Variation of beam quality correction factor with time: Cases for NE2571 and NE2581 chambers in the 6 and 10 MV x-rays Beams

The values of beam quality correction factor kQ that were experimentally determined from year 2002 to 2008 were analysed. As kQ is the function of ionization chamber and beam quality, the analysis were based on three cases namely (a) kQ(NE2571, 6 MV x-rays) that were determined from 17 measurement...

Full description

Bibliographic Details
Main Authors: Samat S.B, Too S.X, Kadni T, Dolah M.T
Format: Article
Language:English
Published: Universiti Kebangsaan Malaysia 2009
Online Access:http://journalarticle.ukm.my/68/1/
_version_ 1796914155030052864
author Samat S.B,
Too S.X,
Kadni T,
Dolah M.T,
author_facet Samat S.B,
Too S.X,
Kadni T,
Dolah M.T,
author_sort Samat S.B,
collection UKM
description The values of beam quality correction factor kQ that were experimentally determined from year 2002 to 2008 were analysed. As kQ is the function of ionization chamber and beam quality, the analysis were based on three cases namely (a) kQ(NE2571, 6 MV x-rays) that were determined from 17 measurements in the duration of 69 months at 6 radiotherapy centres, (b) kQ(NE2571, 10 MV x-rays) from 7 measurements in the duration of 12 months at one radiotherapy centre, and (c) kQ(NE2581, 6 MV x-rays) from 5 measurements in the duration of 5 months also at one radiotherapy centre. The purpose is to examine, in each case, the variation kQ for all the measurements. In other words, to see the variation of kQ with time. Results obtained are 0.993(NE2571, 6 MV), 0.986(NE2571, 10 MV) and 0.986(NE2581, 6 MV). This shows that in each case, despite the difference in the experimental data in getting kQ for all measurement, kQ remains constant with time. Reasons for this are explained.
first_indexed 2024-03-06T03:38:23Z
format Article
id ukm.eprints-68
institution Universiti Kebangsaan Malaysia
language English
last_indexed 2024-03-06T03:38:23Z
publishDate 2009
publisher Universiti Kebangsaan Malaysia
record_format dspace
spelling ukm.eprints-682016-12-14T06:26:19Z http://journalarticle.ukm.my/68/ Variation of beam quality correction factor with time: Cases for NE2571 and NE2581 chambers in the 6 and 10 MV x-rays Beams Samat S.B, Too S.X, Kadni T, Dolah M.T, The values of beam quality correction factor kQ that were experimentally determined from year 2002 to 2008 were analysed. As kQ is the function of ionization chamber and beam quality, the analysis were based on three cases namely (a) kQ(NE2571, 6 MV x-rays) that were determined from 17 measurements in the duration of 69 months at 6 radiotherapy centres, (b) kQ(NE2571, 10 MV x-rays) from 7 measurements in the duration of 12 months at one radiotherapy centre, and (c) kQ(NE2581, 6 MV x-rays) from 5 measurements in the duration of 5 months also at one radiotherapy centre. The purpose is to examine, in each case, the variation kQ for all the measurements. In other words, to see the variation of kQ with time. Results obtained are 0.993(NE2571, 6 MV), 0.986(NE2571, 10 MV) and 0.986(NE2581, 6 MV). This shows that in each case, despite the difference in the experimental data in getting kQ for all measurement, kQ remains constant with time. Reasons for this are explained. Universiti Kebangsaan Malaysia 2009-04 Article PeerReviewed application/pdf en http://journalarticle.ukm.my/68/1/ Samat S.B, and Too S.X, and Kadni T, and Dolah M.T, (2009) Variation of beam quality correction factor with time: Cases for NE2571 and NE2581 chambers in the 6 and 10 MV x-rays Beams. Sains Malaysiana, 38 (2). pp. 203-207. ISSN 0126-6039 http://www.ukm.my/~jsm/kandungan.html
spellingShingle Samat S.B,
Too S.X,
Kadni T,
Dolah M.T,
Variation of beam quality correction factor with time: Cases for NE2571 and NE2581 chambers in the 6 and 10 MV x-rays Beams
title Variation of beam quality correction factor with time: Cases for NE2571 and NE2581 chambers in the 6 and 10 MV x-rays Beams
title_full Variation of beam quality correction factor with time: Cases for NE2571 and NE2581 chambers in the 6 and 10 MV x-rays Beams
title_fullStr Variation of beam quality correction factor with time: Cases for NE2571 and NE2581 chambers in the 6 and 10 MV x-rays Beams
title_full_unstemmed Variation of beam quality correction factor with time: Cases for NE2571 and NE2581 chambers in the 6 and 10 MV x-rays Beams
title_short Variation of beam quality correction factor with time: Cases for NE2571 and NE2581 chambers in the 6 and 10 MV x-rays Beams
title_sort variation of beam quality correction factor with time cases for ne2571 and ne2581 chambers in the 6 and 10 mv x rays beams
url http://journalarticle.ukm.my/68/1/
work_keys_str_mv AT samatsb variationofbeamqualitycorrectionfactorwithtimecasesforne2571andne2581chambersinthe6and10mvxraysbeams
AT toosx variationofbeamqualitycorrectionfactorwithtimecasesforne2571andne2581chambersinthe6and10mvxraysbeams
AT kadnit variationofbeamqualitycorrectionfactorwithtimecasesforne2571andne2581chambersinthe6and10mvxraysbeams
AT dolahmt variationofbeamqualitycorrectionfactorwithtimecasesforne2571andne2581chambersinthe6and10mvxraysbeams