ANALISIS PERBANDINGAN HASIL PENGUKURAN DOSIS SERAP MENGGUNAKAN DETEKTOR ADVANCED MARKUS DAN FARMER IONIZATION CHAMBER PADA BERKAS ENERGI ELEKTRON DENGAN R50 � 4 g/cm2
At the IAEA Code of Practice TRS 398, the measurement of high energy electrons, there is a provision that the detector is used for R50 � 4 g/cm 2 is cylindrical detector or plan parallel detector. R50 is the depth that resulted in dose reduction by 50%. At the time of the calculation R50 for the e...
Main Authors: | , |
---|---|
Format: | Thesis |
Published: |
[Yogyakarta] : Universitas Gadjah Mada
2013
|
Subjects: |
_version_ | 1826047207609991168 |
---|---|
author | , ANISZA OKSELIA , Susila Wardaya |
author_facet | , ANISZA OKSELIA , Susila Wardaya |
author_sort | , ANISZA OKSELIA |
collection | UGM |
description | At the IAEA Code of Practice TRS 398, the measurement of high energy
electrons, there is a provision that the detector is used for R50 � 4 g/cm
2
is
cylindrical detector or plan parallel detector. R50 is the depth that resulted in dose
reduction by 50%. At the time of the calculation R50 for the electron energy which
is owned by RSUP dr. Hasan Sadikin, is energy 4 MeV, 6 MeV, 8 MeV, 10 MeV,
12 MeV and 15 MeV, it was found the energy that has R50 � 4 g/cm
2
is 10 MeV,
12 MeV and 15 MeV. The next step was measured the absorbed dose energy
electron linac by using the farmer ionization chamber as a cylindrical detector and
advanced markus detector as a plan parallel detector is alternately aligned.
The measurement results and then inserted into the calculation of the
absorbed dose maximum depth (zmax). Calculation results are then analyzed by the
authors in order not to exceed the limit of tolerance is 2%. The absorbed dose
measurements obtained using the farmer chamber and measurements with
advanced markus produced at 10 MeV energy is equal to 204 MU and 200.8 MU
with differences in the results of both detectors deviation of 1.6% and at 15 MeV
energy is equal to 206 MU and 203.4 MU with deviation difference in the results
of both detectors at 1.3%. The difference is less than 2% in this study, suggesting
that the two detectors in this study is correct and appropriate when used for
electron energy measurements with R50 � 4 g/cm
2. |
first_indexed | 2024-03-13T23:06:46Z |
format | Thesis |
id | oai:generic.eprints.org:124348 |
institution | Universiti Gadjah Mada |
last_indexed | 2024-03-13T23:06:46Z |
publishDate | 2013 |
publisher | [Yogyakarta] : Universitas Gadjah Mada |
record_format | dspace |
spelling | oai:generic.eprints.org:1243482016-03-04T08:23:16Z https://repository.ugm.ac.id/124348/ ANALISIS PERBANDINGAN HASIL PENGUKURAN DOSIS SERAP MENGGUNAKAN DETEKTOR ADVANCED MARKUS DAN FARMER IONIZATION CHAMBER PADA BERKAS ENERGI ELEKTRON DENGAN R50 � 4 g/cm2 , ANISZA OKSELIA , Susila Wardaya ETD At the IAEA Code of Practice TRS 398, the measurement of high energy electrons, there is a provision that the detector is used for R50 � 4 g/cm 2 is cylindrical detector or plan parallel detector. R50 is the depth that resulted in dose reduction by 50%. At the time of the calculation R50 for the electron energy which is owned by RSUP dr. Hasan Sadikin, is energy 4 MeV, 6 MeV, 8 MeV, 10 MeV, 12 MeV and 15 MeV, it was found the energy that has R50 � 4 g/cm 2 is 10 MeV, 12 MeV and 15 MeV. The next step was measured the absorbed dose energy electron linac by using the farmer ionization chamber as a cylindrical detector and advanced markus detector as a plan parallel detector is alternately aligned. The measurement results and then inserted into the calculation of the absorbed dose maximum depth (zmax). Calculation results are then analyzed by the authors in order not to exceed the limit of tolerance is 2%. The absorbed dose measurements obtained using the farmer chamber and measurements with advanced markus produced at 10 MeV energy is equal to 204 MU and 200.8 MU with differences in the results of both detectors deviation of 1.6% and at 15 MeV energy is equal to 206 MU and 203.4 MU with deviation difference in the results of both detectors at 1.3%. The difference is less than 2% in this study, suggesting that the two detectors in this study is correct and appropriate when used for electron energy measurements with R50 � 4 g/cm 2. [Yogyakarta] : Universitas Gadjah Mada 2013 Thesis NonPeerReviewed , ANISZA OKSELIA and , Susila Wardaya (2013) ANALISIS PERBANDINGAN HASIL PENGUKURAN DOSIS SERAP MENGGUNAKAN DETEKTOR ADVANCED MARKUS DAN FARMER IONIZATION CHAMBER PADA BERKAS ENERGI ELEKTRON DENGAN R50 � 4 g/cm2. UNSPECIFIED thesis, UNSPECIFIED. http://etd.ugm.ac.id/index.php?mod=penelitian_detail&sub=PenelitianDetail&act=view&typ=html&buku_id=64469 |
spellingShingle | ETD , ANISZA OKSELIA , Susila Wardaya ANALISIS PERBANDINGAN HASIL PENGUKURAN DOSIS SERAP MENGGUNAKAN DETEKTOR ADVANCED MARKUS DAN FARMER IONIZATION CHAMBER PADA BERKAS ENERGI ELEKTRON DENGAN R50 � 4 g/cm2 |
title | ANALISIS PERBANDINGAN HASIL PENGUKURAN DOSIS SERAP MENGGUNAKAN DETEKTOR ADVANCED MARKUS DAN FARMER IONIZATION CHAMBER PADA BERKAS ENERGI ELEKTRON DENGAN R50 � 4 g/cm2 |
title_full | ANALISIS PERBANDINGAN HASIL PENGUKURAN DOSIS SERAP MENGGUNAKAN DETEKTOR ADVANCED MARKUS DAN FARMER IONIZATION CHAMBER PADA BERKAS ENERGI ELEKTRON DENGAN R50 � 4 g/cm2 |
title_fullStr | ANALISIS PERBANDINGAN HASIL PENGUKURAN DOSIS SERAP MENGGUNAKAN DETEKTOR ADVANCED MARKUS DAN FARMER IONIZATION CHAMBER PADA BERKAS ENERGI ELEKTRON DENGAN R50 � 4 g/cm2 |
title_full_unstemmed | ANALISIS PERBANDINGAN HASIL PENGUKURAN DOSIS SERAP MENGGUNAKAN DETEKTOR ADVANCED MARKUS DAN FARMER IONIZATION CHAMBER PADA BERKAS ENERGI ELEKTRON DENGAN R50 � 4 g/cm2 |
title_short | ANALISIS PERBANDINGAN HASIL PENGUKURAN DOSIS SERAP MENGGUNAKAN DETEKTOR ADVANCED MARKUS DAN FARMER IONIZATION CHAMBER PADA BERKAS ENERGI ELEKTRON DENGAN R50 � 4 g/cm2 |
title_sort | analisis perbandingan hasil pengukuran dosis serap menggunakan detektor advanced markus dan farmer ionization chamber pada berkas energi elektron dengan r50 a�¥ 4 g cm2 |
topic | ETD |
work_keys_str_mv | AT aniszaokselia analisisperbandinganhasilpengukurandosisserapmenggunakandetektoradvancedmarkusdanfarmerionizationchamberpadaberkasenergielektrondenganr50a4gcm2 AT susilawardaya analisisperbandinganhasilpengukurandosisserapmenggunakandetektoradvancedmarkusdanfarmerionizationchamberpadaberkasenergielektrondenganr50a4gcm2 |