In-vitro evaluation of the effects of insertion and sterilization procedures on the mechanical and surface characteristics of mini screws
PurposeThe aim of the present study was to investigate the effects of insertion andsterilization on primary stability and to examine the mechanical and surfacecharacteristics of mini screws.Materials and methods140 miniscrews (70 Dual-Top; 70 Ortho-Easy) were divided into 3 groups. Group1: control g...
Main Authors: | , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Istanbul University
2019-01-01
|
Series: | European Oral Research |
Subjects: | |
Online Access: | https://dergipark.org.tr/tr/pub/eor/issue/43439/529673 |
_version_ | 1797920336346349568 |
---|---|
author | Nazan Küçükkeleş Mustafa Ateş Yasemin Bahar Acar Civan Aved Hergel |
author_facet | Nazan Küçükkeleş Mustafa Ateş Yasemin Bahar Acar Civan Aved Hergel |
author_sort | Nazan Küçükkeleş |
collection | DOAJ |
description | PurposeThe aim of the present study was to investigate the effects of insertion andsterilization on primary stability and to examine the mechanical and surfacecharacteristics of mini screws.Materials and methods140 miniscrews (70 Dual-Top; 70 Ortho-Easy) were divided into 3 groups. Group1: control group, 10 miniscrews of each brand, evaluated without any primaryprocedure. Group 2: 30 miniscrews of each brand, each inserted into the sawboneonce, then sterilized and tested. Group 3: 30 miniscrews of each brand, eachinserted into the sawbone twice, sterilized after each insertion and then tested. Theminiscrews were evaluated for changes in primary stability, mechanical and surfacecharacteristics with scanning electron microscopy (SEM) analysis, torsion tests,maximum insertion-removal torques and vertical-horizontal pull out strength tests.ResultsThe maximum insertion torque values of the unused miniscrews (Group 1) werefound to be significantly higher than those of the reused (Groups 2, 3) mini screws(p<0.05). Removal torque, vertical-horizontal pull-out strength and torsionalstrength value changes were found to be statistically insignificant. In SEM analysis,wear and atrophy were seen on the threads of used miniscrews especially in theapical region and the oxide layer was seen to have disappeared from some regionsof the coated miniscrews.ConclusionAlthough wear and atrophy were detected in SEM analysis of used miniscrews,the overall primary stability and fracture torque resistance tests did not show anysignificant changes after the first and second insertion and sterilization procedures. |
first_indexed | 2024-04-10T13:59:30Z |
format | Article |
id | doaj.art-fb8d21834b8145268b11e0434851b33d |
institution | Directory Open Access Journal |
issn | 2630-6158 2651-2823 |
language | English |
last_indexed | 2024-04-10T13:59:30Z |
publishDate | 2019-01-01 |
publisher | Istanbul University |
record_format | Article |
series | European Oral Research |
spelling | doaj.art-fb8d21834b8145268b11e0434851b33d2023-02-15T16:10:19ZengIstanbul UniversityEuropean Oral Research2630-61582651-28232019-01-0153125314In-vitro evaluation of the effects of insertion and sterilization procedures on the mechanical and surface characteristics of mini screwsNazan Küçükkeleş0Mustafa Ateş1Yasemin Bahar Acar2Civan Aved Hergel3Private practice,Private practice,Marmara UniversityPrivate practicePurposeThe aim of the present study was to investigate the effects of insertion andsterilization on primary stability and to examine the mechanical and surfacecharacteristics of mini screws.Materials and methods140 miniscrews (70 Dual-Top; 70 Ortho-Easy) were divided into 3 groups. Group1: control group, 10 miniscrews of each brand, evaluated without any primaryprocedure. Group 2: 30 miniscrews of each brand, each inserted into the sawboneonce, then sterilized and tested. Group 3: 30 miniscrews of each brand, eachinserted into the sawbone twice, sterilized after each insertion and then tested. Theminiscrews were evaluated for changes in primary stability, mechanical and surfacecharacteristics with scanning electron microscopy (SEM) analysis, torsion tests,maximum insertion-removal torques and vertical-horizontal pull out strength tests.ResultsThe maximum insertion torque values of the unused miniscrews (Group 1) werefound to be significantly higher than those of the reused (Groups 2, 3) mini screws(p<0.05). Removal torque, vertical-horizontal pull-out strength and torsionalstrength value changes were found to be statistically insignificant. In SEM analysis,wear and atrophy were seen on the threads of used miniscrews especially in theapical region and the oxide layer was seen to have disappeared from some regionsof the coated miniscrews.ConclusionAlthough wear and atrophy were detected in SEM analysis of used miniscrews,the overall primary stability and fracture torque resistance tests did not show anysignificant changes after the first and second insertion and sterilization procedures.https://dergipark.org.tr/tr/pub/eor/issue/43439/529673miniscreworthodontic mini-implantprimary stabilitycanning electron microscopesterilization |
spellingShingle | Nazan Küçükkeleş Mustafa Ateş Yasemin Bahar Acar Civan Aved Hergel In-vitro evaluation of the effects of insertion and sterilization procedures on the mechanical and surface characteristics of mini screws European Oral Research miniscrew orthodontic mini-implant primary stability canning electron microscope sterilization |
title | In-vitro evaluation of the effects of insertion and sterilization procedures on the mechanical and surface characteristics of mini screws |
title_full | In-vitro evaluation of the effects of insertion and sterilization procedures on the mechanical and surface characteristics of mini screws |
title_fullStr | In-vitro evaluation of the effects of insertion and sterilization procedures on the mechanical and surface characteristics of mini screws |
title_full_unstemmed | In-vitro evaluation of the effects of insertion and sterilization procedures on the mechanical and surface characteristics of mini screws |
title_short | In-vitro evaluation of the effects of insertion and sterilization procedures on the mechanical and surface characteristics of mini screws |
title_sort | in vitro evaluation of the effects of insertion and sterilization procedures on the mechanical and surface characteristics of mini screws |
topic | miniscrew orthodontic mini-implant primary stability canning electron microscope sterilization |
url | https://dergipark.org.tr/tr/pub/eor/issue/43439/529673 |
work_keys_str_mv | AT nazankucukkeles invitroevaluationoftheeffectsofinsertionandsterilizationproceduresonthemechanicalandsurfacecharacteristicsofminiscrews AT mustafaates invitroevaluationoftheeffectsofinsertionandsterilizationproceduresonthemechanicalandsurfacecharacteristicsofminiscrews AT yaseminbaharacar invitroevaluationoftheeffectsofinsertionandsterilizationproceduresonthemechanicalandsurfacecharacteristicsofminiscrews AT civanavedhergel invitroevaluationoftheeffectsofinsertionandsterilizationproceduresonthemechanicalandsurfacecharacteristicsofminiscrews |