Low interatrial septal pacing: A simple method

Background: Sinus node disease is associated with widespread structural and electrophysiological changes in the atria in addition to abnormalities at the sinus node. The atrial conduction disorder in patients with atrial pacing results in higher incidence of atrial fibrillation. Studies have shown t...

Full description

Bibliographic Details
Main Authors: Asit Das, Dhiman Kahali, Shibananda Dutta
Format: Article
Language:English
Published: Elsevier 2017-11-01
Series:Indian Heart Journal
Subjects:
Online Access:http://www.sciencedirect.com/science/article/pii/S0019483216302322
Description
Summary:Background: Sinus node disease is associated with widespread structural and electrophysiological changes in the atria in addition to abnormalities at the sinus node. The atrial conduction disorder in patients with atrial pacing results in higher incidence of atrial fibrillation. Studies have shown that low interatrial septal pacing is superior to right atrial appendage pacing in preventing persistent or permanent atrial fibrillation in these patients. However, implantation of active fixation lead in low interatrial septal position is difficult and time consuming with conventional stylet, inhibiting application of this method in routine practice. Method: The technique of implanting atrial pacing lead in low interatrial septum with hand-made stylet is presented in this study with emphasis on fluoroscopic landmark and electrocardiographic P wave pattern. Results: The results indicate acute and short-term success of low interatrial septal pacing in 10 patients out of 11 patients without major complications. Pacing parameters during implantation and 3 months post procedure were within normal limits. Conclusion: The initial favorable results of this study indicate low interatrial septal pacing with conventional active fixation lead using fluoroscopic landmark and electrocardiographic characteristics is feasible and reproducible with a simple technique.
ISSN:0019-4832