Sono-cathodic stripping voltammetry of manganese at a polished boron-doped diamond electrode: application to the determination of manganese in instant tea.

Ultrasonically assisted cathodic stripping voltammetry at a boron-doped diamond electrode was developed for the detection of manganese. Differential-pulse voltammetry was used to give the analytical signal from a cathodic strip of electrodeposited MnO2; linearity was observed from 10(-11) M to at le...

Full description

Bibliographic Details
Main Authors: Saterlay, A, Foord, J, Compton, R
Format: Journal article
Language:English
Published: 1999
Description
Summary:Ultrasonically assisted cathodic stripping voltammetry at a boron-doped diamond electrode was developed for the detection of manganese. Differential-pulse voltammetry was used to give the analytical signal from a cathodic strip of electrodeposited MnO2; linearity was observed from 10(-11) M to at least 3 x 10(-7) M, with 10(-11) M being the detection limit for a 2 min deposition. The procedure involves both ultrasonic-anodic deposition of MnO2 and ultrasonic-cathodic stripping. This novel analytical tool is robust, reproducible, mercury free, oxygen insensitive and highly specific towards manganese. The differential-pulse sono-cathodic stripping voltammetric technique was used to determine successfully the manganese content of two instant tea samples, giving excellent agreement with independent AAS analyses.