Highly trabeculated structure of the human endocardium underlies asymmetrical response to low-energy monophasic shocks

Novel low-energy defibrillation therapies are thought to be driven by virtual-electrodes (VEs), due to the interaction of applied monophasic electric shocks with fine-scale anatomical structures within the heart. Significant inter-species differences in the cardiac (micro)-anatomy exist, however, pa...

Full description

Bibliographic Details
Main Authors: Connolly, A, Robson, M, Schneider, J, Burton, R, Plank, G, Bishop, M
Format: Journal article
Published: AIP Publishing 2017
_version_ 1797091878035783680
author Connolly, A
Robson, M
Schneider, J
Burton, R
Plank, G
Bishop, M
author_facet Connolly, A
Robson, M
Schneider, J
Burton, R
Plank, G
Bishop, M
author_sort Connolly, A
collection OXFORD
description Novel low-energy defibrillation therapies are thought to be driven by virtual-electrodes (VEs), due to the interaction of applied monophasic electric shocks with fine-scale anatomical structures within the heart. Significant inter-species differences in the cardiac (micro)-anatomy exist, however, particularly with respect to the degree of endocardial trabeculations, which may underlie important differences in response to low-energy defibrillation protocols. Understanding the interaction of monophasic electric fields with the specific human micro-anatomy is therefore imperative in facilitating the translation and optimisation of these promising experimental therapies to the clinic. In this study, we sought to investigate how electric fields from implanted devices interact with the highly trabeculated human endocardial surface to better understand shock success in order to help optimise future clinical protocols. A bi-ventricular human computational model was constructed from high resolution (350 μm) ex-vivo MR data, including anatomically accurate endocardial structures. Monophasic shocks were applied between a basal right ventricular catheter and an exterior ground. Shocks of varying strengths were applied with both anodal [positive right ventricle (RV) electrode] and cathodal (negative RV electrode) polarities at different states of tissue refractoriness and during induced arrhythmias. Anodal shocks induced isolated positive VEs at the distal side of "detached" trabeculations, which rapidly spread into hyperpolarised tissue on the surrounding endocardial surfaces following the shock. Anodal shocks thus depolarised more tissue 10 ms after the shock than cathodal shocks where the propagation of activation from VEs induced on the proximal side of "detached" trabeculations was prevented due to refractory endocardium. Anodal shocks increased arrhythmia complexity more than cathodal shocks during failed anti-arrhythmia shocks. In conclusion, multiple detached trabeculations in the human ventricle interact with anodal stimuli to induce multiple secondary sources from VEs, facilitating more rapid shock-induced ventricular excitation compared to cathodal shocks. Such a mechanism may help explain inter-species differences in response to shocks and help to develop novel defibrillation strategies.
first_indexed 2024-03-07T03:38:38Z
format Journal article
id oxford-uuid:bd260f5d-2c99-401c-aad9-4b01f78c1e88
institution University of Oxford
last_indexed 2024-03-07T03:38:38Z
publishDate 2017
publisher AIP Publishing
record_format dspace
spelling oxford-uuid:bd260f5d-2c99-401c-aad9-4b01f78c1e882022-03-27T05:29:31ZHighly trabeculated structure of the human endocardium underlies asymmetrical response to low-energy monophasic shocksJournal articlehttp://purl.org/coar/resource_type/c_dcae04bcuuid:bd260f5d-2c99-401c-aad9-4b01f78c1e88Symplectic Elements at OxfordAIP Publishing2017Connolly, ARobson, MSchneider, JBurton, RPlank, GBishop, MNovel low-energy defibrillation therapies are thought to be driven by virtual-electrodes (VEs), due to the interaction of applied monophasic electric shocks with fine-scale anatomical structures within the heart. Significant inter-species differences in the cardiac (micro)-anatomy exist, however, particularly with respect to the degree of endocardial trabeculations, which may underlie important differences in response to low-energy defibrillation protocols. Understanding the interaction of monophasic electric fields with the specific human micro-anatomy is therefore imperative in facilitating the translation and optimisation of these promising experimental therapies to the clinic. In this study, we sought to investigate how electric fields from implanted devices interact with the highly trabeculated human endocardial surface to better understand shock success in order to help optimise future clinical protocols. A bi-ventricular human computational model was constructed from high resolution (350 μm) ex-vivo MR data, including anatomically accurate endocardial structures. Monophasic shocks were applied between a basal right ventricular catheter and an exterior ground. Shocks of varying strengths were applied with both anodal [positive right ventricle (RV) electrode] and cathodal (negative RV electrode) polarities at different states of tissue refractoriness and during induced arrhythmias. Anodal shocks induced isolated positive VEs at the distal side of "detached" trabeculations, which rapidly spread into hyperpolarised tissue on the surrounding endocardial surfaces following the shock. Anodal shocks thus depolarised more tissue 10 ms after the shock than cathodal shocks where the propagation of activation from VEs induced on the proximal side of "detached" trabeculations was prevented due to refractory endocardium. Anodal shocks increased arrhythmia complexity more than cathodal shocks during failed anti-arrhythmia shocks. In conclusion, multiple detached trabeculations in the human ventricle interact with anodal stimuli to induce multiple secondary sources from VEs, facilitating more rapid shock-induced ventricular excitation compared to cathodal shocks. Such a mechanism may help explain inter-species differences in response to shocks and help to develop novel defibrillation strategies.
spellingShingle Connolly, A
Robson, M
Schneider, J
Burton, R
Plank, G
Bishop, M
Highly trabeculated structure of the human endocardium underlies asymmetrical response to low-energy monophasic shocks
title Highly trabeculated structure of the human endocardium underlies asymmetrical response to low-energy monophasic shocks
title_full Highly trabeculated structure of the human endocardium underlies asymmetrical response to low-energy monophasic shocks
title_fullStr Highly trabeculated structure of the human endocardium underlies asymmetrical response to low-energy monophasic shocks
title_full_unstemmed Highly trabeculated structure of the human endocardium underlies asymmetrical response to low-energy monophasic shocks
title_short Highly trabeculated structure of the human endocardium underlies asymmetrical response to low-energy monophasic shocks
title_sort highly trabeculated structure of the human endocardium underlies asymmetrical response to low energy monophasic shocks
work_keys_str_mv AT connollya highlytrabeculatedstructureofthehumanendocardiumunderliesasymmetricalresponsetolowenergymonophasicshocks
AT robsonm highlytrabeculatedstructureofthehumanendocardiumunderliesasymmetricalresponsetolowenergymonophasicshocks
AT schneiderj highlytrabeculatedstructureofthehumanendocardiumunderliesasymmetricalresponsetolowenergymonophasicshocks
AT burtonr highlytrabeculatedstructureofthehumanendocardiumunderliesasymmetricalresponsetolowenergymonophasicshocks
AT plankg highlytrabeculatedstructureofthehumanendocardiumunderliesasymmetricalresponsetolowenergymonophasicshocks
AT bishopm highlytrabeculatedstructureofthehumanendocardiumunderliesasymmetricalresponsetolowenergymonophasicshocks