Radiation dose measurements in a 256-slice computed tomography scanner
Purpose: The purpose of this study is to compare computed tomography (CT) radiation dose measurement methods proposed by TG111, International Electrotechnical Commission (IEC), and a direct dose profile integral (DPI) measurement method. Methods: Pencil and Farmer ion chambers are used for integrati...
Main Authors: | Victor J Weir, Jie Zhang |
---|---|
Format: | Article |
Language: | English |
Published: |
Wolters Kluwer Medknow Publications
2018-01-01
|
Series: | Journal of Medical Physics |
Subjects: | |
Online Access: | http://www.jmp.org.in/article.asp?issn=0971-6203;year=2018;volume=43;issue=2;spage=85;epage=92;aulast=Weir |