Rapid Iodine Sensing on Mechanically Treated Carbon Nanofibers

In this work, we report on a rapid, efficient electrochemical iodine sensor based on mechanically treated carbon nanofiber (MCNF) electrodes. The electrode’s highly graphitic content, unique microstructure, and the presence of nitrogen heteroatoms in its atomic lattice contribute to increa...

Full description

Bibliographic Details
Main Authors: Eunbyul Cho, Alexandra Perebikovsky, Olivia Benice, Sunshine Holmberg, Marc Madou, Maziar Ghazinejad
Format: Article
Language:English
Published: MDPI AG 2018-05-01
Series:Sensors
Subjects:
Online Access:http://www.mdpi.com/1424-8220/18/5/1486
Description
Summary:In this work, we report on a rapid, efficient electrochemical iodine sensor based on mechanically treated carbon nanofiber (MCNF) electrodes. The electrode’s highly graphitic content, unique microstructure, and the presence of nitrogen heteroatoms in its atomic lattice contribute to increased heterogeneous electron transfer and improved kinetics compared to conventional pyrolytic carbons. The electrode demonstrates selectivity for iodide ions in the presence of both interfering agents and high salt concentrations. The sensor exhibits clinically relevant limits of detection of 0.59 µM and 1.41 µM, in 1X PBS and synthetic urine, respectively, and a wide dynamic range between 5 µM and 700 µM. These results illustrate the advantages of the material’s unique electrochemical properties for iodide sensing, in addition to its simple, inexpensive fabrication. The reported iodine sensor eliminates the need for specimen processing, revealing its aptitude for applications in point-of-care diagnostics.
ISSN:1424-8220