Comparison study of the prosthetics interface pressure profile of air splint socket and ICRC polypropylene socket for upper limb prosthetics

This study examined the interface pressure differences at the stump socket between an ICRC polypropylene socket and an air splint socket for a common wearer of transhumeral amputee using F-socket transducers. Two F-socket sensors arrays were attached to the residual limb. The subject was asked to co...

Full description

Bibliographic Details
Main Authors: Abd Razak, N.A., Abu Osman, Noor Azuan, Ali, S., Gholizadeh, H., Wan Abas, Wan Abu Bakar
Format: Article
Language:English
Published: 2015
Subjects:
Online Access:http://eprints.um.edu.my/15659/1/Comparison_study_of_the_prosthetics_interface_pressure_profile_of_air_splint_socket_and_ICRC.pdf
_version_ 1825720712363507712
author Abd Razak, N.A.
Abu Osman, Noor Azuan
Ali, S.
Gholizadeh, H.
Wan Abas, Wan Abu Bakar
author_facet Abd Razak, N.A.
Abu Osman, Noor Azuan
Ali, S.
Gholizadeh, H.
Wan Abas, Wan Abu Bakar
author_sort Abd Razak, N.A.
collection UM
description This study examined the interface pressure differences at the stump socket between an ICRC polypropylene socket and an air splint socket for a common wearer of transhumeral amputee using F-socket transducers. Two F-socket sensors arrays were attached to the residual limb. The subject was asked to complete the following tasks: Normal position, stand in a normal position without conducting any motion and shoulder movements, flexion/extension and abduction. The results revealed that the interface pressure applied using ICRC polypropylene socket was maximize at the end distal of the residual limb and give more pressure contact to any shoulder movements. Conversely, while using air splint socket, the socket was able to auto-adjust for required socket fitting even for any change while doing shoulder movements. Our result demonstrated how the comparison of pressure applied at the stump socket may lead in chosen the suitable prosthetic's socket for the amputee. The impending development of an auto adjusted socket that uses an air splint system will provide the prosthetic socket with a less contact pressure at the residual limb. (c) 2014 Nalecz Institute of Biocybemetics and Biomedical Engineering. Published by Elsevier Sp. z o.o. All rights reserved.
first_indexed 2024-03-06T05:39:18Z
format Article
id um.eprints-15659
institution Universiti Malaya
language English
last_indexed 2024-03-06T05:39:18Z
publishDate 2015
record_format dspace
spelling um.eprints-156592019-02-07T07:49:54Z http://eprints.um.edu.my/15659/ Comparison study of the prosthetics interface pressure profile of air splint socket and ICRC polypropylene socket for upper limb prosthetics Abd Razak, N.A. Abu Osman, Noor Azuan Ali, S. Gholizadeh, H. Wan Abas, Wan Abu Bakar T Technology (General) TA Engineering (General). Civil engineering (General) This study examined the interface pressure differences at the stump socket between an ICRC polypropylene socket and an air splint socket for a common wearer of transhumeral amputee using F-socket transducers. Two F-socket sensors arrays were attached to the residual limb. The subject was asked to complete the following tasks: Normal position, stand in a normal position without conducting any motion and shoulder movements, flexion/extension and abduction. The results revealed that the interface pressure applied using ICRC polypropylene socket was maximize at the end distal of the residual limb and give more pressure contact to any shoulder movements. Conversely, while using air splint socket, the socket was able to auto-adjust for required socket fitting even for any change while doing shoulder movements. Our result demonstrated how the comparison of pressure applied at the stump socket may lead in chosen the suitable prosthetic's socket for the amputee. The impending development of an auto adjusted socket that uses an air splint system will provide the prosthetic socket with a less contact pressure at the residual limb. (c) 2014 Nalecz Institute of Biocybemetics and Biomedical Engineering. Published by Elsevier Sp. z o.o. All rights reserved. 2015 Article PeerReviewed application/pdf en http://eprints.um.edu.my/15659/1/Comparison_study_of_the_prosthetics_interface_pressure_profile_of_air_splint_socket_and_ICRC.pdf Abd Razak, N.A. and Abu Osman, Noor Azuan and Ali, S. and Gholizadeh, H. and Wan Abas, Wan Abu Bakar (2015) Comparison study of the prosthetics interface pressure profile of air splint socket and ICRC polypropylene socket for upper limb prosthetics. Biocybernetics and Biomedical Engineering, 35 (2). pp. 100-105. ISSN 0208-5216, DOI https://doi.org/10.1016/j.bbe.2014.08.003 <https://doi.org/10.1016/j.bbe.2014.08.003>. http://www.sciencedirect.com/science/article/pii/S0208521614000680 10.1016/j.bbe.2014.08.003
spellingShingle T Technology (General)
TA Engineering (General). Civil engineering (General)
Abd Razak, N.A.
Abu Osman, Noor Azuan
Ali, S.
Gholizadeh, H.
Wan Abas, Wan Abu Bakar
Comparison study of the prosthetics interface pressure profile of air splint socket and ICRC polypropylene socket for upper limb prosthetics
title Comparison study of the prosthetics interface pressure profile of air splint socket and ICRC polypropylene socket for upper limb prosthetics
title_full Comparison study of the prosthetics interface pressure profile of air splint socket and ICRC polypropylene socket for upper limb prosthetics
title_fullStr Comparison study of the prosthetics interface pressure profile of air splint socket and ICRC polypropylene socket for upper limb prosthetics
title_full_unstemmed Comparison study of the prosthetics interface pressure profile of air splint socket and ICRC polypropylene socket for upper limb prosthetics
title_short Comparison study of the prosthetics interface pressure profile of air splint socket and ICRC polypropylene socket for upper limb prosthetics
title_sort comparison study of the prosthetics interface pressure profile of air splint socket and icrc polypropylene socket for upper limb prosthetics
topic T Technology (General)
TA Engineering (General). Civil engineering (General)
url http://eprints.um.edu.my/15659/1/Comparison_study_of_the_prosthetics_interface_pressure_profile_of_air_splint_socket_and_ICRC.pdf
work_keys_str_mv AT abdrazakna comparisonstudyoftheprostheticsinterfacepressureprofileofairsplintsocketandicrcpolypropylenesocketforupperlimbprosthetics
AT abuosmannoorazuan comparisonstudyoftheprostheticsinterfacepressureprofileofairsplintsocketandicrcpolypropylenesocketforupperlimbprosthetics
AT alis comparisonstudyoftheprostheticsinterfacepressureprofileofairsplintsocketandicrcpolypropylenesocketforupperlimbprosthetics
AT gholizadehh comparisonstudyoftheprostheticsinterfacepressureprofileofairsplintsocketandicrcpolypropylenesocketforupperlimbprosthetics
AT wanabaswanabubakar comparisonstudyoftheprostheticsinterfacepressureprofileofairsplintsocketandicrcpolypropylenesocketforupperlimbprosthetics