Effect of Microimplant Neck Design with and without Microthread on Pullout Strength and Destruction Volume

The microthread neck concept has been applied to dental implants. This study investigated the pullout strength and destruction volume of orthodontic microimplants with and without the microthread neck design. Fifteen microimplants (diameter: 1.5 × 10 mm) of three types (Types A and B: without microi...

Full description

Bibliographic Details
Main Authors: Yu-Chuan Tseng, Han-Sheng Chen, Szu-Yu Hsiao, Kun-Jung Hsu, Chun-Ming Chen
Format: Article
Language:English
Published: MDPI AG 2021-10-01
Series:Materials
Subjects:
Online Access:https://www.mdpi.com/1996-1944/14/20/5991
Description
Summary:The microthread neck concept has been applied to dental implants. This study investigated the pullout strength and destruction volume of orthodontic microimplants with and without the microthread neck design. Fifteen microimplants (diameter: 1.5 × 10 mm) of three types (Types A and B: without microimplant neck; Type C: with microimplant neck) were tested. The insertion torque (IT), Periotest value (PTV), horizontal pullout strength (HPS), and horizontal destruction volume (HDV) of each type were measured. Kruskal–Wallis H test and Dunn’s post-hoc comparison test were performed to compare the measured values of the three types of microimplants. The correlations of the measured values were used to perform the Spearman’s correlation coefficient analysis. The ITs of Types B (8.8 Ncm) and C (8.9 Ncm) were significantly higher than those of Type A (5.2 Ncm). Type B yielded the lowest PTV (4.1), and no statistical differences in PTV were observed among the three types. Type A had a significantly lower HPS (158.8 Ncm) than Types B (226.9 Ncm) and C (212.8 Ncm). The three types did not exhibit any significant differences in the HDV. The results of the Spearman’s correlation coefficient test revealed that HDV (ρ = 0.710) and IT (ρ = 0.813) were strongly correlated with HPS, whereas for PTV and HPS, it was not. HPS was strongly and significantly correlated with HDV. The orthodontic microimplant with a microimplant neck design did not perform better than that without a microthread in the mechanical strength test.
ISSN:1996-1944