New Technique for Impact Calibration of Wide-Range Triaxial Force Transducer Using Hopkinson Bar
At the current stage, there is an urgent need for new techniques to dynamically calibrate advanced wide-range (up to 10<sup>4</sup> N~10<sup>5</sup> N) triaxial force transducers. Based on this background, this paper proposes a novel impact calibration method, specifically fo...
Main Authors: | , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
MDPI AG
2022-06-01
|
Series: | Sensors |
Subjects: | |
Online Access: | https://www.mdpi.com/1424-8220/22/13/4885 |
_version_ | 1797441998096957440 |
---|---|
author | Qinghua Wang Feng Xu Weiguo Guo Meng Gao |
author_facet | Qinghua Wang Feng Xu Weiguo Guo Meng Gao |
author_sort | Qinghua Wang |
collection | DOAJ |
description | At the current stage, there is an urgent need for new techniques to dynamically calibrate advanced wide-range (up to 10<sup>4</sup> N~10<sup>5</sup> N) triaxial force transducers. Based on this background, this paper proposes a novel impact calibration method, specifically for the triaxial force transducer, with a wide range and high-frequency response. In this method, the Hopkinson bar, which is typically used to test the dynamic mechanical properties of materials, was used as a generator to generate reference input force for the transducer. In addition, unlike conventional methods, the transverse sensitivities of the transducer were given necessary importance in the proposed method. The calibration result of the triaxial force transducer was expressed in a sensitivity matrix, containing three main sensitivity coefficients and six transverse sensitivity coefficients. The least squares method (LSM) was used to fit the sensitivity matrix linearly. Calibration experiments were performed on a typical triaxial force transducer. Several key issues, involving the validity and the test range, of the method were further investigated numerically. The feasibility and validity of the method were eventually confirmed. The test range of the method can be up to 10<sup>6</sup> N. |
first_indexed | 2024-03-09T12:34:21Z |
format | Article |
id | doaj.art-ec704f30d90d44edba4d74205a88333e |
institution | Directory Open Access Journal |
issn | 1424-8220 |
language | English |
last_indexed | 2024-03-09T12:34:21Z |
publishDate | 2022-06-01 |
publisher | MDPI AG |
record_format | Article |
series | Sensors |
spelling | doaj.art-ec704f30d90d44edba4d74205a88333e2023-11-30T22:26:30ZengMDPI AGSensors1424-82202022-06-012213488510.3390/s22134885New Technique for Impact Calibration of Wide-Range Triaxial Force Transducer Using Hopkinson BarQinghua Wang0Feng Xu1Weiguo Guo2Meng Gao3School of Aeronautics, Northwestern Polytechnical University, Xi’an 710072, ChinaSchool of Astronautics, Northwestern Polytechnical University, Xi’an 710072, ChinaSchool of Aeronautics, Northwestern Polytechnical University, Xi’an 710072, ChinaSchool of Aeronautics, Northwestern Polytechnical University, Xi’an 710072, ChinaAt the current stage, there is an urgent need for new techniques to dynamically calibrate advanced wide-range (up to 10<sup>4</sup> N~10<sup>5</sup> N) triaxial force transducers. Based on this background, this paper proposes a novel impact calibration method, specifically for the triaxial force transducer, with a wide range and high-frequency response. In this method, the Hopkinson bar, which is typically used to test the dynamic mechanical properties of materials, was used as a generator to generate reference input force for the transducer. In addition, unlike conventional methods, the transverse sensitivities of the transducer were given necessary importance in the proposed method. The calibration result of the triaxial force transducer was expressed in a sensitivity matrix, containing three main sensitivity coefficients and six transverse sensitivity coefficients. The least squares method (LSM) was used to fit the sensitivity matrix linearly. Calibration experiments were performed on a typical triaxial force transducer. Several key issues, involving the validity and the test range, of the method were further investigated numerically. The feasibility and validity of the method were eventually confirmed. The test range of the method can be up to 10<sup>6</sup> N.https://www.mdpi.com/1424-8220/22/13/4885triaxial force transducerwide-rangedynamic calibrationHopkinson barsensitivity matrix |
spellingShingle | Qinghua Wang Feng Xu Weiguo Guo Meng Gao New Technique for Impact Calibration of Wide-Range Triaxial Force Transducer Using Hopkinson Bar Sensors triaxial force transducer wide-range dynamic calibration Hopkinson bar sensitivity matrix |
title | New Technique for Impact Calibration of Wide-Range Triaxial Force Transducer Using Hopkinson Bar |
title_full | New Technique for Impact Calibration of Wide-Range Triaxial Force Transducer Using Hopkinson Bar |
title_fullStr | New Technique for Impact Calibration of Wide-Range Triaxial Force Transducer Using Hopkinson Bar |
title_full_unstemmed | New Technique for Impact Calibration of Wide-Range Triaxial Force Transducer Using Hopkinson Bar |
title_short | New Technique for Impact Calibration of Wide-Range Triaxial Force Transducer Using Hopkinson Bar |
title_sort | new technique for impact calibration of wide range triaxial force transducer using hopkinson bar |
topic | triaxial force transducer wide-range dynamic calibration Hopkinson bar sensitivity matrix |
url | https://www.mdpi.com/1424-8220/22/13/4885 |
work_keys_str_mv | AT qinghuawang newtechniqueforimpactcalibrationofwiderangetriaxialforcetransducerusinghopkinsonbar AT fengxu newtechniqueforimpactcalibrationofwiderangetriaxialforcetransducerusinghopkinsonbar AT weiguoguo newtechniqueforimpactcalibrationofwiderangetriaxialforcetransducerusinghopkinsonbar AT menggao newtechniqueforimpactcalibrationofwiderangetriaxialforcetransducerusinghopkinsonbar |