Diagnostic Performance of an Automated System for Assaying Anti-Hepatitis E Virus Immunoglobulins M and G Compared with a Conventional Microplate Assay

To evaluate the diagnostic performance of the Liaison<sup>®</sup> Murex anti-HEV IgM and IgG assays running on the Liaison<sup>®</sup> instrument and compare the results with those obtained with Wantai HEV assays. We tested samples collected in immunocompetent and immunocompr...

Full description

Bibliographic Details
Main Authors: Florence Abravanel, Delphine Parraud, Sabine Chapuy-Regaud, Marcel Miedouge, Estelle Bonnin, Margaux Larrieu, Alexandre Aversenq, Sébastien Lhomme, Jacques Izopet
Format: Article
Language:English
Published: MDPI AG 2022-05-01
Series:Viruses
Subjects:
Online Access:https://www.mdpi.com/1999-4915/14/5/1065
Description
Summary:To evaluate the diagnostic performance of the Liaison<sup>®</sup> Murex anti-HEV IgM and IgG assays running on the Liaison<sup>®</sup> instrument and compare the results with those obtained with Wantai HEV assays. We tested samples collected in immunocompetent and immunocompromised patients during the acute (HEV RNA positive, anti-HEV IgM positive) and the post-viremic phase (HEV RNA negative, anti-HEV IgM positive) of infections. The specificity was assessed by testing HEV RNA negative/anti-HEV IgG-IgM negative samples. The clinical sensitivity of the Liaison<sup>®</sup> IgM assay was 100% for acute-phase samples (56/56) and 57.4% (27/47) for post-viremic samples from immunocompetent patients. It was 93.8% (30/32) for acute-phase (viremic) samples and 71%% (22/31) for post-viremic samples from immunocompromised patients. The clinical sensitivity of the Liaison<sup>®</sup> IgG assay was 100% for viremic samples (56/56) and 94.6% (43/47) for post-viremic samples from immunocompetent patients. It was 84.3% (27/32) for viremic samples and 93.5% (29/31) for post-viremic samples from immunocompromised patients. Specificity was very high (>99%) in both populations. We checked the limit of detection stated for the Liaison<sup>®</sup> IgG assay (0.3 U/mL). The clinical performance of the Liaison<sup>®</sup> ANTI-HEV assays was good. These rapid, automated assays for detecting anti-HEV antibodies will greatly enhance the arsenal for diagnosing HEV infections.
ISSN:1999-4915