Designed mini protein 20 mimicking uricase encapsulated in ZIF-8 as nanozyme biosensor for uric acid detection

This work presents the use of encapsulated mini protein 20 mimicking uricase (mp20)-zeolitic imidazolate framework-8 (ZIF-8) as a bioreceptor for the development of a nanozyme-based electrochemical biosensor for uric acid detection. The electrochemical performance of the biofunctionalized mp20@ZIF-8...

Full description

Bibliographic Details
Main Authors: Abdul Aziz, Siti Fatimah Nur, Salleh, Abu Bakar, Ashari, Siti Efliza, Normi, Yahaya M., Yusof, Nor Azah, Ahmad, Shahrul Ainliah Alang
Format: Article
Published: MDPI 2022
_version_ 1796983980932726784
author Abdul Aziz, Siti Fatimah Nur
Salleh, Abu Bakar
Ashari, Siti Efliza
Normi, Yahaya M.
Yusof, Nor Azah
Ahmad, Shahrul Ainliah Alang
author_facet Abdul Aziz, Siti Fatimah Nur
Salleh, Abu Bakar
Ashari, Siti Efliza
Normi, Yahaya M.
Yusof, Nor Azah
Ahmad, Shahrul Ainliah Alang
author_sort Abdul Aziz, Siti Fatimah Nur
collection UPM
description This work presents the use of encapsulated mini protein 20 mimicking uricase (mp20)-zeolitic imidazolate framework-8 (ZIF-8) as a bioreceptor for the development of a nanozyme-based electrochemical biosensor for uric acid detection. The electrochemical performance of the biofunctionalized mp20@ZIF-8 on the reduced graphene oxide/screen-printed carbon electrode (rGO/SPCE) was investigated by optimizing operating parameters such as pH, deposition potential, and deposition time using a central composite design-response surface methodology (CCD-RSM). The quadratic regression model was developed to correlate the combination of each variable to the oxidation current density as a response. A significant effect on current response was observed under optimized conditions of pH of 7.4 at −0.35 V deposition potential and 56.56 s deposition time, with p < 0.05 for each interacted factor. The obtained coefficient of determination (R2) value of 0.9992 indicated good agreement with the experimental finding. The developed nanozyme biosensor (mp20@ZIF-8/rGO/SPCE) exhibited high selectivity in the presence of the same fold concentration of interfering species with a detection limit of 0.27 μM, over a concentration range of 1 to 34 μM. The practicality of the tailored biosensor in monitoring uric acid in human serum and urine samples was validated with high-performance liquid chromatography (HPLC) and a commercial uric acid meter. Hence, nanozyme-based is a promising platform that offers a rapid, sensitive, selective, and low-cost biosensor for the non-enzymatic detection of uric acid in biological samples.
first_indexed 2024-03-06T11:14:08Z
format Article
id upm.eprints-100902
institution Universiti Putra Malaysia
last_indexed 2024-03-06T11:14:08Z
publishDate 2022
publisher MDPI
record_format dspace
spelling upm.eprints-1009022023-07-14T04:03:06Z http://psasir.upm.edu.my/id/eprint/100902/ Designed mini protein 20 mimicking uricase encapsulated in ZIF-8 as nanozyme biosensor for uric acid detection Abdul Aziz, Siti Fatimah Nur Salleh, Abu Bakar Ashari, Siti Efliza Normi, Yahaya M. Yusof, Nor Azah Ahmad, Shahrul Ainliah Alang This work presents the use of encapsulated mini protein 20 mimicking uricase (mp20)-zeolitic imidazolate framework-8 (ZIF-8) as a bioreceptor for the development of a nanozyme-based electrochemical biosensor for uric acid detection. The electrochemical performance of the biofunctionalized mp20@ZIF-8 on the reduced graphene oxide/screen-printed carbon electrode (rGO/SPCE) was investigated by optimizing operating parameters such as pH, deposition potential, and deposition time using a central composite design-response surface methodology (CCD-RSM). The quadratic regression model was developed to correlate the combination of each variable to the oxidation current density as a response. A significant effect on current response was observed under optimized conditions of pH of 7.4 at −0.35 V deposition potential and 56.56 s deposition time, with p < 0.05 for each interacted factor. The obtained coefficient of determination (R2) value of 0.9992 indicated good agreement with the experimental finding. The developed nanozyme biosensor (mp20@ZIF-8/rGO/SPCE) exhibited high selectivity in the presence of the same fold concentration of interfering species with a detection limit of 0.27 μM, over a concentration range of 1 to 34 μM. The practicality of the tailored biosensor in monitoring uric acid in human serum and urine samples was validated with high-performance liquid chromatography (HPLC) and a commercial uric acid meter. Hence, nanozyme-based is a promising platform that offers a rapid, sensitive, selective, and low-cost biosensor for the non-enzymatic detection of uric acid in biological samples. MDPI 2022-07-04 Article PeerReviewed Abdul Aziz, Siti Fatimah Nur and Salleh, Abu Bakar and Ashari, Siti Efliza and Normi, Yahaya M. and Yusof, Nor Azah and Ahmad, Shahrul Ainliah Alang (2022) Designed mini protein 20 mimicking uricase encapsulated in ZIF-8 as nanozyme biosensor for uric acid detection. Nanomaterials, 12 (13). art. no. 2290. pp. 1-16. ISSN 2079-4991 https://www.mdpi.com/2079-4991/12/13/2290 10.3390/nano12132290
spellingShingle Abdul Aziz, Siti Fatimah Nur
Salleh, Abu Bakar
Ashari, Siti Efliza
Normi, Yahaya M.
Yusof, Nor Azah
Ahmad, Shahrul Ainliah Alang
Designed mini protein 20 mimicking uricase encapsulated in ZIF-8 as nanozyme biosensor for uric acid detection
title Designed mini protein 20 mimicking uricase encapsulated in ZIF-8 as nanozyme biosensor for uric acid detection
title_full Designed mini protein 20 mimicking uricase encapsulated in ZIF-8 as nanozyme biosensor for uric acid detection
title_fullStr Designed mini protein 20 mimicking uricase encapsulated in ZIF-8 as nanozyme biosensor for uric acid detection
title_full_unstemmed Designed mini protein 20 mimicking uricase encapsulated in ZIF-8 as nanozyme biosensor for uric acid detection
title_short Designed mini protein 20 mimicking uricase encapsulated in ZIF-8 as nanozyme biosensor for uric acid detection
title_sort designed mini protein 20 mimicking uricase encapsulated in zif 8 as nanozyme biosensor for uric acid detection
work_keys_str_mv AT abdulazizsitifatimahnur designedminiprotein20mimickinguricaseencapsulatedinzif8asnanozymebiosensorforuricaciddetection
AT sallehabubakar designedminiprotein20mimickinguricaseencapsulatedinzif8asnanozymebiosensorforuricaciddetection
AT asharisitiefliza designedminiprotein20mimickinguricaseencapsulatedinzif8asnanozymebiosensorforuricaciddetection
AT normiyahayam designedminiprotein20mimickinguricaseencapsulatedinzif8asnanozymebiosensorforuricaciddetection
AT yusofnorazah designedminiprotein20mimickinguricaseencapsulatedinzif8asnanozymebiosensorforuricaciddetection
AT ahmadshahrulainliahalang designedminiprotein20mimickinguricaseencapsulatedinzif8asnanozymebiosensorforuricaciddetection