Pressure sensor catheter based on Fiber Tip Fabry-Pérot-Interferometer and Fiber Bragg Grating for temperature compensation

This paper introduces a pressure sensor catheter with two different measurement principles. A Fabry-Perot- Interferometer at the fiber tip is formed by a reflective multilayer membrane of a pressure sensor chip. The cavity provides absolute pressure sensing. 20 millimeter before the fiber tip a Fibe...

Full description

Bibliographic Details
Main Authors: Friedemann Marvin, Voigt Sebastian, Mehner Jan
Format: Article
Language:English
Published: De Gruyter 2022-09-01
Series:Current Directions in Biomedical Engineering
Subjects:
Online Access:https://doi.org/10.1515/cdbme-2022-1103
Description
Summary:This paper introduces a pressure sensor catheter with two different measurement principles. A Fabry-Perot- Interferometer at the fiber tip is formed by a reflective multilayer membrane of a pressure sensor chip. The cavity provides absolute pressure sensing. 20 millimeter before the fiber tip a Fiber Bragg Grating is integrated for additional temperature sensing. The sensing components are completely coated with a silicone mantle of 2 mm diameter. The signal evaluation is both done with a Fiber Bragg Grating interrogation device. First, mechanical structure aspects of the manufactured fiber-optic hybrid sensor as well as the sensing principle is described. Then, experimental results, including quantification and separation of pressure and temperature sensitivity are presented. Out of these characteristics the possibility for absolute pressure sensing with temperature compensation for the application as a pressure sensor catheter is confirmed.
ISSN:2364-5504