Coupled biomechanical response of the cornea assessed by non-contact tonometry. A simulation study.

The mechanical response of the cornea subjected to a non-contact air-jet tonometry diagnostic test represents an interplay between its geometry, the corneal material behavior and the loading. The objective is to study this interplay to better understand and interpret the results obtained with a non-...

Full description

Bibliographic Details
Main Authors: Miguel Á Ariza-Gracia, Jesús F Zurita, David P Piñero, José F Rodriguez-Matas, Begoña Calvo
Format: Article
Language:English
Published: Public Library of Science (PLoS) 2015-01-01
Series:PLoS ONE
Online Access:http://europepmc.org/articles/PMC4364121?pdf=render
_version_ 1830373722262863872
author Miguel Á Ariza-Gracia
Jesús F Zurita
David P Piñero
José F Rodriguez-Matas
Begoña Calvo
author_facet Miguel Á Ariza-Gracia
Jesús F Zurita
David P Piñero
José F Rodriguez-Matas
Begoña Calvo
author_sort Miguel Á Ariza-Gracia
collection DOAJ
description The mechanical response of the cornea subjected to a non-contact air-jet tonometry diagnostic test represents an interplay between its geometry, the corneal material behavior and the loading. The objective is to study this interplay to better understand and interpret the results obtained with a non-contact tonometry test. A patient-specific finite element model of a healthy eye, accounting for the load free configuration, was used. The corneal tissue was modeled as an anisotropic hyperelastic material with two preferential directions. Three different sets of parameters within the human experimental range obtained from inflation tests were considered. The influence of the IOP was studied by considering four pressure levels (10-28 mmHg) whereas the influence of corneal thickness was studied by inducing a uniform variation (300-600 microns). A Computer Fluid Dynamics (CFD) air-jet simulation determined pressure loading exerted on the anterior corneal surface. The maximum apex displacement showed a linear variation with IOP for all materials examined. On the contrary, the maximum apex displacement followed a cubic relation with corneal thickness. In addition, a significant sensitivity of the apical displacement to the corneal stiffness was also obtained. Explanation to this behavior was found in the fact that the cornea experiences bending when subjected to an air-puff loading, causing the anterior surface to work in compression whereas the posterior surface works in tension. Hence, collagen fibers located at the anterior surface do not contribute to load bearing. Non-contact tonometry devices give useful information that could be misleading since the corneal deformation is the result of the interaction between the mechanical properties, IOP, and geometry. Therefore, a non-contact tonometry test is not sufficient to evaluate their individual contribution and a complete in-vivo characterization would require more than one test to independently determine the membrane and bending corneal behavior.
first_indexed 2024-12-20T07:23:49Z
format Article
id doaj.art-dba439f0f01e4645965d39cdaeb2d985
institution Directory Open Access Journal
issn 1932-6203
language English
last_indexed 2024-12-20T07:23:49Z
publishDate 2015-01-01
publisher Public Library of Science (PLoS)
record_format Article
series PLoS ONE
spelling doaj.art-dba439f0f01e4645965d39cdaeb2d9852022-12-21T19:48:36ZengPublic Library of Science (PLoS)PLoS ONE1932-62032015-01-01103e012148610.1371/journal.pone.0121486Coupled biomechanical response of the cornea assessed by non-contact tonometry. A simulation study.Miguel Á Ariza-GraciaJesús F ZuritaDavid P PiñeroJosé F Rodriguez-MatasBegoña CalvoThe mechanical response of the cornea subjected to a non-contact air-jet tonometry diagnostic test represents an interplay between its geometry, the corneal material behavior and the loading. The objective is to study this interplay to better understand and interpret the results obtained with a non-contact tonometry test. A patient-specific finite element model of a healthy eye, accounting for the load free configuration, was used. The corneal tissue was modeled as an anisotropic hyperelastic material with two preferential directions. Three different sets of parameters within the human experimental range obtained from inflation tests were considered. The influence of the IOP was studied by considering four pressure levels (10-28 mmHg) whereas the influence of corneal thickness was studied by inducing a uniform variation (300-600 microns). A Computer Fluid Dynamics (CFD) air-jet simulation determined pressure loading exerted on the anterior corneal surface. The maximum apex displacement showed a linear variation with IOP for all materials examined. On the contrary, the maximum apex displacement followed a cubic relation with corneal thickness. In addition, a significant sensitivity of the apical displacement to the corneal stiffness was also obtained. Explanation to this behavior was found in the fact that the cornea experiences bending when subjected to an air-puff loading, causing the anterior surface to work in compression whereas the posterior surface works in tension. Hence, collagen fibers located at the anterior surface do not contribute to load bearing. Non-contact tonometry devices give useful information that could be misleading since the corneal deformation is the result of the interaction between the mechanical properties, IOP, and geometry. Therefore, a non-contact tonometry test is not sufficient to evaluate their individual contribution and a complete in-vivo characterization would require more than one test to independently determine the membrane and bending corneal behavior.http://europepmc.org/articles/PMC4364121?pdf=render
spellingShingle Miguel Á Ariza-Gracia
Jesús F Zurita
David P Piñero
José F Rodriguez-Matas
Begoña Calvo
Coupled biomechanical response of the cornea assessed by non-contact tonometry. A simulation study.
PLoS ONE
title Coupled biomechanical response of the cornea assessed by non-contact tonometry. A simulation study.
title_full Coupled biomechanical response of the cornea assessed by non-contact tonometry. A simulation study.
title_fullStr Coupled biomechanical response of the cornea assessed by non-contact tonometry. A simulation study.
title_full_unstemmed Coupled biomechanical response of the cornea assessed by non-contact tonometry. A simulation study.
title_short Coupled biomechanical response of the cornea assessed by non-contact tonometry. A simulation study.
title_sort coupled biomechanical response of the cornea assessed by non contact tonometry a simulation study
url http://europepmc.org/articles/PMC4364121?pdf=render
work_keys_str_mv AT miguelaarizagracia coupledbiomechanicalresponseofthecorneaassessedbynoncontacttonometryasimulationstudy
AT jesusfzurita coupledbiomechanicalresponseofthecorneaassessedbynoncontacttonometryasimulationstudy
AT davidppinero coupledbiomechanicalresponseofthecorneaassessedbynoncontacttonometryasimulationstudy
AT josefrodriguezmatas coupledbiomechanicalresponseofthecorneaassessedbynoncontacttonometryasimulationstudy
AT begonacalvo coupledbiomechanicalresponseofthecorneaassessedbynoncontacttonometryasimulationstudy