.3% of the units do not meet some of them. The broadest dose value range was recorded for adult patients. Particularly, during lateral (LAT) lumbar spine radiography the recorded entrance surface dose (ESD) values ranged from 283.5 to 7827 μGy (mean: 2183.3 μGy). Conclusions: It is absolutely necessary to constantly monitor the technical condition of all X-ray units, because it affects population exposure to ionizing radiation. Furthermore, it is essential to raise radiographers’ awareness of the effects that ionizing radiation exposure can have on the human body. Med Pr 2014;65(6):715–721http://medpr.imp.lodz.pl/Ocena-wielkosci-dawek-otrzymywanych-przez-pacjentow-podczas-ogolnodiagnostycznych-badan-rentgenowskich-oraz-ocena-stanu-technicznego-stosowanego-sprzetu,545,0,2.htmlconventional radiologyentrance surface doseESD
spellingShingle Marcin Bekas
Krzysztof Pachocki
Elżbieta Waśniewska
Dagmara Bogucka
Andrzej Magiera
Assessing the dose values received by patients during conventional radiography X-ray examinations and the technical condition of the equipment used for this purpose
Medycyna Pracy
conventional radiology
entrance surface dose
ESD
title Assessing the dose values received by patients during conventional radiography X-ray examinations and the technical condition of the equipment used for this purpose
title_full Assessing the dose values received by patients during conventional radiography X-ray examinations and the technical condition of the equipment used for this purpose
title_fullStr Assessing the dose values received by patients during conventional radiography X-ray examinations and the technical condition of the equipment used for this purpose
title_full_unstemmed Assessing the dose values received by patients during conventional radiography X-ray examinations and the technical condition of the equipment used for this purpose
title_short Assessing the dose values received by patients during conventional radiography X-ray examinations and the technical condition of the equipment used for this purpose
title_sort assessing the dose values received by patients during conventional radiography x ray examinations and the technical condition of the equipment used for this purpose
topic conventional radiology
entrance surface dose
ESD
url http://medpr.imp.lodz.pl/Ocena-wielkosci-dawek-otrzymywanych-przez-pacjentow-podczas-ogolnodiagnostycznych-badan-rentgenowskich-oraz-ocena-stanu-technicznego-stosowanego-sprzetu,545,0,2.html
work_keys_str_mv AT marcinbekas assessingthedosevaluesreceivedbypatientsduringconventionalradiographyxrayexaminationsandthetechnicalconditionoftheequipmentusedforthispurpose
AT krzysztofpachocki assessingthedosevaluesreceivedbypatientsduringconventionalradiographyxrayexaminationsandthetechnicalconditionoftheequipmentusedforthispurpose
AT elzbietawasniewska assessingthedosevaluesreceivedbypatientsduringconventionalradiographyxrayexaminationsandthetechnicalconditionoftheequipmentusedforthispurpose
AT dagmarabogucka assessingthedosevaluesreceivedbypatientsduringconventionalradiographyxrayexaminationsandthetechnicalconditionoftheequipmentusedforthispurpose
AT andrzejmagiera assessingthedosevaluesreceivedbypatientsduringconventionalradiographyxrayexaminationsandthetechnicalconditionoftheequipmentusedforthispurpose