Efficacy of low-dose cone beam computed tomography and metal artifact reduction tool for assessment of peri-implant bone defects: an in vitro study

Abstract Background Early accurate radiographic assessment of peri-implant bone condition is highly important to avoid excessive loss of supporting bone and implant failure. Cone beam computed tomography (CBCT) is the radiographic technique of choice if peri-implant dehiscence and fenestration defec...

Full description

Bibliographic Details
Main Authors: Alaa Shafiek Nomier, Yousria Salah El-Din Gaweesh, Maha R. Taalab, Shaimaa Abu El Sadat
Format: Article
Language:English
Published: BMC 2022-12-01
Series:BMC Oral Health
Subjects:
Online Access:https://doi.org/10.1186/s12903-022-02663-8
_version_ 1811291413518745600
author Alaa Shafiek Nomier
Yousria Salah El-Din Gaweesh
Maha R. Taalab
Shaimaa Abu El Sadat
author_facet Alaa Shafiek Nomier
Yousria Salah El-Din Gaweesh
Maha R. Taalab
Shaimaa Abu El Sadat
author_sort Alaa Shafiek Nomier
collection DOAJ
description Abstract Background Early accurate radiographic assessment of peri-implant bone condition is highly important to avoid excessive loss of supporting bone and implant failure. Cone beam computed tomography (CBCT) is the radiographic technique of choice if peri-implant dehiscence and fenestration defects are suspected. The higher radiation dose and the presence of beam hardening artifacts are the main drawbacks of CBCT imaging techniques. This study aims to evaluate the influence of low-dose cone beam computed tomography (LD-CBCT) and metal artifact reduction (MAR) tool on the assessment of peri-implant dehiscence and fenestration. Methodology Thirty titanium implants were inserted into bovine rib blocks. Twenty had standardized bone defects (10 with dehiscence and 10 with fenestration), while the remaining 10 were used as control group with no defects. Radiographic examinations held with high‐definition CBCT (HD-CBCT) and LD-CBCT with and without application of MAR tool. Images were assessed by four examiners for the presence or absence of peri-implant defects. The area under the area under the receiver operating characteristic (ROC) curve (AUC), sensitivity, specificity, and accuracy were calculated for all radiographic protocols. Results In the absence of MAR tool, there was no difference in AUC and diagnostic values between LD-CBCT and HD-CBCT for detection of both defects. When the MAR tool was applied, the AUC values, sensitivity, and accuracy were higher in HD-CBCT than in LD-CBCT for the detection of both defects, especially for the dehiscence, while specificity remained the same. Conclusion LD-CBCT can be used in the evaluation of peri-implant dehiscence and fenestration without any decrease in diagnostic accuracy. The application of MAR tool decrease the diagnostic ability of both defects, especially for the detection of dehiscence defects.
first_indexed 2024-04-13T04:29:00Z
format Article
id doaj.art-da36c6c4a09c4daaa2ba0fd545b1f3a6
institution Directory Open Access Journal
issn 1472-6831
language English
last_indexed 2024-04-13T04:29:00Z
publishDate 2022-12-01
publisher BMC
record_format Article
series BMC Oral Health
spelling doaj.art-da36c6c4a09c4daaa2ba0fd545b1f3a62022-12-22T03:02:24ZengBMCBMC Oral Health1472-68312022-12-0122111110.1186/s12903-022-02663-8Efficacy of low-dose cone beam computed tomography and metal artifact reduction tool for assessment of peri-implant bone defects: an in vitro studyAlaa Shafiek Nomier0Yousria Salah El-Din Gaweesh1Maha R. Taalab2Shaimaa Abu El Sadat3Department of Oral Medicine, Periodontology, Oral Diagnosis and Oral Radiology, Faculty of Dentistry, Alexandria UniversityDepartment of Oral Medicine, Periodontology, Oral Diagnosis and Oral Radiology, Faculty of Dentistry, Alexandria UniversityDepartment of Oral Medicine, Periodontology, Oral Diagnosis and Oral Radiology, Faculty of Dentistry, Alexandria UniversityDepartment of Oral and Maxillofacial Radiology, Faculty of Dentistry, Ain-Shams UniversityAbstract Background Early accurate radiographic assessment of peri-implant bone condition is highly important to avoid excessive loss of supporting bone and implant failure. Cone beam computed tomography (CBCT) is the radiographic technique of choice if peri-implant dehiscence and fenestration defects are suspected. The higher radiation dose and the presence of beam hardening artifacts are the main drawbacks of CBCT imaging techniques. This study aims to evaluate the influence of low-dose cone beam computed tomography (LD-CBCT) and metal artifact reduction (MAR) tool on the assessment of peri-implant dehiscence and fenestration. Methodology Thirty titanium implants were inserted into bovine rib blocks. Twenty had standardized bone defects (10 with dehiscence and 10 with fenestration), while the remaining 10 were used as control group with no defects. Radiographic examinations held with high‐definition CBCT (HD-CBCT) and LD-CBCT with and without application of MAR tool. Images were assessed by four examiners for the presence or absence of peri-implant defects. The area under the area under the receiver operating characteristic (ROC) curve (AUC), sensitivity, specificity, and accuracy were calculated for all radiographic protocols. Results In the absence of MAR tool, there was no difference in AUC and diagnostic values between LD-CBCT and HD-CBCT for detection of both defects. When the MAR tool was applied, the AUC values, sensitivity, and accuracy were higher in HD-CBCT than in LD-CBCT for the detection of both defects, especially for the dehiscence, while specificity remained the same. Conclusion LD-CBCT can be used in the evaluation of peri-implant dehiscence and fenestration without any decrease in diagnostic accuracy. The application of MAR tool decrease the diagnostic ability of both defects, especially for the detection of dehiscence defects.https://doi.org/10.1186/s12903-022-02663-8CBCTDehiscenceDental implantsFenestrationLow-dose protocolMAR
spellingShingle Alaa Shafiek Nomier
Yousria Salah El-Din Gaweesh
Maha R. Taalab
Shaimaa Abu El Sadat
Efficacy of low-dose cone beam computed tomography and metal artifact reduction tool for assessment of peri-implant bone defects: an in vitro study
BMC Oral Health
CBCT
Dehiscence
Dental implants
Fenestration
Low-dose protocol
MAR
title Efficacy of low-dose cone beam computed tomography and metal artifact reduction tool for assessment of peri-implant bone defects: an in vitro study
title_full Efficacy of low-dose cone beam computed tomography and metal artifact reduction tool for assessment of peri-implant bone defects: an in vitro study
title_fullStr Efficacy of low-dose cone beam computed tomography and metal artifact reduction tool for assessment of peri-implant bone defects: an in vitro study
title_full_unstemmed Efficacy of low-dose cone beam computed tomography and metal artifact reduction tool for assessment of peri-implant bone defects: an in vitro study
title_short Efficacy of low-dose cone beam computed tomography and metal artifact reduction tool for assessment of peri-implant bone defects: an in vitro study
title_sort efficacy of low dose cone beam computed tomography and metal artifact reduction tool for assessment of peri implant bone defects an in vitro study
topic CBCT
Dehiscence
Dental implants
Fenestration
Low-dose protocol
MAR
url https://doi.org/10.1186/s12903-022-02663-8
work_keys_str_mv AT alaashafieknomier efficacyoflowdoseconebeamcomputedtomographyandmetalartifactreductiontoolforassessmentofperiimplantbonedefectsaninvitrostudy
AT yousriasalaheldingaweesh efficacyoflowdoseconebeamcomputedtomographyandmetalartifactreductiontoolforassessmentofperiimplantbonedefectsaninvitrostudy
AT mahartaalab efficacyoflowdoseconebeamcomputedtomographyandmetalartifactreductiontoolforassessmentofperiimplantbonedefectsaninvitrostudy
AT shaimaaabuelsadat efficacyoflowdoseconebeamcomputedtomographyandmetalartifactreductiontoolforassessmentofperiimplantbonedefectsaninvitrostudy