Monte Carlo modeling of small photon fields: quantifying the impact of focal spot size on source occlusion and output factors, and exploring miniphantom design for small-field measurements.

The accuracy with which Monte Carlo models of photon beams generated by linear accelerators (linacs) can describe small-field dose distributions depends on the modeled width of the electron beam profile incident on the linac target. It is known that the electron focal spot width affects penumbra and...

Full description

Bibliographic Details
Main Authors: Scott, A, Nahum, A, Fenwick, J
Format: Journal article
Language:English
Published: 2009
_version_ 1797099005201612800
author Scott, A
Nahum, A
Fenwick, J
author_facet Scott, A
Nahum, A
Fenwick, J
author_sort Scott, A
collection OXFORD
description The accuracy with which Monte Carlo models of photon beams generated by linear accelerators (linacs) can describe small-field dose distributions depends on the modeled width of the electron beam profile incident on the linac target. It is known that the electron focal spot width affects penumbra and cross-field profiles; here, the authors explore the extent to which source occlusion reduces linac output for smaller fields and larger spot sizes. A BEAMnrc Monte Carlo linac model has been used to investigate the variation in penumbra widths and small-field output factors with electron spot size. A formalism is developed separating head scatter factors into source occlusion and flattening filter factors. Differences between head scatter factors defined in terms of in-air energy fluence, collision kerma, and terma are explored using Monte Carlo calculations. Estimates of changes in kerma-based source occlusion and flattening filter factors with field size and focal spot width are obtained by calculating doses deposited in a narrow 2 mm wide virtual "milliphantom" geometry. The impact of focal spot size on phantom scatter is also explored. Modeled electron spot sizes of 0.4-0.7 mm FWHM generate acceptable matches to measured penumbra widths. However the 0.5 cm field output factor is quite sensitive to electron spot width, the measured output only being matched by calculations for a 0.7 mm spot width. Because the spectra of the unscattered primary (psi(pi)) and head-scattered (psi(sigma)) photon energy fluences differ, miniphantom-based collision kerma measurements do not scale precisely with total in-air energy fluence psi = (psi(pi) + psi(sigma) but with (psi(pi)+ 1.2psi(sigma)). For most field sizes, on-axis collision kerma is independent of the focal spot size; but for a 0.5 cm field size and 1.0 mm spot width, it is reduced by around 7% mostly due to source occlusion. The phantom scatter factor of the 0.5 cm field also shows some spot size dependence, decreasing by 6% (relative) as spot size is increased from 0.1 to 1.0 mm. The dependence of small-field source occlusion and output factors on the focal spot size makes this a significant factor in Monte Carlo modeling of small (< 1 cm) fields. Changes in penumbra width with spot size are not sufficiently large to accurately pinpoint spot widths. Consequently, while Monte Carlo models based exclusively on large-field data can quite accurately predict small-field profiles and PDDs, in the absence of experimental methods of determining incident electron beam profiles it will remain necessary to measure small-field output factors, fine-tuning modeled spot sizes to ensure good matching between the Monte Carlo and the measured output factors.
first_indexed 2024-03-07T05:17:35Z
format Journal article
id oxford-uuid:ddbdbd40-ec93-47aa-95e0-376ed7335aaf
institution University of Oxford
language English
last_indexed 2024-03-07T05:17:35Z
publishDate 2009
record_format dspace
spelling oxford-uuid:ddbdbd40-ec93-47aa-95e0-376ed7335aaf2022-03-27T09:27:16ZMonte Carlo modeling of small photon fields: quantifying the impact of focal spot size on source occlusion and output factors, and exploring miniphantom design for small-field measurements.Journal articlehttp://purl.org/coar/resource_type/c_dcae04bcuuid:ddbdbd40-ec93-47aa-95e0-376ed7335aafEnglishSymplectic Elements at Oxford2009Scott, ANahum, AFenwick, JThe accuracy with which Monte Carlo models of photon beams generated by linear accelerators (linacs) can describe small-field dose distributions depends on the modeled width of the electron beam profile incident on the linac target. It is known that the electron focal spot width affects penumbra and cross-field profiles; here, the authors explore the extent to which source occlusion reduces linac output for smaller fields and larger spot sizes. A BEAMnrc Monte Carlo linac model has been used to investigate the variation in penumbra widths and small-field output factors with electron spot size. A formalism is developed separating head scatter factors into source occlusion and flattening filter factors. Differences between head scatter factors defined in terms of in-air energy fluence, collision kerma, and terma are explored using Monte Carlo calculations. Estimates of changes in kerma-based source occlusion and flattening filter factors with field size and focal spot width are obtained by calculating doses deposited in a narrow 2 mm wide virtual "milliphantom" geometry. The impact of focal spot size on phantom scatter is also explored. Modeled electron spot sizes of 0.4-0.7 mm FWHM generate acceptable matches to measured penumbra widths. However the 0.5 cm field output factor is quite sensitive to electron spot width, the measured output only being matched by calculations for a 0.7 mm spot width. Because the spectra of the unscattered primary (psi(pi)) and head-scattered (psi(sigma)) photon energy fluences differ, miniphantom-based collision kerma measurements do not scale precisely with total in-air energy fluence psi = (psi(pi) + psi(sigma) but with (psi(pi)+ 1.2psi(sigma)). For most field sizes, on-axis collision kerma is independent of the focal spot size; but for a 0.5 cm field size and 1.0 mm spot width, it is reduced by around 7% mostly due to source occlusion. The phantom scatter factor of the 0.5 cm field also shows some spot size dependence, decreasing by 6% (relative) as spot size is increased from 0.1 to 1.0 mm. The dependence of small-field source occlusion and output factors on the focal spot size makes this a significant factor in Monte Carlo modeling of small (< 1 cm) fields. Changes in penumbra width with spot size are not sufficiently large to accurately pinpoint spot widths. Consequently, while Monte Carlo models based exclusively on large-field data can quite accurately predict small-field profiles and PDDs, in the absence of experimental methods of determining incident electron beam profiles it will remain necessary to measure small-field output factors, fine-tuning modeled spot sizes to ensure good matching between the Monte Carlo and the measured output factors.
spellingShingle Scott, A
Nahum, A
Fenwick, J
Monte Carlo modeling of small photon fields: quantifying the impact of focal spot size on source occlusion and output factors, and exploring miniphantom design for small-field measurements.
title Monte Carlo modeling of small photon fields: quantifying the impact of focal spot size on source occlusion and output factors, and exploring miniphantom design for small-field measurements.
title_full Monte Carlo modeling of small photon fields: quantifying the impact of focal spot size on source occlusion and output factors, and exploring miniphantom design for small-field measurements.
title_fullStr Monte Carlo modeling of small photon fields: quantifying the impact of focal spot size on source occlusion and output factors, and exploring miniphantom design for small-field measurements.
title_full_unstemmed Monte Carlo modeling of small photon fields: quantifying the impact of focal spot size on source occlusion and output factors, and exploring miniphantom design for small-field measurements.
title_short Monte Carlo modeling of small photon fields: quantifying the impact of focal spot size on source occlusion and output factors, and exploring miniphantom design for small-field measurements.
title_sort monte carlo modeling of small photon fields quantifying the impact of focal spot size on source occlusion and output factors and exploring miniphantom design for small field measurements
work_keys_str_mv AT scotta montecarlomodelingofsmallphotonfieldsquantifyingtheimpactoffocalspotsizeonsourceocclusionandoutputfactorsandexploringminiphantomdesignforsmallfieldmeasurements
AT nahuma montecarlomodelingofsmallphotonfieldsquantifyingtheimpactoffocalspotsizeonsourceocclusionandoutputfactorsandexploringminiphantomdesignforsmallfieldmeasurements
AT fenwickj montecarlomodelingofsmallphotonfieldsquantifyingtheimpactoffocalspotsizeonsourceocclusionandoutputfactorsandexploringminiphantomdesignforsmallfieldmeasurements