The use of wearable pulse oximeters in the prompt detection of hypoxemia and during movement: diagnostic accuracy study

<p><strong>Background:</strong> Commercially available wearable (ambulatory) pulse oximeters have been recommended as a method for managing patients at risk of physiological deterioration, such as active patients with COVID-19 disease receiving care in hospital isolation rooms; how...

Full description

Bibliographic Details
Main Authors: Santos, M, Vollam, S, Pimentel, M, Morgado Areia, C, Young, L, Roman, C, Ede, J, Piper, P, King, E, Harford, M, Shah, A, Gustafson, O, Tarassenko, L, Watkinson, P
Format: Journal article
Language:English
Published: JMIR Publications 2022
_version_ 1826309738289168384
author Santos, M
Vollam, S
Pimentel, M
Morgado Areia, C
Young, L
Roman, C
Ede, J
Piper, P
King, E
Harford, M
Shah, A
Gustafson, O
Tarassenko, L
Watkinson, P
author_facet Santos, M
Vollam, S
Pimentel, M
Morgado Areia, C
Young, L
Roman, C
Ede, J
Piper, P
King, E
Harford, M
Shah, A
Gustafson, O
Tarassenko, L
Watkinson, P
author_sort Santos, M
collection OXFORD
description <p><strong>Background:</strong> Commercially available wearable (ambulatory) pulse oximeters have been recommended as a method for managing patients at risk of physiological deterioration, such as active patients with COVID-19 disease receiving care in hospital isolation rooms; however, their reliability in usual hospital settings is not known.</p> <p><strong>Objective:</strong> We report the performance of wearable pulse oximeters in a simulated clinical setting when challenged by motion and low levels of arterial blood oxygen saturation (SaO<sub>2</sub>).</p> <p><strong>Methods:</strong> The performance of 1 wrist-worn (Wavelet) and 3 finger-worn (CheckMe O2+, AP-20, and WristOx2 3150) wearable, wireless transmission–mode pulse oximeters was evaluated. For this, 7 motion tasks were performed: at rest, sit-to-stand, tapping, rubbing, drinking, turning pages, and using a tablet. Hypoxia exposure followed, in which inspired gases were adjusted to achieve decreasing SaO<sub>2</sub> levels at 100%, 95%, 90%, 87%, 85%, 83%, and 80%. Peripheral oxygen saturation (SpO<sub>2</sub>) estimates were compared with simultaneous SaO<sub>2</sub> samples to calculate the root-mean-square error (RMSE). The area under the receiver operating characteristic curve was used to analyze the detection of hypoxemia (ie, SaO<sub>2</sub><90%).</p> <p><strong>Results:</strong> SpO<sub>2</sub> estimates matching 215 SaO<sub>2</sub> samples in both study phases, from 33 participants, were analyzed. Tapping, rubbing, turning pages, and using a tablet degraded SpO<sub>2</sub> estimation (RMSE>4% for at least 1 device). All finger-worn pulse oximeters detected hypoxemia, with an overall sensitivity of ≥0.87 and specificity of ≥0.80, comparable to that of the Philips MX450 pulse oximeter.</p> <p><strong>Conclusions:</strong> The SpO<sub>2</sub> accuracy of wearable finger-worn pulse oximeters was within that required by the International Organization for Standardization guidelines. Performance was degraded by motion, but all pulse oximeters could detect hypoxemia. Our findings support the use of wearable, wireless transmission–mode pulse oximeters to detect the onset of clinical deterioration in hospital settings.</p>
first_indexed 2024-03-07T07:40:11Z
format Journal article
id oxford-uuid:ee619c02-6662-437b-b059-a9b2e32b3cde
institution University of Oxford
language English
last_indexed 2024-03-07T07:40:11Z
publishDate 2022
publisher JMIR Publications
record_format dspace
spelling oxford-uuid:ee619c02-6662-437b-b059-a9b2e32b3cde2023-04-06T10:38:07ZThe use of wearable pulse oximeters in the prompt detection of hypoxemia and during movement: diagnostic accuracy studyJournal articlehttp://purl.org/coar/resource_type/c_dcae04bcuuid:ee619c02-6662-437b-b059-a9b2e32b3cdeEnglishSymplectic ElementsJMIR Publications2022Santos, MVollam, SPimentel, MMorgado Areia, CYoung, LRoman, CEde, JPiper, PKing, EHarford, MShah, AGustafson, OTarassenko, LWatkinson, P<p><strong>Background:</strong> Commercially available wearable (ambulatory) pulse oximeters have been recommended as a method for managing patients at risk of physiological deterioration, such as active patients with COVID-19 disease receiving care in hospital isolation rooms; however, their reliability in usual hospital settings is not known.</p> <p><strong>Objective:</strong> We report the performance of wearable pulse oximeters in a simulated clinical setting when challenged by motion and low levels of arterial blood oxygen saturation (SaO<sub>2</sub>).</p> <p><strong>Methods:</strong> The performance of 1 wrist-worn (Wavelet) and 3 finger-worn (CheckMe O2+, AP-20, and WristOx2 3150) wearable, wireless transmission–mode pulse oximeters was evaluated. For this, 7 motion tasks were performed: at rest, sit-to-stand, tapping, rubbing, drinking, turning pages, and using a tablet. Hypoxia exposure followed, in which inspired gases were adjusted to achieve decreasing SaO<sub>2</sub> levels at 100%, 95%, 90%, 87%, 85%, 83%, and 80%. Peripheral oxygen saturation (SpO<sub>2</sub>) estimates were compared with simultaneous SaO<sub>2</sub> samples to calculate the root-mean-square error (RMSE). The area under the receiver operating characteristic curve was used to analyze the detection of hypoxemia (ie, SaO<sub>2</sub><90%).</p> <p><strong>Results:</strong> SpO<sub>2</sub> estimates matching 215 SaO<sub>2</sub> samples in both study phases, from 33 participants, were analyzed. Tapping, rubbing, turning pages, and using a tablet degraded SpO<sub>2</sub> estimation (RMSE>4% for at least 1 device). All finger-worn pulse oximeters detected hypoxemia, with an overall sensitivity of ≥0.87 and specificity of ≥0.80, comparable to that of the Philips MX450 pulse oximeter.</p> <p><strong>Conclusions:</strong> The SpO<sub>2</sub> accuracy of wearable finger-worn pulse oximeters was within that required by the International Organization for Standardization guidelines. Performance was degraded by motion, but all pulse oximeters could detect hypoxemia. Our findings support the use of wearable, wireless transmission–mode pulse oximeters to detect the onset of clinical deterioration in hospital settings.</p>
spellingShingle Santos, M
Vollam, S
Pimentel, M
Morgado Areia, C
Young, L
Roman, C
Ede, J
Piper, P
King, E
Harford, M
Shah, A
Gustafson, O
Tarassenko, L
Watkinson, P
The use of wearable pulse oximeters in the prompt detection of hypoxemia and during movement: diagnostic accuracy study
title The use of wearable pulse oximeters in the prompt detection of hypoxemia and during movement: diagnostic accuracy study
title_full The use of wearable pulse oximeters in the prompt detection of hypoxemia and during movement: diagnostic accuracy study
title_fullStr The use of wearable pulse oximeters in the prompt detection of hypoxemia and during movement: diagnostic accuracy study
title_full_unstemmed The use of wearable pulse oximeters in the prompt detection of hypoxemia and during movement: diagnostic accuracy study
title_short The use of wearable pulse oximeters in the prompt detection of hypoxemia and during movement: diagnostic accuracy study
title_sort use of wearable pulse oximeters in the prompt detection of hypoxemia and during movement diagnostic accuracy study
work_keys_str_mv AT santosm theuseofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy
AT vollams theuseofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy
AT pimentelm theuseofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy
AT morgadoareiac theuseofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy
AT youngl theuseofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy
AT romanc theuseofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy
AT edej theuseofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy
AT piperp theuseofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy
AT kinge theuseofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy
AT harfordm theuseofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy
AT shaha theuseofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy
AT gustafsono theuseofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy
AT tarassenkol theuseofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy
AT watkinsonp theuseofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy
AT santosm useofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy
AT vollams useofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy
AT pimentelm useofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy
AT morgadoareiac useofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy
AT youngl useofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy
AT romanc useofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy
AT edej useofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy
AT piperp useofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy
AT kinge useofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy
AT harfordm useofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy
AT shaha useofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy
AT gustafsono useofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy
AT tarassenkol useofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy
AT watkinsonp useofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy