High-resolution ion microscope imaging over wide mass ranges using electrodynamic post-extraction differential acceleration

A time-dependent postextraction differential acceleration (PEDA) potential was used to temporally focus increasingly heavy ions in a stigmatic imaging mass spectrometer, allowing them to be imaged with high mass and spatial resolutions over a broad mass-to-charge (m/z) range. By applying a linearly...

Full description

Bibliographic Details
Main Authors: Guo, A, Burleigh, RJ, Smith, N, Brouard, M, Burt, M
Format: Journal article
Language:English
Published: American Chemical Society 2020
Description
Summary:A time-dependent postextraction differential acceleration (PEDA) potential was used to temporally focus increasingly heavy ions in a stigmatic imaging mass spectrometer, allowing them to be imaged with high mass and spatial resolutions over a broad mass-to-charge (m/z) range. By applying a linearly rising potential to the ion extraction electrode, sequential m/z ratios were subjected to a changing electric field, allowing their foci to coincide at the detector. Using this approach, at least 75% of the maximum mass resolution was obtained over a 300–600 Da range when the ion microscope was focused around 450 Da, representing more than a 10-fold increase over the conventional single-field PEDA method.