Evaluation of a Dedicated Radiofrequency Carotid PET/MRI Coil

Carotid radiofrequency coils inside a PET/MRI system can result in PET quantification errors. We compared the performance of a dedicated PET/MRI carotid coil against a coil for MRI-only use. An <sup>18</sup>F-fluorodeoxyglucose (<sup>18</sup>F-FDG) phantom was scanned without...

Full description

Bibliographic Details
Main Authors: Mueez Aizaz, Jochem A. J. van der Pol, Roel Wierts, Hans Zwart, Abe J. van der Werf, Joachim E. Wildberger, Jan A. Bucerius, Rik P. M. Moonen, Marianne Eline Kooi
Format: Article
Language:English
Published: MDPI AG 2022-05-01
Series:Journal of Clinical Medicine
Subjects:
Online Access:https://www.mdpi.com/2077-0383/11/9/2569
_version_ 1797504172920143872
author Mueez Aizaz
Jochem A. J. van der Pol
Roel Wierts
Hans Zwart
Abe J. van der Werf
Joachim E. Wildberger
Jan A. Bucerius
Rik P. M. Moonen
Marianne Eline Kooi
author_facet Mueez Aizaz
Jochem A. J. van der Pol
Roel Wierts
Hans Zwart
Abe J. van der Werf
Joachim E. Wildberger
Jan A. Bucerius
Rik P. M. Moonen
Marianne Eline Kooi
author_sort Mueez Aizaz
collection DOAJ
description Carotid radiofrequency coils inside a PET/MRI system can result in PET quantification errors. We compared the performance of a dedicated PET/MRI carotid coil against a coil for MRI-only use. An <sup>18</sup>F-fluorodeoxyglucose (<sup>18</sup>F-FDG) phantom was scanned without and with an MRI-only coil and with the PET/MRI coil. The decay-corrected normalized activity was compared for the different coil configurations. Eighteen patients were scanned with the three coil configurations. The maximal standardized uptake values (SUV<sub>max</sub>) and signal-to-noise ratios (SNR) were calculated. Repeated measures ANOVA was performed to assess the differences in SUV<sub>max</sub> and SNR between the coil configurations. In the phantom study, the PET/MRI coil demonstrated a slight decrease (<5%), while the MRI-only coil showed a substantial decrease (up to 10%) in normalized activity at the position of coil elements compared to no dedicated coil configuration. In the patient study, the SUV<sub>max</sub> values for both no surface coil (3.59 ± 0.15) and PET/MRI coil (3.54 ± 0.15) were significantly higher (<i>p</i> = 0.03 and <i>p</i> = 0.04, respectively) as compared to the MRI-only coil (3.28 ± 0.16). No significant difference was observed between PET/MRI and no surface coil (<i>p</i> = 1.0). The SNR values for both PET/MRI (7.31 ± 0.44) and MRI-only (7.62 ± 0.42) configurations demonstrated significantly higher (<i>p</i> < 0.001) SNR values as compared to the no surface coil (3.78 ± 0.22), while no significant difference was observed in SNR between the PET/MRI and MRI-only coil (<i>p</i> = 1.0). This study demonstrated that the PET/MRI coil can be used for PET imaging without requiring attenuation correction while acquiring high-resolution MR images.
first_indexed 2024-03-10T04:00:53Z
format Article
id doaj.art-db3cfc07398a4a5b90811c351d1608c5
institution Directory Open Access Journal
issn 2077-0383
language English
last_indexed 2024-03-10T04:00:53Z
publishDate 2022-05-01
publisher MDPI AG
record_format Article
series Journal of Clinical Medicine
spelling doaj.art-db3cfc07398a4a5b90811c351d1608c52023-11-23T08:34:12ZengMDPI AGJournal of Clinical Medicine2077-03832022-05-01119256910.3390/jcm11092569Evaluation of a Dedicated Radiofrequency Carotid PET/MRI CoilMueez Aizaz0Jochem A. J. van der Pol1Roel Wierts2Hans Zwart3Abe J. van der Werf4Joachim E. Wildberger5Jan A. Bucerius6Rik P. M. Moonen7Marianne Eline Kooi8Department of Radiology and Nuclear Medicine, Maastricht University Medical Center+, 6202 AZ Maastricht, The NetherlandsDepartment of Radiology and Nuclear Medicine, Maastricht University Medical Center+, 6202 AZ Maastricht, The NetherlandsDepartment of Radiology and Nuclear Medicine, Maastricht University Medical Center+, 6202 AZ Maastricht, The NetherlandsMachnet B.V, 9301 LK Roden, The NetherlandsMachnet B.V, 9301 LK Roden, The NetherlandsDepartment of Radiology and Nuclear Medicine, Maastricht University Medical Center+, 6202 AZ Maastricht, The NetherlandsDepartment of Radiology and Nuclear Medicine, Maastricht University Medical Center+, 6202 AZ Maastricht, The NetherlandsDepartment of Radiology and Nuclear Medicine, Maastricht University Medical Center+, 6202 AZ Maastricht, The NetherlandsDepartment of Radiology and Nuclear Medicine, Maastricht University Medical Center+, 6202 AZ Maastricht, The NetherlandsCarotid radiofrequency coils inside a PET/MRI system can result in PET quantification errors. We compared the performance of a dedicated PET/MRI carotid coil against a coil for MRI-only use. An <sup>18</sup>F-fluorodeoxyglucose (<sup>18</sup>F-FDG) phantom was scanned without and with an MRI-only coil and with the PET/MRI coil. The decay-corrected normalized activity was compared for the different coil configurations. Eighteen patients were scanned with the three coil configurations. The maximal standardized uptake values (SUV<sub>max</sub>) and signal-to-noise ratios (SNR) were calculated. Repeated measures ANOVA was performed to assess the differences in SUV<sub>max</sub> and SNR between the coil configurations. In the phantom study, the PET/MRI coil demonstrated a slight decrease (<5%), while the MRI-only coil showed a substantial decrease (up to 10%) in normalized activity at the position of coil elements compared to no dedicated coil configuration. In the patient study, the SUV<sub>max</sub> values for both no surface coil (3.59 ± 0.15) and PET/MRI coil (3.54 ± 0.15) were significantly higher (<i>p</i> = 0.03 and <i>p</i> = 0.04, respectively) as compared to the MRI-only coil (3.28 ± 0.16). No significant difference was observed between PET/MRI and no surface coil (<i>p</i> = 1.0). The SNR values for both PET/MRI (7.31 ± 0.44) and MRI-only (7.62 ± 0.42) configurations demonstrated significantly higher (<i>p</i> < 0.001) SNR values as compared to the no surface coil (3.78 ± 0.22), while no significant difference was observed in SNR between the PET/MRI and MRI-only coil (<i>p</i> = 1.0). This study demonstrated that the PET/MRI coil can be used for PET imaging without requiring attenuation correction while acquiring high-resolution MR images.https://www.mdpi.com/2077-0383/11/9/2569carotid imagingflexible MRI coilsPET/MRIattenuation correction
spellingShingle Mueez Aizaz
Jochem A. J. van der Pol
Roel Wierts
Hans Zwart
Abe J. van der Werf
Joachim E. Wildberger
Jan A. Bucerius
Rik P. M. Moonen
Marianne Eline Kooi
Evaluation of a Dedicated Radiofrequency Carotid PET/MRI Coil
Journal of Clinical Medicine
carotid imaging
flexible MRI coils
PET/MRI
attenuation correction
title Evaluation of a Dedicated Radiofrequency Carotid PET/MRI Coil
title_full Evaluation of a Dedicated Radiofrequency Carotid PET/MRI Coil
title_fullStr Evaluation of a Dedicated Radiofrequency Carotid PET/MRI Coil
title_full_unstemmed Evaluation of a Dedicated Radiofrequency Carotid PET/MRI Coil
title_short Evaluation of a Dedicated Radiofrequency Carotid PET/MRI Coil
title_sort evaluation of a dedicated radiofrequency carotid pet mri coil
topic carotid imaging
flexible MRI coils
PET/MRI
attenuation correction
url https://www.mdpi.com/2077-0383/11/9/2569
work_keys_str_mv AT mueezaizaz evaluationofadedicatedradiofrequencycarotidpetmricoil
AT jochemajvanderpol evaluationofadedicatedradiofrequencycarotidpetmricoil
AT roelwierts evaluationofadedicatedradiofrequencycarotidpetmricoil
AT hanszwart evaluationofadedicatedradiofrequencycarotidpetmricoil
AT abejvanderwerf evaluationofadedicatedradiofrequencycarotidpetmricoil
AT joachimewildberger evaluationofadedicatedradiofrequencycarotidpetmricoil
AT janabucerius evaluationofadedicatedradiofrequencycarotidpetmricoil
AT rikpmmoonen evaluationofadedicatedradiofrequencycarotidpetmricoil
AT marianneelinekooi evaluationofadedicatedradiofrequencycarotidpetmricoil