Accelerated parallel magnetic resonance imaging reconstruction using joint estimation with a sparse signal model

Accelerating magnetic resonance imaging (MRI) by reducing the number of acquired k-space scan lines benefits conventional MRI significantly by decreasing the time subjects remain in the magnet. In this paper, we formulate a novel method for Joint estimation from Undersampled LinEs in Parallel MRI (J...

Full description

Bibliographic Details
Main Authors: Weller, Daniel S., Polimeni, Jonathan R., Grady, Leo, Wald, Lawrence L., Adalsteinsson, Elfar, Goyal, Vivek K.
Other Authors: Massachusetts Institute of Technology. Department of Electrical Engineering and Computer Science
Format: Article
Language:en_US
Published: Institute of Electrical and Electronics Engineers (IEEE) 2014
Online Access:http://hdl.handle.net/1721.1/85876
https://orcid.org/0000-0002-7637-2914
Description
Summary:Accelerating magnetic resonance imaging (MRI) by reducing the number of acquired k-space scan lines benefits conventional MRI significantly by decreasing the time subjects remain in the magnet. In this paper, we formulate a novel method for Joint estimation from Undersampled LinEs in Parallel MRI (JULEP) that simultaneously calibrates the GeneRalized Autocalibrating Partially Parallel Acquisitions (GRAPPA) reconstruction kernel and reconstructs the full multi-channel k-space. We employ a joint sparsity signal model for the channel images in conjunction with observation models for both the acquired data and GRAPPA reconstructed k-space. We demonstrate using real MRI data that JULEP outperforms conventional GRAPPA reconstruction at high levels of undersampling, increasing the peak-signal-to-noise ratio by up to 10 dB.