Experimental validation of a finite element model of a human cadaveric tibia.

Finite element (FE) models of long bones are widely used to analyze implant designs. Experimental validation has been used to examine the accuracy of FE models of cadaveric femurs; however, although convergence tests have been carried out, no FE models of an intact and implanted human cadaveric tibi...

Full description

Bibliographic Details
Main Authors: Gray, H, Taddei, F, Zavatsky, A, Cristofolini, L, Gill, H
Format: Journal article
Language:English
Published: 2008
_version_ 1826270073755533312
author Gray, H
Taddei, F
Zavatsky, A
Cristofolini, L
Gill, H
author_facet Gray, H
Taddei, F
Zavatsky, A
Cristofolini, L
Gill, H
author_sort Gray, H
collection OXFORD
description Finite element (FE) models of long bones are widely used to analyze implant designs. Experimental validation has been used to examine the accuracy of FE models of cadaveric femurs; however, although convergence tests have been carried out, no FE models of an intact and implanted human cadaveric tibia have been validated using a range of experimental loading conditions. The aim of the current study was to create FE models of a human cadaveric tibia, both intact and implanted with a unicompartmental knee replacement, and to validate the models against results obtained from a comprehensive set of experiments. Seventeen strain rosettes were attached to a human cadaveric tibia. Surface strains and displacements were measured under 17 loading conditions, which consisted of axial, torsional, and bending loads. The tibia was tested both before and after implantation of the knee replacement. FE models were created based on computed tomography (CT) scans of the cadaveric tibia. The models consisted of ten-node tetrahedral elements and used 600 material properties derived from the CT scans. The experiments were simulated on the models and the results compared to experimental results. Experimental strain measurements were highly repeatable and the measured stiffnesses compared well to published results. For the intact tibia under axial loading, the regression line through a plot of strains predicted by the FE model versus experimentally measured strains had a slope of 1.15, an intercept of 5.5 microstrain, and an R(2) value of 0.98. For the implanted tibia, the comparable regression line had a slope of 1.25, an intercept of 12.3 microstrain, and an R(2) value of 0.97. The root mean square errors were 6.0% and 8.8% for the intact and implanted models under axial loads, respectively. The model produced by the current study provides a tool for simulating mechanical test conditions on a human tibia. This has considerable value in reducing the costs of physical testing by pre-selecting the most appropriate test conditions or most favorable prosthetic designs for final mechanical testing. It can also be used to gain insight into the results of physical testing, by allowing the prediction of those variables difficult or impossible to measure directly.
first_indexed 2024-03-06T21:35:09Z
format Journal article
id oxford-uuid:45fbc08d-1300-4eb9-af9d-77cfd31d1375
institution University of Oxford
language English
last_indexed 2024-03-06T21:35:09Z
publishDate 2008
record_format dspace
spelling oxford-uuid:45fbc08d-1300-4eb9-af9d-77cfd31d13752022-03-26T15:11:06ZExperimental validation of a finite element model of a human cadaveric tibia.Journal articlehttp://purl.org/coar/resource_type/c_dcae04bcuuid:45fbc08d-1300-4eb9-af9d-77cfd31d1375EnglishSymplectic Elements at Oxford2008Gray, HTaddei, FZavatsky, ACristofolini, LGill, HFinite element (FE) models of long bones are widely used to analyze implant designs. Experimental validation has been used to examine the accuracy of FE models of cadaveric femurs; however, although convergence tests have been carried out, no FE models of an intact and implanted human cadaveric tibia have been validated using a range of experimental loading conditions. The aim of the current study was to create FE models of a human cadaveric tibia, both intact and implanted with a unicompartmental knee replacement, and to validate the models against results obtained from a comprehensive set of experiments. Seventeen strain rosettes were attached to a human cadaveric tibia. Surface strains and displacements were measured under 17 loading conditions, which consisted of axial, torsional, and bending loads. The tibia was tested both before and after implantation of the knee replacement. FE models were created based on computed tomography (CT) scans of the cadaveric tibia. The models consisted of ten-node tetrahedral elements and used 600 material properties derived from the CT scans. The experiments were simulated on the models and the results compared to experimental results. Experimental strain measurements were highly repeatable and the measured stiffnesses compared well to published results. For the intact tibia under axial loading, the regression line through a plot of strains predicted by the FE model versus experimentally measured strains had a slope of 1.15, an intercept of 5.5 microstrain, and an R(2) value of 0.98. For the implanted tibia, the comparable regression line had a slope of 1.25, an intercept of 12.3 microstrain, and an R(2) value of 0.97. The root mean square errors were 6.0% and 8.8% for the intact and implanted models under axial loads, respectively. The model produced by the current study provides a tool for simulating mechanical test conditions on a human tibia. This has considerable value in reducing the costs of physical testing by pre-selecting the most appropriate test conditions or most favorable prosthetic designs for final mechanical testing. It can also be used to gain insight into the results of physical testing, by allowing the prediction of those variables difficult or impossible to measure directly.
spellingShingle Gray, H
Taddei, F
Zavatsky, A
Cristofolini, L
Gill, H
Experimental validation of a finite element model of a human cadaveric tibia.
title Experimental validation of a finite element model of a human cadaveric tibia.
title_full Experimental validation of a finite element model of a human cadaveric tibia.
title_fullStr Experimental validation of a finite element model of a human cadaveric tibia.
title_full_unstemmed Experimental validation of a finite element model of a human cadaveric tibia.
title_short Experimental validation of a finite element model of a human cadaveric tibia.
title_sort experimental validation of a finite element model of a human cadaveric tibia
work_keys_str_mv AT grayh experimentalvalidationofafiniteelementmodelofahumancadaverictibia
AT taddeif experimentalvalidationofafiniteelementmodelofahumancadaverictibia
AT zavatskya experimentalvalidationofafiniteelementmodelofahumancadaverictibia
AT cristofolinil experimentalvalidationofafiniteelementmodelofahumancadaverictibia
AT gillh experimentalvalidationofafiniteelementmodelofahumancadaverictibia