Increase of the particle hit rate in a laser single-particle mass spectrometer by pulse delayed extraction technology

<p>A single-particle mass spectrometer (SPMS) can provide a wealth of valuable information on chemical and physical parameters of individual particles in real time. One of the main performance criteria of the instrument is efficiency of particle detection (hit rate). Most SPMS instruments use...

Full description

Bibliographic Details
Main Authors: Y. Chen, V. Kozlovskiy, X. Du, J. Lv, S. Nikiforov, J. Yu, A. Kolosov, W. Gao, Z. Zhou, Z. Huang, L. Li
Format: Article
Language:English
Published: Copernicus Publications 2020-02-01
Series:Atmospheric Measurement Techniques
Online Access:https://www.atmos-meas-tech.net/13/941/2020/amt-13-941-2020.pdf
_version_ 1831734899548094464
author Y. Chen
Y. Chen
V. Kozlovskiy
X. Du
X. Du
J. Lv
S. Nikiforov
J. Yu
J. Yu
A. Kolosov
W. Gao
W. Gao
Z. Zhou
Z. Zhou
Z. Huang
Z. Huang
L. Li
L. Li
author_facet Y. Chen
Y. Chen
V. Kozlovskiy
X. Du
X. Du
J. Lv
S. Nikiforov
J. Yu
J. Yu
A. Kolosov
W. Gao
W. Gao
Z. Zhou
Z. Zhou
Z. Huang
Z. Huang
L. Li
L. Li
author_sort Y. Chen
collection DOAJ
description <p>A single-particle mass spectrometer (SPMS) can provide a wealth of valuable information on chemical and physical parameters of individual particles in real time. One of the main performance criteria of the instrument is efficiency of particle detection (hit rate). Most SPMS instruments use constant electrical field (DC) extraction, where stationary high voltage is applied to the extraction electrodes. As the aerosol particles initially carry a certain charge, those with a high amount to charge can be deflected by this electric field and lost, thus decreasing the hit rate. We realized that the delayed extraction technique can eliminate the stochastic dispersion of the particle beam caused by their deflection in the stationary electric field. As the result, the hit rate of the instrument can be significantly improved. Also, as the effect of the deflection in the electric field is mass dependent, it can cause distortion of the measured size distribution of the particles. Hence, the delayed extraction technique can bring the recorded distribution closer to the actual one. We found that the delayed extraction technique provides a mass resolution improvement as well as increases the hit rate. The gain in the hit rate depends on the type of particles. It can be 2 orders of magnitude for model particles and up to 2–4 times for ambient particles. In the present work we report experiments and results showing the effect of the delayed extraction on the beam divergence caused by particle charge, the hit rate improvement, and the effect of the delayed extraction on the measured particle size distribution.</p>
first_indexed 2024-12-21T12:16:41Z
format Article
id doaj.art-e91dbcc197d84cf386203b462e4d8fe4
institution Directory Open Access Journal
issn 1867-1381
1867-8548
language English
last_indexed 2024-12-21T12:16:41Z
publishDate 2020-02-01
publisher Copernicus Publications
record_format Article
series Atmospheric Measurement Techniques
spelling doaj.art-e91dbcc197d84cf386203b462e4d8fe42022-12-21T19:04:25ZengCopernicus PublicationsAtmospheric Measurement Techniques1867-13811867-85482020-02-011394194910.5194/amt-13-941-2020Increase of the particle hit rate in a laser single-particle mass spectrometer by pulse delayed extraction technologyY. Chen0Y. Chen1V. Kozlovskiy2X. Du3X. Du4J. Lv5S. Nikiforov6J. Yu7J. Yu8A. Kolosov9W. Gao10W. Gao11Z. Zhou12Z. Zhou13Z. Huang14Z. Huang15L. Li16L. Li17Institute of Mass Spectrometer and Atmospheric Environment, Jinan University, Guangzhou, 510632, ChinaGuangdong Provincial Engineering Research Center for On-Line Source Apportionment System of Air Pollution, Guangzhou, 510632, ChinaChernogolovka Branch of the N. N. Semenov Federal Research Center for Chemical Physics, Russian Academy of Sciences, Chernogolovka, 142432, RussiaInstitute of Mass Spectrometer and Atmospheric Environment, Jinan University, Guangzhou, 510632, ChinaGuangdong Provincial Engineering Research Center for On-Line Source Apportionment System of Air Pollution, Guangzhou, 510632, ChinaGuangzhou Hexin Analytical Instrument Company, Ltd., Guangzhou, 510530, ChinaProkhorov General Physics Institute, Russian Academy of Science, Moscow, 119991, RussiaInstitute of Mass Spectrometer and Atmospheric Environment, Jinan University, Guangzhou, 510632, ChinaGuangdong Provincial Engineering Research Center for On-Line Source Apportionment System of Air Pollution, Guangzhou, 510632, ChinaN. N. Semenov Federal Research Center for Chemical Physics, Russian Academy of Sciences, Moscow, 119991, RussiaInstitute of Mass Spectrometer and Atmospheric Environment, Jinan University, Guangzhou, 510632, ChinaGuangdong Provincial Engineering Research Center for On-Line Source Apportionment System of Air Pollution, Guangzhou, 510632, ChinaInstitute of Mass Spectrometer and Atmospheric Environment, Jinan University, Guangzhou, 510632, ChinaGuangdong Provincial Engineering Research Center for On-Line Source Apportionment System of Air Pollution, Guangzhou, 510632, ChinaInstitute of Mass Spectrometer and Atmospheric Environment, Jinan University, Guangzhou, 510632, ChinaGuangdong Provincial Engineering Research Center for On-Line Source Apportionment System of Air Pollution, Guangzhou, 510632, ChinaInstitute of Mass Spectrometer and Atmospheric Environment, Jinan University, Guangzhou, 510632, ChinaGuangdong Provincial Engineering Research Center for On-Line Source Apportionment System of Air Pollution, Guangzhou, 510632, China<p>A single-particle mass spectrometer (SPMS) can provide a wealth of valuable information on chemical and physical parameters of individual particles in real time. One of the main performance criteria of the instrument is efficiency of particle detection (hit rate). Most SPMS instruments use constant electrical field (DC) extraction, where stationary high voltage is applied to the extraction electrodes. As the aerosol particles initially carry a certain charge, those with a high amount to charge can be deflected by this electric field and lost, thus decreasing the hit rate. We realized that the delayed extraction technique can eliminate the stochastic dispersion of the particle beam caused by their deflection in the stationary electric field. As the result, the hit rate of the instrument can be significantly improved. Also, as the effect of the deflection in the electric field is mass dependent, it can cause distortion of the measured size distribution of the particles. Hence, the delayed extraction technique can bring the recorded distribution closer to the actual one. We found that the delayed extraction technique provides a mass resolution improvement as well as increases the hit rate. The gain in the hit rate depends on the type of particles. It can be 2 orders of magnitude for model particles and up to 2–4 times for ambient particles. In the present work we report experiments and results showing the effect of the delayed extraction on the beam divergence caused by particle charge, the hit rate improvement, and the effect of the delayed extraction on the measured particle size distribution.</p>https://www.atmos-meas-tech.net/13/941/2020/amt-13-941-2020.pdf
spellingShingle Y. Chen
Y. Chen
V. Kozlovskiy
X. Du
X. Du
J. Lv
S. Nikiforov
J. Yu
J. Yu
A. Kolosov
W. Gao
W. Gao
Z. Zhou
Z. Zhou
Z. Huang
Z. Huang
L. Li
L. Li
Increase of the particle hit rate in a laser single-particle mass spectrometer by pulse delayed extraction technology
Atmospheric Measurement Techniques
title Increase of the particle hit rate in a laser single-particle mass spectrometer by pulse delayed extraction technology
title_full Increase of the particle hit rate in a laser single-particle mass spectrometer by pulse delayed extraction technology
title_fullStr Increase of the particle hit rate in a laser single-particle mass spectrometer by pulse delayed extraction technology
title_full_unstemmed Increase of the particle hit rate in a laser single-particle mass spectrometer by pulse delayed extraction technology
title_short Increase of the particle hit rate in a laser single-particle mass spectrometer by pulse delayed extraction technology
title_sort increase of the particle hit rate in a laser single particle mass spectrometer by pulse delayed extraction technology
url https://www.atmos-meas-tech.net/13/941/2020/amt-13-941-2020.pdf
work_keys_str_mv AT ychen increaseoftheparticlehitrateinalasersingleparticlemassspectrometerbypulsedelayedextractiontechnology
AT ychen increaseoftheparticlehitrateinalasersingleparticlemassspectrometerbypulsedelayedextractiontechnology
AT vkozlovskiy increaseoftheparticlehitrateinalasersingleparticlemassspectrometerbypulsedelayedextractiontechnology
AT xdu increaseoftheparticlehitrateinalasersingleparticlemassspectrometerbypulsedelayedextractiontechnology
AT xdu increaseoftheparticlehitrateinalasersingleparticlemassspectrometerbypulsedelayedextractiontechnology
AT jlv increaseoftheparticlehitrateinalasersingleparticlemassspectrometerbypulsedelayedextractiontechnology
AT snikiforov increaseoftheparticlehitrateinalasersingleparticlemassspectrometerbypulsedelayedextractiontechnology
AT jyu increaseoftheparticlehitrateinalasersingleparticlemassspectrometerbypulsedelayedextractiontechnology
AT jyu increaseoftheparticlehitrateinalasersingleparticlemassspectrometerbypulsedelayedextractiontechnology
AT akolosov increaseoftheparticlehitrateinalasersingleparticlemassspectrometerbypulsedelayedextractiontechnology
AT wgao increaseoftheparticlehitrateinalasersingleparticlemassspectrometerbypulsedelayedextractiontechnology
AT wgao increaseoftheparticlehitrateinalasersingleparticlemassspectrometerbypulsedelayedextractiontechnology
AT zzhou increaseoftheparticlehitrateinalasersingleparticlemassspectrometerbypulsedelayedextractiontechnology
AT zzhou increaseoftheparticlehitrateinalasersingleparticlemassspectrometerbypulsedelayedextractiontechnology
AT zhuang increaseoftheparticlehitrateinalasersingleparticlemassspectrometerbypulsedelayedextractiontechnology
AT zhuang increaseoftheparticlehitrateinalasersingleparticlemassspectrometerbypulsedelayedextractiontechnology
AT lli increaseoftheparticlehitrateinalasersingleparticlemassspectrometerbypulsedelayedextractiontechnology
AT lli increaseoftheparticlehitrateinalasersingleparticlemassspectrometerbypulsedelayedextractiontechnology