Super-Nernstian Potentiometric pH Sensor based on the Electrodeposition of Iridium Oxide Nanoparticles

This work reports an investigation into the synthesis and electrodeposition of iridium oxide nanoparticles to fabricate an Au-based super-Nernstian potentiometric pH sensor. Monodisperse ultrafine iridium oxide nanoparticles with a mean particle diameter of 1–2 nm were successfully synthesized by th...

Full description

Bibliographic Details
Main Authors: Munawar Khalil, Ning Liu, Robert Lee
Format: Article
Language:English
Published: Universitas Indonesia 2018-04-01
Series:International Journal of Technology
Subjects:
Online Access:http://ijtech.eng.ui.ac.id/article/view/1263
Description
Summary:This work reports an investigation into the synthesis and electrodeposition of iridium oxide nanoparticles to fabricate an Au-based super-Nernstian potentiometric pH sensor. Monodisperse ultrafine iridium oxide nanoparticles with a mean particle diameter of 1–2 nm were successfully synthesized by the facile alkaline hydrolysis method and electrodeposited on the surface of Au substrate using Cyclic Voltammetry (CV). Based on the result, it was observed that the iridium oxide deposited Au substrate had a rough surface morphology. It was also found that the as-prepared sensor exhibited an excellent pH sensitivity and good stability over a long period, with a super-Nernstian response value of -73.7 mV/pH.
ISSN:2086-9614
2087-2100