The dependence of Ge detectors efficiency on the density of the samples in gamma-ray spectrometry

The effect of the density of environmental samples on the counting efficiency of Ge detectors used in gamma-spectrometry was studied. The dependence εff(ρ) was determined for two HPGe detectors (relative efficiencies 18% and 20%) using five radioactive standard reference materials (silicone resins,...

Full description

Bibliographic Details
Main Authors: Vukašinović Ivana, Todorović Dragana, Popović Dragana R.
Format: Article
Language:English
Published: VINCA Institute of Nuclear Sciences 2007-01-01
Series:Nuclear Technology and Radiation Protection
Subjects:
Online Access:http://www.doiserbia.nb.rs/img/doi/1451-3994/2007/1451-39940702058V.pdf
Description
Summary:The effect of the density of environmental samples on the counting efficiency of Ge detectors used in gamma-spectrometry was studied. The dependence εff(ρ) was determined for two HPGe detectors (relative efficiencies 18% and 20%) using five radioactive standard reference materials (silicone resins, epoxy resin, milk powder, soil) with different matrix densities (0.45-1.22 g/cm3) in Marinelli beakers (V = 500 cm3). The dependence of efficiency vs. density was found to be linear and the regression parameters for energies in the range of 60-2000 keV were determined, too. The effect of variation in density on the counting efficiency of Ge detectors is dominant in the range of lower energies (60-600 keV) and de creases with energies in the higher energy range.
ISSN:1451-3994