Interstudy reproducibility of the second generation, Fourier domain optical coherence tomography in patients with coronary artery disease and comparison with intravascular ultrasound: a study applying automated contour detection

Recently, Fourier domain OCT (FD-OCT) has been introduced for clinical use. This approach allows in vivo, high resolution (15 micron) imaging with very fast data acquisition, however, it requires brief flushing of the lumen during imaging. The reproducibility of such fast data acquisition under intr...

Full description

Bibliographic Details
Main Authors: Jamil, Z., Bruining, N., Sihan, K., Soest, G., Ligthart, J., Domburg, R., Regar, E., Tearney, Guillermo J., Bouma, Brett E.
Other Authors: Harvard University--MIT Division of Health Sciences and Technology
Format: Article
Language:en_US
Published: Springer-Verlag 2013
Online Access:http://hdl.handle.net/1721.1/82657
_version_ 1811092954145619968
author Jamil, Z.
Bruining, N.
Sihan, K.
Soest, G.
Ligthart, J.
Domburg, R.
Regar, E.
Tearney, Guillermo J.
Bouma, Brett E.
author2 Harvard University--MIT Division of Health Sciences and Technology
author_facet Harvard University--MIT Division of Health Sciences and Technology
Jamil, Z.
Bruining, N.
Sihan, K.
Soest, G.
Ligthart, J.
Domburg, R.
Regar, E.
Tearney, Guillermo J.
Bouma, Brett E.
author_sort Jamil, Z.
collection MIT
description Recently, Fourier domain OCT (FD-OCT) has been introduced for clinical use. This approach allows in vivo, high resolution (15 micron) imaging with very fast data acquisition, however, it requires brief flushing of the lumen during imaging. The reproducibility of such fast data acquisition under intracoronary flush application is poorly understood. To assess the inter-study variability of FD-OCT and to compare lumen morphometry to the established invasive imaging method, IVUS. 18 consecutive patients with coronary artery disease scheduled for PCI were included. In each target vessel a FD-OCT pullback (MGH system, light source 1,310 nm, 105 fps, pullback speed 20 mm/s) was acquired during brief (3 s) injection of X-ray contrast (flow 3 ml/s) through the guiding catheter. A second pullback was repeated under the same conditions after re-introduction of the FD OCT catheter into the coronary artery. IVUS and OCT imaging was performed in random order. FD-OCT and IVUS pullback data were analyzed using a recently developed software employing semi automated lumen contour and stent strut detection algorithms. Corresponding ROI were matched based on anatomical landmarks such as side branches and/or stent edges. Inter-study variability is presented as the absolute difference between the two pullbacks. FD-OCT showed remarkably good reproducibility. Inter-study variability in native vessels (cohort A) was very low for mean and minimal luminal area (0.10 ± 0.38, 0.19 ± 0.57 mm[superscript 2], respectively). Likewise inter-study variability was very low in stented coronary segments (cohort B) for mean lumen, mean stent, minimal luminal and minimal stent area (0.06 ± 0.08, 0.07 ± 0.10, 0.04 ± 0.09, 0.04 ± 0.10 mm[superscript 2], respectively). Comparison to IVUS morphometry revealed no significant differences. The differences between both imaging methods, OCT and IVUS, were very low for mean lumen, mean stent, minimal luminal and minimal stent area (0.10 ± 0.45, 0.10 ± 0.36, 0.26 ± 0.54, 0.05 ± 0.47 mm[superscript 2], respectively). FD-OCT shows excellent reproducibility and very low inter-study variability in both, native and stented coronary segments. No significant differences in quantitative lumen morphometry were observed between FD-OCT and IVUS. Evaluating these results suggest that FD-OCT is a reliable imaging tool to apply in longitudinal coronary artery disease studies
first_indexed 2024-09-23T15:36:07Z
format Article
id mit-1721.1/82657
institution Massachusetts Institute of Technology
language en_US
last_indexed 2024-09-23T15:36:07Z
publishDate 2013
publisher Springer-Verlag
record_format dspace
spelling mit-1721.1/826572022-09-29T14:59:19Z Interstudy reproducibility of the second generation, Fourier domain optical coherence tomography in patients with coronary artery disease and comparison with intravascular ultrasound: a study applying automated contour detection Jamil, Z. Bruining, N. Sihan, K. Soest, G. Ligthart, J. Domburg, R. Regar, E. Tearney, Guillermo J. Bouma, Brett E. Harvard University--MIT Division of Health Sciences and Technology Tearney, Guillermo J. Bouma, Brett E. Recently, Fourier domain OCT (FD-OCT) has been introduced for clinical use. This approach allows in vivo, high resolution (15 micron) imaging with very fast data acquisition, however, it requires brief flushing of the lumen during imaging. The reproducibility of such fast data acquisition under intracoronary flush application is poorly understood. To assess the inter-study variability of FD-OCT and to compare lumen morphometry to the established invasive imaging method, IVUS. 18 consecutive patients with coronary artery disease scheduled for PCI were included. In each target vessel a FD-OCT pullback (MGH system, light source 1,310 nm, 105 fps, pullback speed 20 mm/s) was acquired during brief (3 s) injection of X-ray contrast (flow 3 ml/s) through the guiding catheter. A second pullback was repeated under the same conditions after re-introduction of the FD OCT catheter into the coronary artery. IVUS and OCT imaging was performed in random order. FD-OCT and IVUS pullback data were analyzed using a recently developed software employing semi automated lumen contour and stent strut detection algorithms. Corresponding ROI were matched based on anatomical landmarks such as side branches and/or stent edges. Inter-study variability is presented as the absolute difference between the two pullbacks. FD-OCT showed remarkably good reproducibility. Inter-study variability in native vessels (cohort A) was very low for mean and minimal luminal area (0.10 ± 0.38, 0.19 ± 0.57 mm[superscript 2], respectively). Likewise inter-study variability was very low in stented coronary segments (cohort B) for mean lumen, mean stent, minimal luminal and minimal stent area (0.06 ± 0.08, 0.07 ± 0.10, 0.04 ± 0.09, 0.04 ± 0.10 mm[superscript 2], respectively). Comparison to IVUS morphometry revealed no significant differences. The differences between both imaging methods, OCT and IVUS, were very low for mean lumen, mean stent, minimal luminal and minimal stent area (0.10 ± 0.45, 0.10 ± 0.36, 0.26 ± 0.54, 0.05 ± 0.47 mm[superscript 2], respectively). FD-OCT shows excellent reproducibility and very low inter-study variability in both, native and stented coronary segments. No significant differences in quantitative lumen morphometry were observed between FD-OCT and IVUS. Evaluating these results suggest that FD-OCT is a reliable imaging tool to apply in longitudinal coronary artery disease studies 2013-12-06T14:40:35Z 2013-12-06T14:40:35Z 2012-05 2011-09 Article http://purl.org/eprint/type/JournalArticle 1569-5794 1573-0743 http://hdl.handle.net/1721.1/82657 Jamil, Z., G. Tearney, N. Bruining, K. Sihan, G. Soest, J. Ligthart, R. Domburg, B. Bouma, and E. Regar. “Interstudy reproducibility of the second generation, Fourier domain optical coherence tomography in patients with coronary artery disease and comparison with intravascular ultrasound: a study applying automated contour detection.” The International Journal of Cardiovascular Imaging 29, no. 1 (January 26, 2013): 39-51. en_US http://dx.doi.org/10.1007/s10554-012-0067-8 The International Journal of Cardiovascular Imaging http://creativecommons.org/licenses/by/2.0 application/pdf Springer-Verlag PMC
spellingShingle Jamil, Z.
Bruining, N.
Sihan, K.
Soest, G.
Ligthart, J.
Domburg, R.
Regar, E.
Tearney, Guillermo J.
Bouma, Brett E.
Interstudy reproducibility of the second generation, Fourier domain optical coherence tomography in patients with coronary artery disease and comparison with intravascular ultrasound: a study applying automated contour detection
title Interstudy reproducibility of the second generation, Fourier domain optical coherence tomography in patients with coronary artery disease and comparison with intravascular ultrasound: a study applying automated contour detection
title_full Interstudy reproducibility of the second generation, Fourier domain optical coherence tomography in patients with coronary artery disease and comparison with intravascular ultrasound: a study applying automated contour detection
title_fullStr Interstudy reproducibility of the second generation, Fourier domain optical coherence tomography in patients with coronary artery disease and comparison with intravascular ultrasound: a study applying automated contour detection
title_full_unstemmed Interstudy reproducibility of the second generation, Fourier domain optical coherence tomography in patients with coronary artery disease and comparison with intravascular ultrasound: a study applying automated contour detection
title_short Interstudy reproducibility of the second generation, Fourier domain optical coherence tomography in patients with coronary artery disease and comparison with intravascular ultrasound: a study applying automated contour detection
title_sort interstudy reproducibility of the second generation fourier domain optical coherence tomography in patients with coronary artery disease and comparison with intravascular ultrasound a study applying automated contour detection
url http://hdl.handle.net/1721.1/82657
work_keys_str_mv AT jamilz interstudyreproducibilityofthesecondgenerationfourierdomainopticalcoherencetomographyinpatientswithcoronaryarterydiseaseandcomparisonwithintravascularultrasoundastudyapplyingautomatedcontourdetection
AT bruiningn interstudyreproducibilityofthesecondgenerationfourierdomainopticalcoherencetomographyinpatientswithcoronaryarterydiseaseandcomparisonwithintravascularultrasoundastudyapplyingautomatedcontourdetection
AT sihank interstudyreproducibilityofthesecondgenerationfourierdomainopticalcoherencetomographyinpatientswithcoronaryarterydiseaseandcomparisonwithintravascularultrasoundastudyapplyingautomatedcontourdetection
AT soestg interstudyreproducibilityofthesecondgenerationfourierdomainopticalcoherencetomographyinpatientswithcoronaryarterydiseaseandcomparisonwithintravascularultrasoundastudyapplyingautomatedcontourdetection
AT ligthartj interstudyreproducibilityofthesecondgenerationfourierdomainopticalcoherencetomographyinpatientswithcoronaryarterydiseaseandcomparisonwithintravascularultrasoundastudyapplyingautomatedcontourdetection
AT domburgr interstudyreproducibilityofthesecondgenerationfourierdomainopticalcoherencetomographyinpatientswithcoronaryarterydiseaseandcomparisonwithintravascularultrasoundastudyapplyingautomatedcontourdetection
AT regare interstudyreproducibilityofthesecondgenerationfourierdomainopticalcoherencetomographyinpatientswithcoronaryarterydiseaseandcomparisonwithintravascularultrasoundastudyapplyingautomatedcontourdetection
AT tearneyguillermoj interstudyreproducibilityofthesecondgenerationfourierdomainopticalcoherencetomographyinpatientswithcoronaryarterydiseaseandcomparisonwithintravascularultrasoundastudyapplyingautomatedcontourdetection
AT boumabrette interstudyreproducibilityofthesecondgenerationfourierdomainopticalcoherencetomographyinpatientswithcoronaryarterydiseaseandcomparisonwithintravascularultrasoundastudyapplyingautomatedcontourdetection