Different Configurations of Radio-Frequency Atomic Magnetometers—A Comparative Study
We comprehensively explore different optical configurations of a radio-frequency atomic magnetometer in the context of sensor miniaturisation. Similarities and differences in operation principles of the magnetometer arrangements are discussed. Through analysis of the radio-frequency and noise spectr...
Main Authors: | , |
---|---|
Format: | Article |
Language: | English |
Published: |
MDPI AG
2022-12-01
|
Series: | Sensors |
Subjects: | |
Online Access: | https://www.mdpi.com/1424-8220/22/24/9741 |
Summary: | We comprehensively explore different optical configurations of a radio-frequency atomic magnetometer in the context of sensor miniaturisation. Similarities and differences in operation principles of the magnetometer arrangements are discussed. Through analysis of the radio-frequency and noise spectra, we demonstrate that all configurations provide the same level of atomic polarisation and signal-to-noise ratio, but the optimum performance is achieved for significantly different laser powers and frequencies. We conclude with possible strategies for system miniaturisation. |
---|---|
ISSN: | 1424-8220 |