Comparison of NEMA characterizations for Discovery MI and Discovery MI-DR TOF PET/CT systems at different sites and with other commercial PET/CT systems

Abstract Background This article compares the physical performance of the 4-ring digital Discovery MI (DMI) and PMT-based Discovery MI-DR (DMI-DR) PET/CT systems. Physical performance was assessed according to the NEMA NU 2-2012 standards. Performance measures included spatial resolution, image qual...

Full description

Bibliographic Details
Main Authors: Alexandre Chicheportiche, Rami Marciano, Marina Orevi
Format: Article
Language:English
Published: SpringerOpen 2020-01-01
Series:EJNMMI Physics
Subjects:
Online Access:https://doi.org/10.1186/s40658-020-0271-x
_version_ 1811271009787969536
author Alexandre Chicheportiche
Rami Marciano
Marina Orevi
author_facet Alexandre Chicheportiche
Rami Marciano
Marina Orevi
author_sort Alexandre Chicheportiche
collection DOAJ
description Abstract Background This article compares the physical performance of the 4-ring digital Discovery MI (DMI) and PMT-based Discovery MI-DR (DMI-DR) PET/CT systems. Physical performance was assessed according to the NEMA NU 2-2012 standards. Performance measures included spatial resolution, image quality, scatter fraction and count rate performance, and sensitivity. Energy and timing resolutions were also measured. Published DMI and DMI-DR performance studies from other centers are reviewed and compared. Results 4-ring DMI spatial resolution at 1-cm radial offset in the radial, tangential and axial directions was 4.62, 4.18 and 4.57 mm, respectively, compared with the DMI-DR system values of 4.58, 4.52, and 5.31 mm. Measured sensitivity was 13.3 kcps/MBq at the center of the FOV and 13.4 kcps/MBq 10 cm off-center for the SiPM-based DMI system. DMI-DR system sensitivity was 6.3 kcps/MBq at the center of the FOV and 6.8 kcps/MBq at 10 cm off-center. DMI measured noise equivalent count rate peak was 175.6 kcps at 20.1 kBq/ml; DMI-DR was 146.7 kcps at 31.7 kBq/ml. Scatter fraction was 40.5% and 36.6%, respectively. DMI image contrast recovery (CR) values ranged from 73.2% (10 mm sphere) to 91.0% (37 mm sphere); DMI-DR, values ranged from 68.4% to 91.4%. DMI background variability (BV) was 1.8%–6.5%; DMI-DR was 2.3%–9.1%. The Q.Clear algorithm improved image quality, increasing CR and decreasing BV in both systems. The photopeak energy resolution was 9.63% and 12.19% for DMI and DMI-DR, respectively. The time-of-flight (TOF) resolution was 377.26 ps and 552.71 ps, respectively. Compared with measurements in other centers, results were similar and showed an absolute mean relative deviation of 6% for DMI and 7% for DMI-DR overall performance results. Conclusions Performance measures were higher for the 4-ring DMI than the DMI-DR system. The biggest advantages of the 4-ring DMI vs DMI-DR are improved sensitivity and count rate performance. This should allow a better image signal-to-noise ratio (SNR) for the same acquisition times or, similar SNR with lower acquisition times or injected activity. In its 3-ring configuration, the DMI showed worse performance results than the PMT-based system in terms of count rate scatter fraction and image quality (for similar axial FOV).
first_indexed 2024-04-12T22:12:28Z
format Article
id doaj.art-e49b6b66770043b59a122a1dcae6e8ac
institution Directory Open Access Journal
issn 2197-7364
language English
last_indexed 2024-04-12T22:12:28Z
publishDate 2020-01-01
publisher SpringerOpen
record_format Article
series EJNMMI Physics
spelling doaj.art-e49b6b66770043b59a122a1dcae6e8ac2022-12-22T03:14:41ZengSpringerOpenEJNMMI Physics2197-73642020-01-017112010.1186/s40658-020-0271-xComparison of NEMA characterizations for Discovery MI and Discovery MI-DR TOF PET/CT systems at different sites and with other commercial PET/CT systemsAlexandre Chicheportiche0Rami MarcianoMarina Orevi1Department of Nuclear Medicine and Biophysics, Hadassah Hebrew University Medical CenterDepartment of Nuclear Medicine and Biophysics, Hadassah Hebrew University Medical CenterAbstract Background This article compares the physical performance of the 4-ring digital Discovery MI (DMI) and PMT-based Discovery MI-DR (DMI-DR) PET/CT systems. Physical performance was assessed according to the NEMA NU 2-2012 standards. Performance measures included spatial resolution, image quality, scatter fraction and count rate performance, and sensitivity. Energy and timing resolutions were also measured. Published DMI and DMI-DR performance studies from other centers are reviewed and compared. Results 4-ring DMI spatial resolution at 1-cm radial offset in the radial, tangential and axial directions was 4.62, 4.18 and 4.57 mm, respectively, compared with the DMI-DR system values of 4.58, 4.52, and 5.31 mm. Measured sensitivity was 13.3 kcps/MBq at the center of the FOV and 13.4 kcps/MBq 10 cm off-center for the SiPM-based DMI system. DMI-DR system sensitivity was 6.3 kcps/MBq at the center of the FOV and 6.8 kcps/MBq at 10 cm off-center. DMI measured noise equivalent count rate peak was 175.6 kcps at 20.1 kBq/ml; DMI-DR was 146.7 kcps at 31.7 kBq/ml. Scatter fraction was 40.5% and 36.6%, respectively. DMI image contrast recovery (CR) values ranged from 73.2% (10 mm sphere) to 91.0% (37 mm sphere); DMI-DR, values ranged from 68.4% to 91.4%. DMI background variability (BV) was 1.8%–6.5%; DMI-DR was 2.3%–9.1%. The Q.Clear algorithm improved image quality, increasing CR and decreasing BV in both systems. The photopeak energy resolution was 9.63% and 12.19% for DMI and DMI-DR, respectively. The time-of-flight (TOF) resolution was 377.26 ps and 552.71 ps, respectively. Compared with measurements in other centers, results were similar and showed an absolute mean relative deviation of 6% for DMI and 7% for DMI-DR overall performance results. Conclusions Performance measures were higher for the 4-ring DMI than the DMI-DR system. The biggest advantages of the 4-ring DMI vs DMI-DR are improved sensitivity and count rate performance. This should allow a better image signal-to-noise ratio (SNR) for the same acquisition times or, similar SNR with lower acquisition times or injected activity. In its 3-ring configuration, the DMI showed worse performance results than the PMT-based system in terms of count rate scatter fraction and image quality (for similar axial FOV).https://doi.org/10.1186/s40658-020-0271-xNEMAPET/CTPhysical performanceDiscovery MIDiscovery MI-DR
spellingShingle Alexandre Chicheportiche
Rami Marciano
Marina Orevi
Comparison of NEMA characterizations for Discovery MI and Discovery MI-DR TOF PET/CT systems at different sites and with other commercial PET/CT systems
EJNMMI Physics
NEMA
PET/CT
Physical performance
Discovery MI
Discovery MI-DR
title Comparison of NEMA characterizations for Discovery MI and Discovery MI-DR TOF PET/CT systems at different sites and with other commercial PET/CT systems
title_full Comparison of NEMA characterizations for Discovery MI and Discovery MI-DR TOF PET/CT systems at different sites and with other commercial PET/CT systems
title_fullStr Comparison of NEMA characterizations for Discovery MI and Discovery MI-DR TOF PET/CT systems at different sites and with other commercial PET/CT systems
title_full_unstemmed Comparison of NEMA characterizations for Discovery MI and Discovery MI-DR TOF PET/CT systems at different sites and with other commercial PET/CT systems
title_short Comparison of NEMA characterizations for Discovery MI and Discovery MI-DR TOF PET/CT systems at different sites and with other commercial PET/CT systems
title_sort comparison of nema characterizations for discovery mi and discovery mi dr tof pet ct systems at different sites and with other commercial pet ct systems
topic NEMA
PET/CT
Physical performance
Discovery MI
Discovery MI-DR
url https://doi.org/10.1186/s40658-020-0271-x
work_keys_str_mv AT alexandrechicheportiche comparisonofnemacharacterizationsfordiscoverymianddiscoverymidrtofpetctsystemsatdifferentsitesandwithothercommercialpetctsystems
AT ramimarciano comparisonofnemacharacterizationsfordiscoverymianddiscoverymidrtofpetctsystemsatdifferentsitesandwithothercommercialpetctsystems
AT marinaorevi comparisonofnemacharacterizationsfordiscoverymianddiscoverymidrtofpetctsystemsatdifferentsitesandwithothercommercialpetctsystems