Clinical validation of a contactless respiration rate monitor
Abstract Respiratory rate (RR) is an often underestimated and underreported vital sign with tremendous clinical value. As a predictor of cardiopulmonary arrest, chronic obstructive pulmonary disease (COPD) exacerbation or indicator of health state for example in COVID-19 patients, respiratory rate c...
Main Authors: | , , , , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Nature Portfolio
2023-03-01
|
Series: | Scientific Reports |
Online Access: | https://doi.org/10.1038/s41598-023-30171-4 |
_version_ | 1827984349674340352 |
---|---|
author | Bartosz Bujan Tobit Fischer Sarah Dietz-Terjung Aribert Bauerfeind Piotr Jedrysiak Martina Große Sundrup Janne Hamann Christoph Schöbel |
author_facet | Bartosz Bujan Tobit Fischer Sarah Dietz-Terjung Aribert Bauerfeind Piotr Jedrysiak Martina Große Sundrup Janne Hamann Christoph Schöbel |
author_sort | Bartosz Bujan |
collection | DOAJ |
description | Abstract Respiratory rate (RR) is an often underestimated and underreported vital sign with tremendous clinical value. As a predictor of cardiopulmonary arrest, chronic obstructive pulmonary disease (COPD) exacerbation or indicator of health state for example in COVID-19 patients, respiratory rate could be especially valuable in remote long-term patient monitoring, which is challenging to implement. Contactless devices for home use aim to overcome these challenges. In this study, the contactless Sleepiz One+ respiration monitor for home use during sleep was validated against the thoracic effort belt. The agreement of instantaneous breathing rate and breathing rate statistics between the Sleepiz One+ device and the thoracic effort belt was initially evaluated during a 20-min sleep window under controlled conditions (no body movement) on a cohort of 19 participants and secondly in a more natural setting (uncontrolled for body movement) during a whole night on a cohort of 139 participants. Excellent agreement was shown for instantaneous breathing rate to be within 3 breaths per minute (Brpm) compared to thoracic effort band with an accuracy of 100% and mean absolute error (MAE) of 0.39 Brpm for the setting controlled for movement, and an accuracy of 99.5% with a MAE of 0.48 Brpm for the whole night measurement, respectively. Excellent agreement was also achieved for the respiratory rate statistics over the whole night with absolute errors of 0.43, 0.39 and 0.67 Brpm for the 10th, 50th and 90th percentiles, respectively. Based on these results we conclude that the Sleepiz One+ can estimate instantaneous respiratory rate and its summary statistics at high accuracy in a clinical setting. Further studies are required to evaluate the performance in the home environment, however, it is expected that the performance is at similar level, as the measurement conditions for the Sleepiz One+ device are better at home than in a clinical setting. |
first_indexed | 2024-04-09T22:58:54Z |
format | Article |
id | doaj.art-e92c319cb9394f78b85b3f710bc40c90 |
institution | Directory Open Access Journal |
issn | 2045-2322 |
language | English |
last_indexed | 2024-04-09T22:58:54Z |
publishDate | 2023-03-01 |
publisher | Nature Portfolio |
record_format | Article |
series | Scientific Reports |
spelling | doaj.art-e92c319cb9394f78b85b3f710bc40c902023-03-22T11:09:27ZengNature PortfolioScientific Reports2045-23222023-03-0113111210.1038/s41598-023-30171-4Clinical validation of a contactless respiration rate monitorBartosz Bujan0Tobit Fischer1Sarah Dietz-Terjung2Aribert Bauerfeind3Piotr Jedrysiak4Martina Große Sundrup5Janne Hamann6Christoph Schöbel7Klinik Lengg AG, Neurorehabilitation CenterEssen University Hospital, RuhrlandklinikEssen University Hospital, RuhrlandklinikKlinik Lengg AG, Swiss Epilepsy CenterEssen University Hospital, Neurorehabilitation CenterEssen University Hospital, RuhrlandklinikKlinik Lengg AG, Swiss Epilepsy CenterEssen University Hospital, RuhrlandklinikAbstract Respiratory rate (RR) is an often underestimated and underreported vital sign with tremendous clinical value. As a predictor of cardiopulmonary arrest, chronic obstructive pulmonary disease (COPD) exacerbation or indicator of health state for example in COVID-19 patients, respiratory rate could be especially valuable in remote long-term patient monitoring, which is challenging to implement. Contactless devices for home use aim to overcome these challenges. In this study, the contactless Sleepiz One+ respiration monitor for home use during sleep was validated against the thoracic effort belt. The agreement of instantaneous breathing rate and breathing rate statistics between the Sleepiz One+ device and the thoracic effort belt was initially evaluated during a 20-min sleep window under controlled conditions (no body movement) on a cohort of 19 participants and secondly in a more natural setting (uncontrolled for body movement) during a whole night on a cohort of 139 participants. Excellent agreement was shown for instantaneous breathing rate to be within 3 breaths per minute (Brpm) compared to thoracic effort band with an accuracy of 100% and mean absolute error (MAE) of 0.39 Brpm for the setting controlled for movement, and an accuracy of 99.5% with a MAE of 0.48 Brpm for the whole night measurement, respectively. Excellent agreement was also achieved for the respiratory rate statistics over the whole night with absolute errors of 0.43, 0.39 and 0.67 Brpm for the 10th, 50th and 90th percentiles, respectively. Based on these results we conclude that the Sleepiz One+ can estimate instantaneous respiratory rate and its summary statistics at high accuracy in a clinical setting. Further studies are required to evaluate the performance in the home environment, however, it is expected that the performance is at similar level, as the measurement conditions for the Sleepiz One+ device are better at home than in a clinical setting.https://doi.org/10.1038/s41598-023-30171-4 |
spellingShingle | Bartosz Bujan Tobit Fischer Sarah Dietz-Terjung Aribert Bauerfeind Piotr Jedrysiak Martina Große Sundrup Janne Hamann Christoph Schöbel Clinical validation of a contactless respiration rate monitor Scientific Reports |
title | Clinical validation of a contactless respiration rate monitor |
title_full | Clinical validation of a contactless respiration rate monitor |
title_fullStr | Clinical validation of a contactless respiration rate monitor |
title_full_unstemmed | Clinical validation of a contactless respiration rate monitor |
title_short | Clinical validation of a contactless respiration rate monitor |
title_sort | clinical validation of a contactless respiration rate monitor |
url | https://doi.org/10.1038/s41598-023-30171-4 |
work_keys_str_mv | AT bartoszbujan clinicalvalidationofacontactlessrespirationratemonitor AT tobitfischer clinicalvalidationofacontactlessrespirationratemonitor AT sarahdietzterjung clinicalvalidationofacontactlessrespirationratemonitor AT aribertbauerfeind clinicalvalidationofacontactlessrespirationratemonitor AT piotrjedrysiak clinicalvalidationofacontactlessrespirationratemonitor AT martinagroßesundrup clinicalvalidationofacontactlessrespirationratemonitor AT jannehamann clinicalvalidationofacontactlessrespirationratemonitor AT christophschobel clinicalvalidationofacontactlessrespirationratemonitor |