Performance of diethylene glycol-based particle counters in the sub-3 nm size range
When studying new particle formation, the uncertainty in determining the "true" nucleation rate is considerably reduced when using condensation particle counters (CPCs) capable of measuring concentrations of aerosol particles at sizes close to or even at the critical cluster size (1–2...
Main Authors: | , , , , , , , , , , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Copernicus Publications
2013-07-01
|
Series: | Atmospheric Measurement Techniques |
Online Access: | http://www.atmos-meas-tech.net/6/1793/2013/amt-6-1793-2013.pdf |
_version_ | 1828232259845488640 |
---|---|
author | D. Wimmer K. Lehtipalo A. Franchin J. Kangasluoma F. Kreissl A. Kürten A. Kupc A. Metzger J. Mikkilä T. Petäjä F. Riccobono J. Vanhanen M. Kulmala J. Curtius |
author_facet | D. Wimmer K. Lehtipalo A. Franchin J. Kangasluoma F. Kreissl A. Kürten A. Kupc A. Metzger J. Mikkilä T. Petäjä F. Riccobono J. Vanhanen M. Kulmala J. Curtius |
author_sort | D. Wimmer |
collection | DOAJ |
description | When studying new particle formation, the uncertainty in determining the "true" nucleation rate is considerably reduced when using condensation particle counters (CPCs) capable of measuring concentrations of aerosol particles at sizes close to or even at the critical cluster size (1–2 nm). Recently, CPCs able to reliably detect particles below 2 nm in size and even close to 1 nm became available. Using these instruments, the corrections needed for calculating nucleation rates are substantially reduced compared to scaling the observed formation rate to the nucleation rate at the critical cluster size. However, this improved instrumentation requires a careful characterization of their cut-off size and the shape of the detection efficiency curve because relatively small shifts in the cut-off size can translate into larger relative errors when measuring particles close to the cut-off size. <br><br> Here we describe the development of two continuous-flow CPCs using diethylene glycol (DEG) as the working fluid. The design is based on two TSI 3776 counters. Several sets of measurements to characterize their performance at different temperature settings were carried out. Furthermore, two mixing-type particle size magnifiers (PSM) A09 from Airmodus were characterized in parallel. One PSM was operated at the highest mixing ratio (1 L min<sup>−1</sup> saturator flow), and the other was operated in a scanning mode, where the mixing ratios are changed periodically, resulting in a range of cut-off sizes. The mixing ratios are determined by varying the saturator flow, where the aerosol flow stays constant at 2.5 L min<sup>−1</sup>. Different test aerosols were generated using a nano-differential mobility analyser (nano-DMA) or a high-resolution DMA, to obtain detection efficiency curves for all four CPCs. One calibration setup included a high-resolution mass spectrometer (APi-TOF) for the determination of the chemical composition of the generated clusters. The lowest cut-off sizes were achieved with negatively charged ammonium sulfate clusters, resulting in cut-offs of 1.4 nm for the laminar flow CPCs and 1.2 and 1.1 nm for the PSMs. A comparison of one of the laminar-flow CPCs and one of the PSMs measuring ambient and laboratory air showed good agreement between the instruments. |
first_indexed | 2024-04-12T19:18:39Z |
format | Article |
id | doaj.art-d350fef44e00439687a5faf971cc00a2 |
institution | Directory Open Access Journal |
issn | 1867-1381 1867-8548 |
language | English |
last_indexed | 2024-04-12T19:18:39Z |
publishDate | 2013-07-01 |
publisher | Copernicus Publications |
record_format | Article |
series | Atmospheric Measurement Techniques |
spelling | doaj.art-d350fef44e00439687a5faf971cc00a22022-12-22T03:19:39ZengCopernicus PublicationsAtmospheric Measurement Techniques1867-13811867-85482013-07-01671793180410.5194/amt-6-1793-2013Performance of diethylene glycol-based particle counters in the sub-3 nm size rangeD. WimmerK. LehtipaloA. FranchinJ. KangasluomaF. KreisslA. KürtenA. KupcA. MetzgerJ. MikkiläT. PetäjäF. RiccobonoJ. VanhanenM. KulmalaJ. CurtiusWhen studying new particle formation, the uncertainty in determining the "true" nucleation rate is considerably reduced when using condensation particle counters (CPCs) capable of measuring concentrations of aerosol particles at sizes close to or even at the critical cluster size (1–2 nm). Recently, CPCs able to reliably detect particles below 2 nm in size and even close to 1 nm became available. Using these instruments, the corrections needed for calculating nucleation rates are substantially reduced compared to scaling the observed formation rate to the nucleation rate at the critical cluster size. However, this improved instrumentation requires a careful characterization of their cut-off size and the shape of the detection efficiency curve because relatively small shifts in the cut-off size can translate into larger relative errors when measuring particles close to the cut-off size. <br><br> Here we describe the development of two continuous-flow CPCs using diethylene glycol (DEG) as the working fluid. The design is based on two TSI 3776 counters. Several sets of measurements to characterize their performance at different temperature settings were carried out. Furthermore, two mixing-type particle size magnifiers (PSM) A09 from Airmodus were characterized in parallel. One PSM was operated at the highest mixing ratio (1 L min<sup>−1</sup> saturator flow), and the other was operated in a scanning mode, where the mixing ratios are changed periodically, resulting in a range of cut-off sizes. The mixing ratios are determined by varying the saturator flow, where the aerosol flow stays constant at 2.5 L min<sup>−1</sup>. Different test aerosols were generated using a nano-differential mobility analyser (nano-DMA) or a high-resolution DMA, to obtain detection efficiency curves for all four CPCs. One calibration setup included a high-resolution mass spectrometer (APi-TOF) for the determination of the chemical composition of the generated clusters. The lowest cut-off sizes were achieved with negatively charged ammonium sulfate clusters, resulting in cut-offs of 1.4 nm for the laminar flow CPCs and 1.2 and 1.1 nm for the PSMs. A comparison of one of the laminar-flow CPCs and one of the PSMs measuring ambient and laboratory air showed good agreement between the instruments.http://www.atmos-meas-tech.net/6/1793/2013/amt-6-1793-2013.pdf |
spellingShingle | D. Wimmer K. Lehtipalo A. Franchin J. Kangasluoma F. Kreissl A. Kürten A. Kupc A. Metzger J. Mikkilä T. Petäjä F. Riccobono J. Vanhanen M. Kulmala J. Curtius Performance of diethylene glycol-based particle counters in the sub-3 nm size range Atmospheric Measurement Techniques |
title | Performance of diethylene glycol-based particle counters in the sub-3 nm size range |
title_full | Performance of diethylene glycol-based particle counters in the sub-3 nm size range |
title_fullStr | Performance of diethylene glycol-based particle counters in the sub-3 nm size range |
title_full_unstemmed | Performance of diethylene glycol-based particle counters in the sub-3 nm size range |
title_short | Performance of diethylene glycol-based particle counters in the sub-3 nm size range |
title_sort | performance of diethylene glycol based particle counters in the sub 3 nm size range |
url | http://www.atmos-meas-tech.net/6/1793/2013/amt-6-1793-2013.pdf |
work_keys_str_mv | AT dwimmer performanceofdiethyleneglycolbasedparticlecountersinthesub3nmsizerange AT klehtipalo performanceofdiethyleneglycolbasedparticlecountersinthesub3nmsizerange AT afranchin performanceofdiethyleneglycolbasedparticlecountersinthesub3nmsizerange AT jkangasluoma performanceofdiethyleneglycolbasedparticlecountersinthesub3nmsizerange AT fkreissl performanceofdiethyleneglycolbasedparticlecountersinthesub3nmsizerange AT akurten performanceofdiethyleneglycolbasedparticlecountersinthesub3nmsizerange AT akupc performanceofdiethyleneglycolbasedparticlecountersinthesub3nmsizerange AT ametzger performanceofdiethyleneglycolbasedparticlecountersinthesub3nmsizerange AT jmikkila performanceofdiethyleneglycolbasedparticlecountersinthesub3nmsizerange AT tpetaja performanceofdiethyleneglycolbasedparticlecountersinthesub3nmsizerange AT friccobono performanceofdiethyleneglycolbasedparticlecountersinthesub3nmsizerange AT jvanhanen performanceofdiethyleneglycolbasedparticlecountersinthesub3nmsizerange AT mkulmala performanceofdiethyleneglycolbasedparticlecountersinthesub3nmsizerange AT jcurtius performanceofdiethyleneglycolbasedparticlecountersinthesub3nmsizerange |