Inadvertent malposition of a permanent ventricular lead into the middle cardiac vein was misdiagnosed as lead perforation

Abstract A 54‐year‐old man had a dual‐chamber pacemaker implantation 9 years ago because of sick sinus syndrome at a different facility. The patient did not undergo any evaluation of his pacemaker for a long time with cardiologist. The patient was admitted to another hospital manifesting dyspnea and...

Full description

Bibliographic Details
Main Authors: Mingxian Chen, Zhihong Wu, Zhenjiang Liu, Lin Hu, Xuping Li, Qiming Liu, Shenghua Zhou
Format: Article
Language:English
Published: Wiley 2022-07-01
Series:Annals of Noninvasive Electrocardiology
Subjects:
Online Access:https://doi.org/10.1111/anec.12949
_version_ 1828401300033765376
author Mingxian Chen
Zhihong Wu
Zhenjiang Liu
Lin Hu
Xuping Li
Qiming Liu
Shenghua Zhou
author_facet Mingxian Chen
Zhihong Wu
Zhenjiang Liu
Lin Hu
Xuping Li
Qiming Liu
Shenghua Zhou
author_sort Mingxian Chen
collection DOAJ
description Abstract A 54‐year‐old man had a dual‐chamber pacemaker implantation 9 years ago because of sick sinus syndrome at a different facility. The patient did not undergo any evaluation of his pacemaker for a long time with cardiologist. The patient was admitted to another hospital manifesting dyspnea and palpitation with atrial fibrillation for 1 month, and he was diagnosed with ventricular lead perforation. For further treatment, he was referred to our hospital, and an elective replacement indicator (ERI) of the battery state and a malpositioned ventricular lead into the middle cardiac vein were found. Finally, the pacing lead was left in the primary place and the pacemaker was replaced.
first_indexed 2024-12-10T09:45:01Z
format Article
id doaj.art-db98b4f65cc7448e8cdd9a379f53e5f7
institution Directory Open Access Journal
issn 1082-720X
1542-474X
language English
last_indexed 2024-12-10T09:45:01Z
publishDate 2022-07-01
publisher Wiley
record_format Article
series Annals of Noninvasive Electrocardiology
spelling doaj.art-db98b4f65cc7448e8cdd9a379f53e5f72022-12-22T01:53:53ZengWileyAnnals of Noninvasive Electrocardiology1082-720X1542-474X2022-07-01274n/an/a10.1111/anec.12949Inadvertent malposition of a permanent ventricular lead into the middle cardiac vein was misdiagnosed as lead perforationMingxian Chen0Zhihong Wu1Zhenjiang Liu2Lin Hu3Xuping Li4Qiming Liu5Shenghua Zhou6Department of Cardiology The Second Xiangya Hospital of Central South University Changsha ChinaDepartment of Cardiology The Second Xiangya Hospital of Central South University Changsha ChinaDepartment of Cardiology The Second Xiangya Hospital of Central South University Changsha ChinaDepartment of Cardiology The Second Xiangya Hospital of Central South University Changsha ChinaDepartment of Cardiology The Second Xiangya Hospital of Central South University Changsha ChinaDepartment of Cardiology The Second Xiangya Hospital of Central South University Changsha ChinaDepartment of Cardiology The Second Xiangya Hospital of Central South University Changsha ChinaAbstract A 54‐year‐old man had a dual‐chamber pacemaker implantation 9 years ago because of sick sinus syndrome at a different facility. The patient did not undergo any evaluation of his pacemaker for a long time with cardiologist. The patient was admitted to another hospital manifesting dyspnea and palpitation with atrial fibrillation for 1 month, and he was diagnosed with ventricular lead perforation. For further treatment, he was referred to our hospital, and an elective replacement indicator (ERI) of the battery state and a malpositioned ventricular lead into the middle cardiac vein were found. Finally, the pacing lead was left in the primary place and the pacemaker was replaced.https://doi.org/10.1111/anec.12949malpositionmiddle cardiac veinpacemakerpacing lead
spellingShingle Mingxian Chen
Zhihong Wu
Zhenjiang Liu
Lin Hu
Xuping Li
Qiming Liu
Shenghua Zhou
Inadvertent malposition of a permanent ventricular lead into the middle cardiac vein was misdiagnosed as lead perforation
Annals of Noninvasive Electrocardiology
malposition
middle cardiac vein
pacemaker
pacing lead
title Inadvertent malposition of a permanent ventricular lead into the middle cardiac vein was misdiagnosed as lead perforation
title_full Inadvertent malposition of a permanent ventricular lead into the middle cardiac vein was misdiagnosed as lead perforation
title_fullStr Inadvertent malposition of a permanent ventricular lead into the middle cardiac vein was misdiagnosed as lead perforation
title_full_unstemmed Inadvertent malposition of a permanent ventricular lead into the middle cardiac vein was misdiagnosed as lead perforation
title_short Inadvertent malposition of a permanent ventricular lead into the middle cardiac vein was misdiagnosed as lead perforation
title_sort inadvertent malposition of a permanent ventricular lead into the middle cardiac vein was misdiagnosed as lead perforation
topic malposition
middle cardiac vein
pacemaker
pacing lead
url https://doi.org/10.1111/anec.12949
work_keys_str_mv AT mingxianchen inadvertentmalpositionofapermanentventricularleadintothemiddlecardiacveinwasmisdiagnosedasleadperforation
AT zhihongwu inadvertentmalpositionofapermanentventricularleadintothemiddlecardiacveinwasmisdiagnosedasleadperforation
AT zhenjiangliu inadvertentmalpositionofapermanentventricularleadintothemiddlecardiacveinwasmisdiagnosedasleadperforation
AT linhu inadvertentmalpositionofapermanentventricularleadintothemiddlecardiacveinwasmisdiagnosedasleadperforation
AT xupingli inadvertentmalpositionofapermanentventricularleadintothemiddlecardiacveinwasmisdiagnosedasleadperforation
AT qimingliu inadvertentmalpositionofapermanentventricularleadintothemiddlecardiacveinwasmisdiagnosedasleadperforation
AT shenghuazhou inadvertentmalpositionofapermanentventricularleadintothemiddlecardiacveinwasmisdiagnosedasleadperforation