ANALISIS PARAMETER FISIS SERAPAN DAN DIMENSI CITRA PHANTOM HASIL TOMOGRAFI KOMPUTER MENGGUNAKAN SOFTWARE BRILLIANCE CT DAN IMAGEJ

Physical parameters analysis of absorption and dimension on Computed Tomography (CT) phantom image produced by 64-multi-slice CT at the Dr. Sardjito General Hospital Yogyakarta has been accomplished. Scanning process based on the same protocol for scanning the patients� brain, abdomen and chest. F...

Full description

Bibliographic Details
Main Authors: , EMERITA SETYOWATI, , Dr. Gede Bayu Suparta
Format: Thesis
Published: [Yogyakarta] : Universitas Gadjah Mada 2013
Subjects:
ETD
_version_ 1826047247198978048
author , EMERITA SETYOWATI
, Dr. Gede Bayu Suparta
author_facet , EMERITA SETYOWATI
, Dr. Gede Bayu Suparta
author_sort , EMERITA SETYOWATI
collection UGM
description Physical parameters analysis of absorption and dimension on Computed Tomography (CT) phantom image produced by 64-multi-slice CT at the Dr. Sardjito General Hospital Yogyakarta has been accomplished. Scanning process based on the same protocol for scanning the patients� brain, abdomen and chest. Furthermore, the reconstructed image was analyzed based on material absorption, diameter dimensions, and seven rows acrylic pin hole dimension readings of multi-pin layer using computed tomography software in Digital Imaging Communications in Medicine (DICOM) image format and ImageJ software in DICOM and Joint Photographic Experts Group (JPEG) image format. The readings on material absorption measurements using Brilliance CT software in Hounsfield Units (HU) was linear in accordance to the reference value, as well as the measurement using ImageJ software on DICOM image format. On measurement results using ImageJ, the JPEG format was not linear due to the effect of compression of 16 bits to 8 bits images. Diameter dimensions measurements using Brilliance CT software in DICOM format and ImageJ both DICOM and JPEG format image reconstruction results were linear with respect to the actual dimensions. In the seven row acrylic pin holes in multi-pin layer measurement using Brilliance CT software was linear with respect to the actual dimensions. The smallest dimension that can be measured was 2 mm. Measurements using ImageJ software for DICOM and JPEG format was linear with respect to the actual dimensions. The smallest dimension that can be measured using ImageJ software was 1 mm.
first_indexed 2024-03-13T23:07:25Z
format Thesis
id oai:generic.eprints.org:124548
institution Universiti Gadjah Mada
last_indexed 2024-03-13T23:07:25Z
publishDate 2013
publisher [Yogyakarta] : Universitas Gadjah Mada
record_format dspace
spelling oai:generic.eprints.org:1245482016-03-04T08:33:22Z https://repository.ugm.ac.id/124548/ ANALISIS PARAMETER FISIS SERAPAN DAN DIMENSI CITRA PHANTOM HASIL TOMOGRAFI KOMPUTER MENGGUNAKAN SOFTWARE BRILLIANCE CT DAN IMAGEJ , EMERITA SETYOWATI , Dr. Gede Bayu Suparta ETD Physical parameters analysis of absorption and dimension on Computed Tomography (CT) phantom image produced by 64-multi-slice CT at the Dr. Sardjito General Hospital Yogyakarta has been accomplished. Scanning process based on the same protocol for scanning the patients� brain, abdomen and chest. Furthermore, the reconstructed image was analyzed based on material absorption, diameter dimensions, and seven rows acrylic pin hole dimension readings of multi-pin layer using computed tomography software in Digital Imaging Communications in Medicine (DICOM) image format and ImageJ software in DICOM and Joint Photographic Experts Group (JPEG) image format. The readings on material absorption measurements using Brilliance CT software in Hounsfield Units (HU) was linear in accordance to the reference value, as well as the measurement using ImageJ software on DICOM image format. On measurement results using ImageJ, the JPEG format was not linear due to the effect of compression of 16 bits to 8 bits images. Diameter dimensions measurements using Brilliance CT software in DICOM format and ImageJ both DICOM and JPEG format image reconstruction results were linear with respect to the actual dimensions. In the seven row acrylic pin holes in multi-pin layer measurement using Brilliance CT software was linear with respect to the actual dimensions. The smallest dimension that can be measured was 2 mm. Measurements using ImageJ software for DICOM and JPEG format was linear with respect to the actual dimensions. The smallest dimension that can be measured using ImageJ software was 1 mm. [Yogyakarta] : Universitas Gadjah Mada 2013 Thesis NonPeerReviewed , EMERITA SETYOWATI and , Dr. Gede Bayu Suparta (2013) ANALISIS PARAMETER FISIS SERAPAN DAN DIMENSI CITRA PHANTOM HASIL TOMOGRAFI KOMPUTER MENGGUNAKAN SOFTWARE BRILLIANCE CT DAN IMAGEJ. UNSPECIFIED thesis, UNSPECIFIED. http://etd.ugm.ac.id/index.php?mod=penelitian_detail&sub=PenelitianDetail&act=view&typ=html&buku_id=64701
spellingShingle ETD
, EMERITA SETYOWATI
, Dr. Gede Bayu Suparta
ANALISIS PARAMETER FISIS SERAPAN DAN DIMENSI CITRA PHANTOM HASIL TOMOGRAFI KOMPUTER MENGGUNAKAN SOFTWARE BRILLIANCE CT DAN IMAGEJ
title ANALISIS PARAMETER FISIS SERAPAN DAN DIMENSI CITRA PHANTOM HASIL TOMOGRAFI KOMPUTER MENGGUNAKAN SOFTWARE BRILLIANCE CT DAN IMAGEJ
title_full ANALISIS PARAMETER FISIS SERAPAN DAN DIMENSI CITRA PHANTOM HASIL TOMOGRAFI KOMPUTER MENGGUNAKAN SOFTWARE BRILLIANCE CT DAN IMAGEJ
title_fullStr ANALISIS PARAMETER FISIS SERAPAN DAN DIMENSI CITRA PHANTOM HASIL TOMOGRAFI KOMPUTER MENGGUNAKAN SOFTWARE BRILLIANCE CT DAN IMAGEJ
title_full_unstemmed ANALISIS PARAMETER FISIS SERAPAN DAN DIMENSI CITRA PHANTOM HASIL TOMOGRAFI KOMPUTER MENGGUNAKAN SOFTWARE BRILLIANCE CT DAN IMAGEJ
title_short ANALISIS PARAMETER FISIS SERAPAN DAN DIMENSI CITRA PHANTOM HASIL TOMOGRAFI KOMPUTER MENGGUNAKAN SOFTWARE BRILLIANCE CT DAN IMAGEJ
title_sort analisis parameter fisis serapan dan dimensi citra phantom hasil tomografi komputer menggunakan software brilliance ct dan imagej
topic ETD
work_keys_str_mv AT emeritasetyowati analisisparameterfisisserapandandimensicitraphantomhasiltomografikomputermenggunakansoftwarebrilliancectdanimagej
AT drgedebayusuparta analisisparameterfisisserapandandimensicitraphantomhasiltomografikomputermenggunakansoftwarebrilliancectdanimagej