Multi frequency phase fluorimetry (MFPF) for oxygen partial pressure measurement: ex vivo validation by polarographic clark-type electrode.

BACKGROUND: Measurement of partial pressure of oxygen (PO2) at high temporal resolution remains a technological challenge. This study introduces a novel PO2 sensing technology based on Multi-Frequency Phase Fluorimetry (MFPF). The aim was to validate MFPF against polarographic Clark-type electrode (...

Full description

Bibliographic Details
Main Authors: Stefan Boehme, Bastian Duenges, Klaus U Klein, Volker Hartwich, Beate Mayr, Jolanda Consiglio, James E Baumgardner, Klaus Markstaller, Reto Basciani, Andreas Vogt
Format: Article
Language:English
Published: Public Library of Science (PLoS) 2013-01-01
Series:PLoS ONE
Online Access:http://europepmc.org/articles/PMC3614895?pdf=render
_version_ 1818302007830642688
author Stefan Boehme
Bastian Duenges
Klaus U Klein
Volker Hartwich
Beate Mayr
Jolanda Consiglio
James E Baumgardner
Klaus Markstaller
Reto Basciani
Andreas Vogt
author_facet Stefan Boehme
Bastian Duenges
Klaus U Klein
Volker Hartwich
Beate Mayr
Jolanda Consiglio
James E Baumgardner
Klaus Markstaller
Reto Basciani
Andreas Vogt
author_sort Stefan Boehme
collection DOAJ
description BACKGROUND: Measurement of partial pressure of oxygen (PO2) at high temporal resolution remains a technological challenge. This study introduces a novel PO2 sensing technology based on Multi-Frequency Phase Fluorimetry (MFPF). The aim was to validate MFPF against polarographic Clark-type electrode (CTE) PO2 measurements. METHODOLOGY/PRINCIPAL FINDINGS: MFPF technology was first investigated in N = 8 anaesthetised pigs at FIO2 of 0.21, 0.4, 0.6, 0.8 and 1.0. At each FIO2 level, blood samples were withdrawn and PO2 was measured in vitro with MFPF using two FOXY-AL300 probes immediately followed by CTE measurement. Secondly, MFPF-PO2 readings were compared to CTE in an artificial circulatory setup (human packed red blood cells, haematocrit of 30%). The impacts of temperature (20, 30, 40°C) and blood flow (0.8, 1.6, 2.4, 3.2, 4.0 L min(-1)) on MFPF-PO2 measurements were assessed. MFPF response time in the gas- and blood-phase was determined. Porcine MFPF-PO2 ranged from 63 to 749 mmHg; the corresponding CTE samples from 43 to 712 mmHg. Linear regression: CTE = 15.59+1.18*MFPF (R(2) = 0.93; P<0.0001). Bland Altman analysis: meandiff 69.2 mmHg, rangediff -50.1/215.6 mmHg, 1.96-SD limits -56.3/194.8 mmHg. In artificial circulatory setup, MFPF-PO2 ranged from 20 to 567 mmHg and CTE samples from 11 to 575 mmHg. Linear regression: CTE = -8.73+1.05*MFPF (R(2) = 0.99; P<0.0001). Bland-Altman analysis: meandiff 6.6 mmHg, rangediff -9.7/20.5 mmHg, 1.96-SD limits -12.7/25.8 mmHg. Differences between MFPF and CTE-PO2 due to variations of temperature were less than 6 mmHg (range 0-140 mmHg) and less than 35 mmHg (range 140-750 mmHg); differences due to variations in blood flow were less than 15 mmHg (all P-values>0.05). MFPF response-time (monoexponential) was 1.48±0.26 s for the gas-phase and 1.51±0.20 s for the blood-phase. CONCLUSIONS/SIGNIFICANCE: MFPF-derived PO2 readings were reproducible and showed excellent correlation and good agreement with Clark-type electrode-based PO2 measurements. There was no relevant impact of temperature and blood flow upon MFPF-PO2 measurements. The response time of the MFPF FOXY-AL300 probe was adequate for real-time sensing in the blood phase.
first_indexed 2024-12-13T05:32:04Z
format Article
id doaj.art-e2342126ffa6444e950530538140ff8f
institution Directory Open Access Journal
issn 1932-6203
language English
last_indexed 2024-12-13T05:32:04Z
publishDate 2013-01-01
publisher Public Library of Science (PLoS)
record_format Article
series PLoS ONE
spelling doaj.art-e2342126ffa6444e950530538140ff8f2022-12-21T23:58:02ZengPublic Library of Science (PLoS)PLoS ONE1932-62032013-01-0184e6059110.1371/journal.pone.0060591Multi frequency phase fluorimetry (MFPF) for oxygen partial pressure measurement: ex vivo validation by polarographic clark-type electrode.Stefan BoehmeBastian DuengesKlaus U KleinVolker HartwichBeate MayrJolanda ConsiglioJames E BaumgardnerKlaus MarkstallerReto BascianiAndreas VogtBACKGROUND: Measurement of partial pressure of oxygen (PO2) at high temporal resolution remains a technological challenge. This study introduces a novel PO2 sensing technology based on Multi-Frequency Phase Fluorimetry (MFPF). The aim was to validate MFPF against polarographic Clark-type electrode (CTE) PO2 measurements. METHODOLOGY/PRINCIPAL FINDINGS: MFPF technology was first investigated in N = 8 anaesthetised pigs at FIO2 of 0.21, 0.4, 0.6, 0.8 and 1.0. At each FIO2 level, blood samples were withdrawn and PO2 was measured in vitro with MFPF using two FOXY-AL300 probes immediately followed by CTE measurement. Secondly, MFPF-PO2 readings were compared to CTE in an artificial circulatory setup (human packed red blood cells, haematocrit of 30%). The impacts of temperature (20, 30, 40°C) and blood flow (0.8, 1.6, 2.4, 3.2, 4.0 L min(-1)) on MFPF-PO2 measurements were assessed. MFPF response time in the gas- and blood-phase was determined. Porcine MFPF-PO2 ranged from 63 to 749 mmHg; the corresponding CTE samples from 43 to 712 mmHg. Linear regression: CTE = 15.59+1.18*MFPF (R(2) = 0.93; P<0.0001). Bland Altman analysis: meandiff 69.2 mmHg, rangediff -50.1/215.6 mmHg, 1.96-SD limits -56.3/194.8 mmHg. In artificial circulatory setup, MFPF-PO2 ranged from 20 to 567 mmHg and CTE samples from 11 to 575 mmHg. Linear regression: CTE = -8.73+1.05*MFPF (R(2) = 0.99; P<0.0001). Bland-Altman analysis: meandiff 6.6 mmHg, rangediff -9.7/20.5 mmHg, 1.96-SD limits -12.7/25.8 mmHg. Differences between MFPF and CTE-PO2 due to variations of temperature were less than 6 mmHg (range 0-140 mmHg) and less than 35 mmHg (range 140-750 mmHg); differences due to variations in blood flow were less than 15 mmHg (all P-values>0.05). MFPF response-time (monoexponential) was 1.48±0.26 s for the gas-phase and 1.51±0.20 s for the blood-phase. CONCLUSIONS/SIGNIFICANCE: MFPF-derived PO2 readings were reproducible and showed excellent correlation and good agreement with Clark-type electrode-based PO2 measurements. There was no relevant impact of temperature and blood flow upon MFPF-PO2 measurements. The response time of the MFPF FOXY-AL300 probe was adequate for real-time sensing in the blood phase.http://europepmc.org/articles/PMC3614895?pdf=render
spellingShingle Stefan Boehme
Bastian Duenges
Klaus U Klein
Volker Hartwich
Beate Mayr
Jolanda Consiglio
James E Baumgardner
Klaus Markstaller
Reto Basciani
Andreas Vogt
Multi frequency phase fluorimetry (MFPF) for oxygen partial pressure measurement: ex vivo validation by polarographic clark-type electrode.
PLoS ONE
title Multi frequency phase fluorimetry (MFPF) for oxygen partial pressure measurement: ex vivo validation by polarographic clark-type electrode.
title_full Multi frequency phase fluorimetry (MFPF) for oxygen partial pressure measurement: ex vivo validation by polarographic clark-type electrode.
title_fullStr Multi frequency phase fluorimetry (MFPF) for oxygen partial pressure measurement: ex vivo validation by polarographic clark-type electrode.
title_full_unstemmed Multi frequency phase fluorimetry (MFPF) for oxygen partial pressure measurement: ex vivo validation by polarographic clark-type electrode.
title_short Multi frequency phase fluorimetry (MFPF) for oxygen partial pressure measurement: ex vivo validation by polarographic clark-type electrode.
title_sort multi frequency phase fluorimetry mfpf for oxygen partial pressure measurement ex vivo validation by polarographic clark type electrode
url http://europepmc.org/articles/PMC3614895?pdf=render
work_keys_str_mv AT stefanboehme multifrequencyphasefluorimetrymfpfforoxygenpartialpressuremeasurementexvivovalidationbypolarographicclarktypeelectrode
AT bastianduenges multifrequencyphasefluorimetrymfpfforoxygenpartialpressuremeasurementexvivovalidationbypolarographicclarktypeelectrode
AT klausuklein multifrequencyphasefluorimetrymfpfforoxygenpartialpressuremeasurementexvivovalidationbypolarographicclarktypeelectrode
AT volkerhartwich multifrequencyphasefluorimetrymfpfforoxygenpartialpressuremeasurementexvivovalidationbypolarographicclarktypeelectrode
AT beatemayr multifrequencyphasefluorimetrymfpfforoxygenpartialpressuremeasurementexvivovalidationbypolarographicclarktypeelectrode
AT jolandaconsiglio multifrequencyphasefluorimetrymfpfforoxygenpartialpressuremeasurementexvivovalidationbypolarographicclarktypeelectrode
AT jamesebaumgardner multifrequencyphasefluorimetrymfpfforoxygenpartialpressuremeasurementexvivovalidationbypolarographicclarktypeelectrode
AT klausmarkstaller multifrequencyphasefluorimetrymfpfforoxygenpartialpressuremeasurementexvivovalidationbypolarographicclarktypeelectrode
AT retobasciani multifrequencyphasefluorimetrymfpfforoxygenpartialpressuremeasurementexvivovalidationbypolarographicclarktypeelectrode
AT andreasvogt multifrequencyphasefluorimetrymfpfforoxygenpartialpressuremeasurementexvivovalidationbypolarographicclarktypeelectrode