Influence of object location in cone beam computed tomography (NewTom 5G and 3D Accuitomo 170) on gray value measurements at an implant site

The aim of this study was to determine the gray value variation at an implant site with different object location within the selected field of view (FOV) in two cone beam computed tomography (CBCT) scanners. A 1-cm-thick section from the edentulous region of a dry human mandible was scanned by two C...

Full description

Bibliographic Details
Main Authors: Parsa, A., Ibrahim, N., Hassan, B., van der Stelt, P., Wismeijer, D.
Format: Article
Language:English
Published: Springer Verlag (Germany) 2014
Subjects:
Online Access:http://eprints.um.edu.my/10633/1/Influence_of_object_location_in_cone_beam_computed_tomography_%28NewTom_5G_and_3D_Accuitomo_170%29_on_gray_value_measurements_at_an_implant_site.pdf
_version_ 1825719891355762688
author Parsa, A.
Ibrahim, N.
Hassan, B.
van der Stelt, P.
Wismeijer, D.
author_facet Parsa, A.
Ibrahim, N.
Hassan, B.
van der Stelt, P.
Wismeijer, D.
author_sort Parsa, A.
collection UM
description The aim of this study was to determine the gray value variation at an implant site with different object location within the selected field of view (FOV) in two cone beam computed tomography (CBCT) scanners. A 1-cm-thick section from the edentulous region of a dry human mandible was scanned by two CBCT scanners: 3D Accuitomo 170 (J. Morita, Kyoto, Japan) and NewTom 5G (QR Verona, Verona, Italy). Five FOVs were used with each CBCT scanner. Within each FOV, the specimen was located at different positions. The scans were converted to DICOM format. Data analysis was performed using 3Diagnosys (ver. 3.1, 3DIEMME, Cantu, Italy) and Geomagic software (Studio 2012, Morrisville, NC). On one of the scans, a probe designating the site for pre-operative implant placement was selected. The inserted virtual implant was transformed on the same region on each CBCT scan by a three-dimensional registration algorithm. The mean voxel gray value of the region around the probe was derived separately from all CBCT scans. The influence of object location within each FOV on variability of voxel gray values was assessed. In both CBCT systems, object location had a significant influence on gray value measurements (F (4,16) = 3.71, p = 0.0255 for Accuitomo and F (4,16) = 9.31, p = 0.0000 for NewTom). Gray level values from CBCT images are influenced by object location within the FOV.
first_indexed 2024-03-06T05:26:39Z
format Article
id um.eprints-10633
institution Universiti Malaya
language English
last_indexed 2024-03-06T05:26:39Z
publishDate 2014
publisher Springer Verlag (Germany)
record_format dspace
spelling um.eprints-106332015-01-06T03:20:40Z http://eprints.um.edu.my/10633/ Influence of object location in cone beam computed tomography (NewTom 5G and 3D Accuitomo 170) on gray value measurements at an implant site Parsa, A. Ibrahim, N. Hassan, B. van der Stelt, P. Wismeijer, D. RK Dentistry The aim of this study was to determine the gray value variation at an implant site with different object location within the selected field of view (FOV) in two cone beam computed tomography (CBCT) scanners. A 1-cm-thick section from the edentulous region of a dry human mandible was scanned by two CBCT scanners: 3D Accuitomo 170 (J. Morita, Kyoto, Japan) and NewTom 5G (QR Verona, Verona, Italy). Five FOVs were used with each CBCT scanner. Within each FOV, the specimen was located at different positions. The scans were converted to DICOM format. Data analysis was performed using 3Diagnosys (ver. 3.1, 3DIEMME, Cantu, Italy) and Geomagic software (Studio 2012, Morrisville, NC). On one of the scans, a probe designating the site for pre-operative implant placement was selected. The inserted virtual implant was transformed on the same region on each CBCT scan by a three-dimensional registration algorithm. The mean voxel gray value of the region around the probe was derived separately from all CBCT scans. The influence of object location within each FOV on variability of voxel gray values was assessed. In both CBCT systems, object location had a significant influence on gray value measurements (F (4,16) = 3.71, p = 0.0255 for Accuitomo and F (4,16) = 9.31, p = 0.0000 for NewTom). Gray level values from CBCT images are influenced by object location within the FOV. Springer Verlag (Germany) 2014-05 Article PeerReviewed application/pdf en http://eprints.um.edu.my/10633/1/Influence_of_object_location_in_cone_beam_computed_tomography_%28NewTom_5G_and_3D_Accuitomo_170%29_on_gray_value_measurements_at_an_implant_site.pdf Parsa, A. and Ibrahim, N. and Hassan, B. and van der Stelt, P. and Wismeijer, D. (2014) Influence of object location in cone beam computed tomography (NewTom 5G and 3D Accuitomo 170) on gray value measurements at an implant site. Oral Radiology, 30 (2). pp. 153-159. ISSN 0911-6028, DOI https://doi.org/10.1007/s11282-013-0157-x <https://doi.org/10.1007/s11282-013-0157-x>. http://link.springer.com/article/10.1007%2Fs11282-013-0157-x http://dx.doi.org/10.1007/s11282-013-0157-x
spellingShingle RK Dentistry
Parsa, A.
Ibrahim, N.
Hassan, B.
van der Stelt, P.
Wismeijer, D.
Influence of object location in cone beam computed tomography (NewTom 5G and 3D Accuitomo 170) on gray value measurements at an implant site
title Influence of object location in cone beam computed tomography (NewTom 5G and 3D Accuitomo 170) on gray value measurements at an implant site
title_full Influence of object location in cone beam computed tomography (NewTom 5G and 3D Accuitomo 170) on gray value measurements at an implant site
title_fullStr Influence of object location in cone beam computed tomography (NewTom 5G and 3D Accuitomo 170) on gray value measurements at an implant site
title_full_unstemmed Influence of object location in cone beam computed tomography (NewTom 5G and 3D Accuitomo 170) on gray value measurements at an implant site
title_short Influence of object location in cone beam computed tomography (NewTom 5G and 3D Accuitomo 170) on gray value measurements at an implant site
title_sort influence of object location in cone beam computed tomography newtom 5g and 3d accuitomo 170 on gray value measurements at an implant site
topic RK Dentistry
url http://eprints.um.edu.my/10633/1/Influence_of_object_location_in_cone_beam_computed_tomography_%28NewTom_5G_and_3D_Accuitomo_170%29_on_gray_value_measurements_at_an_implant_site.pdf
work_keys_str_mv AT parsaa influenceofobjectlocationinconebeamcomputedtomographynewtom5gand3daccuitomo170ongrayvaluemeasurementsatanimplantsite
AT ibrahimn influenceofobjectlocationinconebeamcomputedtomographynewtom5gand3daccuitomo170ongrayvaluemeasurementsatanimplantsite
AT hassanb influenceofobjectlocationinconebeamcomputedtomographynewtom5gand3daccuitomo170ongrayvaluemeasurementsatanimplantsite
AT vandersteltp influenceofobjectlocationinconebeamcomputedtomographynewtom5gand3daccuitomo170ongrayvaluemeasurementsatanimplantsite
AT wismeijerd influenceofobjectlocationinconebeamcomputedtomographynewtom5gand3daccuitomo170ongrayvaluemeasurementsatanimplantsite