Evaluation of the Ohmeda 3700 pulse oximeter.

Arterial oxygen saturation values (Sao2) from 60% to 98% were measured by the Ohmeda 3700 pulse oximeter with the three types of probe available and compared with values of oxygen saturation estimated from direct arterial sampling (arterial oxygen and carbon dioxide tensions and pH) on 65 occasions....

Full description

Bibliographic Details
Main Authors: Warley, A, Mitchell, J, Stradling, JR
Format: Journal article
Language:English
Published: BMJ Publishing Group 1987
_version_ 1797105947676508160
author Warley, A
Mitchell, J
Stradling, JR
author_facet Warley, A
Mitchell, J
Stradling, JR
author_sort Warley, A
collection OXFORD
description Arterial oxygen saturation values (Sao2) from 60% to 98% were measured by the Ohmeda 3700 pulse oximeter with the three types of probe available and compared with values of oxygen saturation estimated from direct arterial sampling (arterial oxygen and carbon dioxide tensions and pH) on 65 occasions. The response time of the oximeter was measured after a sudden rise in inspired oxygen concentration. Artefact rejection was assessed by arterial compression proximal to the probe site, and by simultaneous recordings of overnight Sao2 on opposite hands. The ability to recreate patterns of oscillating Sao2 from the data stored in the oximeter was also investigated. With the best probe system the oximeter measured Sao2, relative to arterial values estimated from Pao2, with a mean (SD) difference of -0.4% (1.8%). The response time was comparable with those of previous oximeters. It was not possible to generate artefactual dips in excess of 2% Sao2, and the dual overnight recordings rarely showed even small dips on one tracing alone. The stored data can recreate oscillating Sao2 signals with wavelengths down to about 35 seconds, but not below. The Ohmeda 3700 pulse oximeter appears to be suitable for unattended overnight recordings of Sao2.
first_indexed 2024-03-07T06:54:40Z
format Journal article
id oxford-uuid:fdb8522a-cf0e-4ad8-adfb-a80ac58bb125
institution University of Oxford
language English
last_indexed 2024-03-07T06:54:40Z
publishDate 1987
publisher BMJ Publishing Group
record_format dspace
spelling oxford-uuid:fdb8522a-cf0e-4ad8-adfb-a80ac58bb1252022-03-27T13:30:59ZEvaluation of the Ohmeda 3700 pulse oximeter.Journal articlehttp://purl.org/coar/resource_type/c_dcae04bcuuid:fdb8522a-cf0e-4ad8-adfb-a80ac58bb125EnglishSymplectic Elements at OxfordBMJ Publishing Group1987Warley, AMitchell, JStradling, JRArterial oxygen saturation values (Sao2) from 60% to 98% were measured by the Ohmeda 3700 pulse oximeter with the three types of probe available and compared with values of oxygen saturation estimated from direct arterial sampling (arterial oxygen and carbon dioxide tensions and pH) on 65 occasions. The response time of the oximeter was measured after a sudden rise in inspired oxygen concentration. Artefact rejection was assessed by arterial compression proximal to the probe site, and by simultaneous recordings of overnight Sao2 on opposite hands. The ability to recreate patterns of oscillating Sao2 from the data stored in the oximeter was also investigated. With the best probe system the oximeter measured Sao2, relative to arterial values estimated from Pao2, with a mean (SD) difference of -0.4% (1.8%). The response time was comparable with those of previous oximeters. It was not possible to generate artefactual dips in excess of 2% Sao2, and the dual overnight recordings rarely showed even small dips on one tracing alone. The stored data can recreate oscillating Sao2 signals with wavelengths down to about 35 seconds, but not below. The Ohmeda 3700 pulse oximeter appears to be suitable for unattended overnight recordings of Sao2.
spellingShingle Warley, A
Mitchell, J
Stradling, JR
Evaluation of the Ohmeda 3700 pulse oximeter.
title Evaluation of the Ohmeda 3700 pulse oximeter.
title_full Evaluation of the Ohmeda 3700 pulse oximeter.
title_fullStr Evaluation of the Ohmeda 3700 pulse oximeter.
title_full_unstemmed Evaluation of the Ohmeda 3700 pulse oximeter.
title_short Evaluation of the Ohmeda 3700 pulse oximeter.
title_sort evaluation of the ohmeda 3700 pulse oximeter
work_keys_str_mv AT warleya evaluationoftheohmeda3700pulseoximeter
AT mitchellj evaluationoftheohmeda3700pulseoximeter
AT stradlingjr evaluationoftheohmeda3700pulseoximeter