Determination of the effective SSD as the function of various energy and field size for siemens digital mevatron linear accelerator model type MXE 2 in HUSM

(SSDetr) as the function of various energies and field sizes for the SIEMENS DIGITAL MEV ATRON (model MXE 2) linear acceleralor in HUSM. The experiment involv~ electrons energies which deputized the lowest energy, the middle energy and the highest functional energy provided by this machine .The e...

Full description

Bibliographic Details
Main Author: Zakaria, Noor Ainzza
Format: Article
Language:English
Published: Pusat Pengajian Sains Kesihatan, Universiti Sains Malaysia 2004
Subjects:
Online Access:http://eprints.usm.my/45965/1/LI...Noor%20Ainzza%20Binti%20Zakaria...2004...24%20pages.pdf
_version_ 1825835252008878080
author Zakaria, Noor Ainzza
author_facet Zakaria, Noor Ainzza
author_sort Zakaria, Noor Ainzza
collection USM
description (SSDetr) as the function of various energies and field sizes for the SIEMENS DIGITAL MEV ATRON (model MXE 2) linear acceleralor in HUSM. The experiment involv~ electrons energies which deputized the lowest energy, the middle energy and the highest functional energy provided by this machine .The energies were 5 MeV, 7 MeV, and 9 MeV electron beams. The measurement for the energies level were done together with the application of the available applicators which carried the field size Scm x Scm, 1 Ocm x 1 Ocm, 1 Scm x I Scm, 20cm x 20cm ,and 25cmx25cm. Several materials and devices were used for the experiment which included the parallel plate chamber, solid water phantom and an electrometer. The method of measurement to determine the effective SSD was such that the parallel plate ionization chamber was located at the depth of dose maximum for the specific energies and fields sizes applied and after that a series of output measurements were made at that depth. The measurements started with zero air gap between an applicator surface and the phantom surface for a selected energy .Every single shoot, the gap between applicator surface and phantom surface was increased about 1 em until reach 15 em. The same procedures were repeated for all defined energies and applicators. A few calculations involved in the process of determining effective SSD. For all of the energies used, the deviations of the measurement reading from the inverse square law estimation were found larger up to more than SO% for the smaller field size applied. An effective SSD, SSDetr had been determined at dmax for each energy as a function of field size. For 5 MeV electron energy with 5x5 field size, 10 x10 field size,15x 15 field size,20 x 20 and 25 x 25, the values are 37.07 em, 79.65 cm,95.15cm,l00cm and 105.26.For 7 MeV energy with the same field size as mentioned above, the values are 46.47cm, 91.18cm, 96.83cm,107.3lcm, and 105.18cm.While for 9 MeV with all the field size as mentioned before, the values are 57.78cm,96.26cm,104.60cm,104.56cm and 104.77cm. The effective SSD values determined for the energies and fields sizes are fixed for a specific condition
first_indexed 2024-03-06T15:34:56Z
format Article
id usm.eprints-45965
institution Universiti Sains Malaysia
language English
last_indexed 2024-03-06T15:34:56Z
publishDate 2004
publisher Pusat Pengajian Sains Kesihatan, Universiti Sains Malaysia
record_format dspace
spelling usm.eprints-459652020-01-22T04:08:16Z http://eprints.usm.my/45965/ Determination of the effective SSD as the function of various energy and field size for siemens digital mevatron linear accelerator model type MXE 2 in HUSM Zakaria, Noor Ainzza R Medicine (SSDetr) as the function of various energies and field sizes for the SIEMENS DIGITAL MEV ATRON (model MXE 2) linear acceleralor in HUSM. The experiment involv~ electrons energies which deputized the lowest energy, the middle energy and the highest functional energy provided by this machine .The energies were 5 MeV, 7 MeV, and 9 MeV electron beams. The measurement for the energies level were done together with the application of the available applicators which carried the field size Scm x Scm, 1 Ocm x 1 Ocm, 1 Scm x I Scm, 20cm x 20cm ,and 25cmx25cm. Several materials and devices were used for the experiment which included the parallel plate chamber, solid water phantom and an electrometer. The method of measurement to determine the effective SSD was such that the parallel plate ionization chamber was located at the depth of dose maximum for the specific energies and fields sizes applied and after that a series of output measurements were made at that depth. The measurements started with zero air gap between an applicator surface and the phantom surface for a selected energy .Every single shoot, the gap between applicator surface and phantom surface was increased about 1 em until reach 15 em. The same procedures were repeated for all defined energies and applicators. A few calculations involved in the process of determining effective SSD. For all of the energies used, the deviations of the measurement reading from the inverse square law estimation were found larger up to more than SO% for the smaller field size applied. An effective SSD, SSDetr had been determined at dmax for each energy as a function of field size. For 5 MeV electron energy with 5x5 field size, 10 x10 field size,15x 15 field size,20 x 20 and 25 x 25, the values are 37.07 em, 79.65 cm,95.15cm,l00cm and 105.26.For 7 MeV energy with the same field size as mentioned above, the values are 46.47cm, 91.18cm, 96.83cm,107.3lcm, and 105.18cm.While for 9 MeV with all the field size as mentioned before, the values are 57.78cm,96.26cm,104.60cm,104.56cm and 104.77cm. The effective SSD values determined for the energies and fields sizes are fixed for a specific condition Pusat Pengajian Sains Kesihatan, Universiti Sains Malaysia 2004 Article NonPeerReviewed application/pdf en http://eprints.usm.my/45965/1/LI...Noor%20Ainzza%20Binti%20Zakaria...2004...24%20pages.pdf Zakaria, Noor Ainzza (2004) Determination of the effective SSD as the function of various energy and field size for siemens digital mevatron linear accelerator model type MXE 2 in HUSM. Determination of the effective SSD as the function of various energy and field size for siemens digital mevatron linear accelerator model type MXE 2 in HUSM. (Submitted)
spellingShingle R Medicine
Zakaria, Noor Ainzza
Determination of the effective SSD as the function of various energy and field size for siemens digital mevatron linear accelerator model type MXE 2 in HUSM
title Determination of the effective SSD as the function of various energy and field size for siemens digital mevatron linear accelerator model type MXE 2 in HUSM
title_full Determination of the effective SSD as the function of various energy and field size for siemens digital mevatron linear accelerator model type MXE 2 in HUSM
title_fullStr Determination of the effective SSD as the function of various energy and field size for siemens digital mevatron linear accelerator model type MXE 2 in HUSM
title_full_unstemmed Determination of the effective SSD as the function of various energy and field size for siemens digital mevatron linear accelerator model type MXE 2 in HUSM
title_short Determination of the effective SSD as the function of various energy and field size for siemens digital mevatron linear accelerator model type MXE 2 in HUSM
title_sort determination of the effective ssd as the function of various energy and field size for siemens digital mevatron linear accelerator model type mxe 2 in husm
topic R Medicine
url http://eprints.usm.my/45965/1/LI...Noor%20Ainzza%20Binti%20Zakaria...2004...24%20pages.pdf
work_keys_str_mv AT zakarianoorainzza determinationoftheeffectivessdasthefunctionofvariousenergyandfieldsizeforsiemensdigitalmevatronlinearacceleratormodeltypemxe2inhusm