Numerical and in-vitro experimental assessment of the performance of a novel designed expanded-polytetrafluoroethylene stentless bi-leaflet valve for aortic valve replacement

The expanded polytetrafluoroethylene (ePTFE) heart valve can serve as a viable option for prosthetic aortic valve. In this study, an ePTFE bi-leaflet valve design for aortic valve replacement (AVR) is presented, and the performance of the proposed valve was assessed numerically and experimentally. T...

Full description

Bibliographic Details
Main Authors: Munirah Ismail, Zhu, Guangyu, Nakao, Masakazu, Yuan, Qi, Yeo, Joon Hock
Other Authors: Pacini, Davide
Format: Journal Article
Language:English
Published: 2019
Subjects:
Online Access:https://hdl.handle.net/10356/105942
http://hdl.handle.net/10220/48806
_version_ 1811679821735919616
author Munirah Ismail
Zhu, Guangyu
Nakao, Masakazu
Yuan, Qi
Yeo, Joon Hock
author2 Pacini, Davide
author_facet Pacini, Davide
Munirah Ismail
Zhu, Guangyu
Nakao, Masakazu
Yuan, Qi
Yeo, Joon Hock
author_sort Munirah Ismail
collection NTU
description The expanded polytetrafluoroethylene (ePTFE) heart valve can serve as a viable option for prosthetic aortic valve. In this study, an ePTFE bi-leaflet valve design for aortic valve replacement (AVR) is presented, and the performance of the proposed valve was assessed numerically and experimentally. The valve was designed using CAE software. The dynamic behavior of the newly designed bi-leaflet valve under time-varying physiological pressure loading was first investigated by using commercial finite element code. Then, in-vitro tests were performed to validate the simulation and to assess the hemodynamic performance of the proposed design. A tri-leaflet ePTFE valve was tested in-vitro under the same conditions as a reference. The maximum leaflet coaptation area of the bi-leaflet valve during diastole was 216.3 mm2. When fully closed, no leakage gap was observed and the free edges of the molded valve formed S-shaped lines. The maximum Von Mises stress during a full cardiac cycle was 4.20 MPa. The dynamic performance of the bi-leaflet valve was validated by the in-vitro test under physiological aortic pressure pulse. The effective orifice area (EOA), mean pressure gradient, regurgitant volume, leakage volume and energy loss of the proposed valve were 3.14 cm2, 8.74 mmHg, 5.93 ml/beat, 1.55 ml/beat and 98.99 mJ, respectively. This study reports a novel bi-leaflet valve design for AVR. The performance of the proposed valve was numerically and experimentally assessed. Compared with the reference valve, the proposed design exhibited better structural and hemodynamic performances, which improved valve competency. Moreover, the performance of the bi-leaflet design is comparable to commercialized valves available on the market. The results of the present study provide a viable option for the future clinical applications.
first_indexed 2024-10-01T03:15:15Z
format Journal Article
id ntu-10356/105942
institution Nanyang Technological University
language English
last_indexed 2024-10-01T03:15:15Z
publishDate 2019
record_format dspace
spelling ntu-10356/1059422023-03-04T17:15:53Z Numerical and in-vitro experimental assessment of the performance of a novel designed expanded-polytetrafluoroethylene stentless bi-leaflet valve for aortic valve replacement Munirah Ismail Zhu, Guangyu Nakao, Masakazu Yuan, Qi Yeo, Joon Hock Pacini, Davide School of Mechanical and Aerospace Engineering DRNTU::Engineering::Mechanical engineering Aortic Valve Prosthetics The expanded polytetrafluoroethylene (ePTFE) heart valve can serve as a viable option for prosthetic aortic valve. In this study, an ePTFE bi-leaflet valve design for aortic valve replacement (AVR) is presented, and the performance of the proposed valve was assessed numerically and experimentally. The valve was designed using CAE software. The dynamic behavior of the newly designed bi-leaflet valve under time-varying physiological pressure loading was first investigated by using commercial finite element code. Then, in-vitro tests were performed to validate the simulation and to assess the hemodynamic performance of the proposed design. A tri-leaflet ePTFE valve was tested in-vitro under the same conditions as a reference. The maximum leaflet coaptation area of the bi-leaflet valve during diastole was 216.3 mm2. When fully closed, no leakage gap was observed and the free edges of the molded valve formed S-shaped lines. The maximum Von Mises stress during a full cardiac cycle was 4.20 MPa. The dynamic performance of the bi-leaflet valve was validated by the in-vitro test under physiological aortic pressure pulse. The effective orifice area (EOA), mean pressure gradient, regurgitant volume, leakage volume and energy loss of the proposed valve were 3.14 cm2, 8.74 mmHg, 5.93 ml/beat, 1.55 ml/beat and 98.99 mJ, respectively. This study reports a novel bi-leaflet valve design for AVR. The performance of the proposed valve was numerically and experimentally assessed. Compared with the reference valve, the proposed design exhibited better structural and hemodynamic performances, which improved valve competency. Moreover, the performance of the bi-leaflet design is comparable to commercialized valves available on the market. The results of the present study provide a viable option for the future clinical applications. NMRC (Natl Medical Research Council, S’pore) Published version 2019-06-19T01:04:28Z 2019-12-06T22:01:13Z 2019-06-19T01:04:28Z 2019-12-06T22:01:13Z 2019 Journal Article Zhu, G., Munirah Ismail, Nakao, M., Yuan, Q., & Yeo, J. H. (2019). Numerical and in-vitro experimental assessment of the performance of a novel designed expanded-polytetrafluoroethylene stentless bi-leaflet valve for aortic valve replacement. PLOS ONE, 14(1), e0210780-. doi:10.1371/journal.pone.0210780 https://hdl.handle.net/10356/105942 http://hdl.handle.net/10220/48806 10.1371/journal.pone.0210780 en PLOS ONE © 2019 Zhu et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. 27 p. application/pdf
spellingShingle DRNTU::Engineering::Mechanical engineering
Aortic Valve
Prosthetics
Munirah Ismail
Zhu, Guangyu
Nakao, Masakazu
Yuan, Qi
Yeo, Joon Hock
Numerical and in-vitro experimental assessment of the performance of a novel designed expanded-polytetrafluoroethylene stentless bi-leaflet valve for aortic valve replacement
title Numerical and in-vitro experimental assessment of the performance of a novel designed expanded-polytetrafluoroethylene stentless bi-leaflet valve for aortic valve replacement
title_full Numerical and in-vitro experimental assessment of the performance of a novel designed expanded-polytetrafluoroethylene stentless bi-leaflet valve for aortic valve replacement
title_fullStr Numerical and in-vitro experimental assessment of the performance of a novel designed expanded-polytetrafluoroethylene stentless bi-leaflet valve for aortic valve replacement
title_full_unstemmed Numerical and in-vitro experimental assessment of the performance of a novel designed expanded-polytetrafluoroethylene stentless bi-leaflet valve for aortic valve replacement
title_short Numerical and in-vitro experimental assessment of the performance of a novel designed expanded-polytetrafluoroethylene stentless bi-leaflet valve for aortic valve replacement
title_sort numerical and in vitro experimental assessment of the performance of a novel designed expanded polytetrafluoroethylene stentless bi leaflet valve for aortic valve replacement
topic DRNTU::Engineering::Mechanical engineering
Aortic Valve
Prosthetics
url https://hdl.handle.net/10356/105942
http://hdl.handle.net/10220/48806
work_keys_str_mv AT munirahismail numericalandinvitroexperimentalassessmentoftheperformanceofanoveldesignedexpandedpolytetrafluoroethylenestentlessbileafletvalveforaorticvalvereplacement
AT zhuguangyu numericalandinvitroexperimentalassessmentoftheperformanceofanoveldesignedexpandedpolytetrafluoroethylenestentlessbileafletvalveforaorticvalvereplacement
AT nakaomasakazu numericalandinvitroexperimentalassessmentoftheperformanceofanoveldesignedexpandedpolytetrafluoroethylenestentlessbileafletvalveforaorticvalvereplacement
AT yuanqi numericalandinvitroexperimentalassessmentoftheperformanceofanoveldesignedexpandedpolytetrafluoroethylenestentlessbileafletvalveforaorticvalvereplacement
AT yeojoonhock numericalandinvitroexperimentalassessmentoftheperformanceofanoveldesignedexpandedpolytetrafluoroethylenestentlessbileafletvalveforaorticvalvereplacement