Problems with capnography at high altitude.

The aim of this study was to investigate why capnographs malfunction at high altitude, in order to suggest ways of improving instrument reliability. Four machines were decompressed to simulated high altitude and measurements taken. The study showed that the reduced air density at altitude has effect...

Celý popis

Podrobná bibliografie
Hlavní autoři: Pattinson, K, Myers, S, Gardner-Thorpe, C
Médium: Journal article
Jazyk:English
Vydáno: 2004
_version_ 1826281810636570624
author Pattinson, K
Myers, S
Gardner-Thorpe, C
author_facet Pattinson, K
Myers, S
Gardner-Thorpe, C
author_sort Pattinson, K
collection OXFORD
description The aim of this study was to investigate why capnographs malfunction at high altitude, in order to suggest ways of improving instrument reliability. Four machines were decompressed to simulated high altitude and measurements taken. The study showed that the reduced air density at altitude has effects upon the pump, causing difficulty in entraining gas, and that high altitude has additional effects upon calibration of the instruments.
first_indexed 2024-03-07T00:34:24Z
format Journal article
id oxford-uuid:80e48e79-828a-4f8b-93c4-2d2f1c6c8517
institution University of Oxford
language English
last_indexed 2024-03-07T00:34:24Z
publishDate 2004
record_format dspace
spelling oxford-uuid:80e48e79-828a-4f8b-93c4-2d2f1c6c85172022-03-26T21:26:30ZProblems with capnography at high altitude.Journal articlehttp://purl.org/coar/resource_type/c_dcae04bcuuid:80e48e79-828a-4f8b-93c4-2d2f1c6c8517EnglishSymplectic Elements at Oxford2004Pattinson, KMyers, SGardner-Thorpe, CThe aim of this study was to investigate why capnographs malfunction at high altitude, in order to suggest ways of improving instrument reliability. Four machines were decompressed to simulated high altitude and measurements taken. The study showed that the reduced air density at altitude has effects upon the pump, causing difficulty in entraining gas, and that high altitude has additional effects upon calibration of the instruments.
spellingShingle Pattinson, K
Myers, S
Gardner-Thorpe, C
Problems with capnography at high altitude.
title Problems with capnography at high altitude.
title_full Problems with capnography at high altitude.
title_fullStr Problems with capnography at high altitude.
title_full_unstemmed Problems with capnography at high altitude.
title_short Problems with capnography at high altitude.
title_sort problems with capnography at high altitude
work_keys_str_mv AT pattinsonk problemswithcapnographyathighaltitude
AT myerss problemswithcapnographyathighaltitude
AT gardnerthorpec problemswithcapnographyathighaltitude