Summary: | To evaluate the properties of two nickel–titanium (NiTi) reciprocating endodontic instruments (commercially known as Procodile and Reziflow), a total of 40 size 25 and 0.06 taper new Procodile and Reziflow instruments (<i>n</i> = 20) were subjected to cyclic fatigue tests (60° angle of curvature, 5-mm radius) at 20 °C and 37 °C and a torsional test based on ISO 3630-1. The fracture surface of each fragment was examined. The morphological, mechanical, chemical, thermal, and phase composition characteristics of the files were investigated by field-emission gun scanning electron microscopy (FEG-SEM) equipped with an energy-dispersive X-ray (EDX) detector, focused ion beam analysis (FIB), micro-Raman spectroscopy, X-ray diffraction (XRD), differential scanning calorimetry (DSC), and Auger electron spectroscopy (AES). Reziflow showed higher cyclic fatigue resistance than Procodile at 37 °C (<i>p</i> < 0.05). The maximum torsional strength of Procodile was lower than that of Reziflow (<i>p</i> < 0.05). No difference was found between their angular rotations to fracture (<i>p</i> > 0.05). SEM, FIB, Micro-Raman, and AES analyses revealed the presence of an Nb/Nb<sub>2</sub>O<sub>5</sub> coating on the Procodile surface. DSC and XRD analysis confirmed that both files consist of an almost austenitic phase structure at 37 °C. The cyclic fatigue resistance of Procodile and Reziflow significantly decreases upon exposure to body temperature.
|