Computed and conventional chest radiography: A comparison of image quality and radiation dose

The aim of this study was to compare the image quality and entrance skin dose (ESD) for film-screen and computed chest radiography. Analysis of the image quality and dose on chest radiography was carried out on a conventional X-ray unit using film-screen, storage phosphor plates and selenium drum di...

Disgrifiad llawn

Manylion Llyfryddiaeth
Prif Awduron: Ramli, Khadijah, Abdullah, Basri Johan Jeet, Ng, Kwan Hoong, Mahmud, Rozi, Hussain, Amir Fuad
Fformat: Erthygl
Iaith:English
English
Cyhoeddwyd: Royal Australian and New Zealand College of Radiologists 2005
Pynciau:
Mynediad Ar-lein:http://psasir.upm.edu.my/id/eprint/18327/1/Computed%20and%20conventional%20chest%20radiography.pdf
_version_ 1825946172319072256
author Ramli, Khadijah
Abdullah, Basri Johan Jeet
Ng, Kwan Hoong
Mahmud, Rozi
Hussain, Amir Fuad
author_facet Ramli, Khadijah
Abdullah, Basri Johan Jeet
Ng, Kwan Hoong
Mahmud, Rozi
Hussain, Amir Fuad
author_sort Ramli, Khadijah
collection UPM
description The aim of this study was to compare the image quality and entrance skin dose (ESD) for film-screen and computed chest radiography. Analysis of the image quality and dose on chest radiography was carried out on a conventional X-ray unit using film-screen, storage phosphor plates and selenium drum direct chest radiography. For each receptor, ESD was measured in 60 patients using thermoluminescent dosemeters. Images were printed on 35 × 43 cm films. Image quality was assessed subjectively by evaluation of anatomic features and estimation of the image quality, following the guidelines established by the protocols of the Commission of the European Communities. There was no statistically significant difference noted between the computed and conventional images (Wilcoxon rank sum test, P > 0.05). Imaging of the mediastinum and peripheral lung structures were better visualized with the storage phosphor and selenium drum technique than with the film-screen combination. The patients' mean ESD for chest radiography using the storage phosphor, film-screen combination and selenium drum was 0.20, 0.20 and 0.25 mGy, respectively, with no statistically significant difference with P > 0.05 (χ2 tests). © 2005 Royal Australian and New Zealand College of Radiologists
first_indexed 2024-03-06T07:42:54Z
format Article
id upm.eprints-18327
institution Universiti Putra Malaysia
language English
English
last_indexed 2024-03-06T07:42:54Z
publishDate 2005
publisher Royal Australian and New Zealand College of Radiologists
record_format dspace
spelling upm.eprints-183272015-10-29T07:08:32Z http://psasir.upm.edu.my/id/eprint/18327/ Computed and conventional chest radiography: A comparison of image quality and radiation dose Ramli, Khadijah Abdullah, Basri Johan Jeet Ng, Kwan Hoong Mahmud, Rozi Hussain, Amir Fuad The aim of this study was to compare the image quality and entrance skin dose (ESD) for film-screen and computed chest radiography. Analysis of the image quality and dose on chest radiography was carried out on a conventional X-ray unit using film-screen, storage phosphor plates and selenium drum direct chest radiography. For each receptor, ESD was measured in 60 patients using thermoluminescent dosemeters. Images were printed on 35 × 43 cm films. Image quality was assessed subjectively by evaluation of anatomic features and estimation of the image quality, following the guidelines established by the protocols of the Commission of the European Communities. There was no statistically significant difference noted between the computed and conventional images (Wilcoxon rank sum test, P > 0.05). Imaging of the mediastinum and peripheral lung structures were better visualized with the storage phosphor and selenium drum technique than with the film-screen combination. The patients' mean ESD for chest radiography using the storage phosphor, film-screen combination and selenium drum was 0.20, 0.20 and 0.25 mGy, respectively, with no statistically significant difference with P > 0.05 (χ2 tests). © 2005 Royal Australian and New Zealand College of Radiologists Royal Australian and New Zealand College of Radiologists 2005-12 Article PeerReviewed application/pdf en http://psasir.upm.edu.my/id/eprint/18327/1/Computed%20and%20conventional%20chest%20radiography.pdf Ramli, Khadijah and Abdullah, Basri Johan Jeet and Ng, Kwan Hoong and Mahmud, Rozi and Hussain, Amir Fuad (2005) Computed and conventional chest radiography: A comparison of image quality and radiation dose. Australasian Radiology, 49 (6). pp. 460-466. ISSN 0004-8461 http://dx.doi.org/10.1111/j.1440-1673.2005.01497.x Chest - Radiography Radiography - Image quality Radiation dosimetry doi:10.1111/j.1440-1673.2005.01497.x English
spellingShingle Chest - Radiography
Radiography - Image quality
Radiation dosimetry
Ramli, Khadijah
Abdullah, Basri Johan Jeet
Ng, Kwan Hoong
Mahmud, Rozi
Hussain, Amir Fuad
Computed and conventional chest radiography: A comparison of image quality and radiation dose
title Computed and conventional chest radiography: A comparison of image quality and radiation dose
title_full Computed and conventional chest radiography: A comparison of image quality and radiation dose
title_fullStr Computed and conventional chest radiography: A comparison of image quality and radiation dose
title_full_unstemmed Computed and conventional chest radiography: A comparison of image quality and radiation dose
title_short Computed and conventional chest radiography: A comparison of image quality and radiation dose
title_sort computed and conventional chest radiography a comparison of image quality and radiation dose
topic Chest - Radiography
Radiography - Image quality
Radiation dosimetry
url http://psasir.upm.edu.my/id/eprint/18327/1/Computed%20and%20conventional%20chest%20radiography.pdf
work_keys_str_mv AT ramlikhadijah computedandconventionalchestradiographyacomparisonofimagequalityandradiationdose
AT abdullahbasrijohanjeet computedandconventionalchestradiographyacomparisonofimagequalityandradiationdose
AT ngkwanhoong computedandconventionalchestradiographyacomparisonofimagequalityandradiationdose
AT mahmudrozi computedandconventionalchestradiographyacomparisonofimagequalityandradiationdose
AT hussainamirfuad computedandconventionalchestradiographyacomparisonofimagequalityandradiationdose