Monte-Carlo modelling to determine optimum filter choices for sub-microsecond optical pyrometry

When designing a spectral-band pyrometer for use at high time resolutions (sub-μs), there is ambiguity regarding the optimum characteristics for a spectral filter(s). In particular, while prior work has discussed uncertainties in spectral-band pyrometry, there has been little discussion of the effec...

Full description

Bibliographic Details
Main Authors: Ota, TA, Chapman, DJ, Eakins, DE
Format: Journal article
Language:English
Published: American Institute of Physics 2017
_version_ 1826291588672782336
author Ota, TA
Chapman, DJ
Eakins, DE
author_facet Ota, TA
Chapman, DJ
Eakins, DE
author_sort Ota, TA
collection OXFORD
description When designing a spectral-band pyrometer for use at high time resolutions (sub-μs), there is ambiguity regarding the optimum characteristics for a spectral filter(s). In particular, while prior work has discussed uncertainties in spectral-band pyrometry, there has been little discussion of the effects of noise which is an important consideration in time-resolved, high speed experiments. Using a Monte-Carlo process to simulate the effects of noise, a model of collection from a black body has been developed to give insights into the optimum choices for centre wavelength and passband width. The model was validated and then used to explore the effects of centre wavelength and passband width on measurement uncertainty. This reveals a transition centre wavelength below which uncertainties in calculated temperature are high. To further investigate system performance, simultaneous variation of the centre wavelength and bandpass width of a filter is investigated. Using data reduction, the effects of temperature and noise levels are illustrated and an empirical approximation is determined. The results presented show that filter choice can significantly affect instrument performance and, while best practice requires detailed modelling to achieve optimal performance, the expression presented can be used to aid filter selection.
first_indexed 2024-03-07T03:01:40Z
format Journal article
id oxford-uuid:b127ead5-162b-414e-b26e-fa63c793ab53
institution University of Oxford
language English
last_indexed 2024-03-07T03:01:40Z
publishDate 2017
publisher American Institute of Physics
record_format dspace
spelling oxford-uuid:b127ead5-162b-414e-b26e-fa63c793ab532022-03-27T04:01:55ZMonte-Carlo modelling to determine optimum filter choices for sub-microsecond optical pyrometryJournal articlehttp://purl.org/coar/resource_type/c_dcae04bcuuid:b127ead5-162b-414e-b26e-fa63c793ab53EnglishSymplectic Elements at OxfordAmerican Institute of Physics2017Ota, TAChapman, DJEakins, DEWhen designing a spectral-band pyrometer for use at high time resolutions (sub-μs), there is ambiguity regarding the optimum characteristics for a spectral filter(s). In particular, while prior work has discussed uncertainties in spectral-band pyrometry, there has been little discussion of the effects of noise which is an important consideration in time-resolved, high speed experiments. Using a Monte-Carlo process to simulate the effects of noise, a model of collection from a black body has been developed to give insights into the optimum choices for centre wavelength and passband width. The model was validated and then used to explore the effects of centre wavelength and passband width on measurement uncertainty. This reveals a transition centre wavelength below which uncertainties in calculated temperature are high. To further investigate system performance, simultaneous variation of the centre wavelength and bandpass width of a filter is investigated. Using data reduction, the effects of temperature and noise levels are illustrated and an empirical approximation is determined. The results presented show that filter choice can significantly affect instrument performance and, while best practice requires detailed modelling to achieve optimal performance, the expression presented can be used to aid filter selection.
spellingShingle Ota, TA
Chapman, DJ
Eakins, DE
Monte-Carlo modelling to determine optimum filter choices for sub-microsecond optical pyrometry
title Monte-Carlo modelling to determine optimum filter choices for sub-microsecond optical pyrometry
title_full Monte-Carlo modelling to determine optimum filter choices for sub-microsecond optical pyrometry
title_fullStr Monte-Carlo modelling to determine optimum filter choices for sub-microsecond optical pyrometry
title_full_unstemmed Monte-Carlo modelling to determine optimum filter choices for sub-microsecond optical pyrometry
title_short Monte-Carlo modelling to determine optimum filter choices for sub-microsecond optical pyrometry
title_sort monte carlo modelling to determine optimum filter choices for sub microsecond optical pyrometry
work_keys_str_mv AT otata montecarlomodellingtodetermineoptimumfilterchoicesforsubmicrosecondopticalpyrometry
AT chapmandj montecarlomodellingtodetermineoptimumfilterchoicesforsubmicrosecondopticalpyrometry
AT eakinsde montecarlomodellingtodetermineoptimumfilterchoicesforsubmicrosecondopticalpyrometry