Study on the factors affecting glandular dose in digital breast tomosynthesis

Digital breast tomosynthesis (DBT) is the type of 3 -dimensional mammography that construct three-dimensional images based on the series of 2-dimensional images. DBT is said to has lower radiation dose than the digital mammography. This study focuses on the factors that affecting the average glandul...

Celý popis

Podrobná bibliografie
Hlavní autor: Ab Aziz, Nur Alia Nabihah
Médium: Diplomová práce
Jazyk:English
Vydáno: 2020
Témata:
On-line přístup:http://eprints.usm.my/50697/1/Nur%20Alia%20Nabihah-24%20pages.pdf
_version_ 1825906342222626816
author Ab Aziz, Nur Alia Nabihah
author_facet Ab Aziz, Nur Alia Nabihah
author_sort Ab Aziz, Nur Alia Nabihah
collection USM
description Digital breast tomosynthesis (DBT) is the type of 3 -dimensional mammography that construct three-dimensional images based on the series of 2-dimensional images. DBT is said to has lower radiation dose than the digital mammography. This study focuses on the factors that affecting the average glandular dose (AGD) in DBT. A total of 224 dose data were included in this study, which is 57 samples for RCC, 55 samples for LCC, 57 samples for LMLO, and 55 samples for LMLO projections. All examinations were performed using the same mammographic unit which is Siemens MAMMOMAT Revelation (Siemens Healthineer, German) at Imaging Unit, Advanced Medical and Dental Institute (AMDI), USM, Penang. For the comparison between 2D and 3D mode in DBT, the different is statistically significant where 3D shown a higher mean of AGD than 2D modes. The multiple regression and correlation tests were used to evaluate the correlation between AGD with exposure setting and breast thickness. There is a positive correlation between the AGD with exposure settings and breast thickness. The mAs and kV settings have a linear relationship with AGD with R2 value of 0.562 and 0.605 respectively for right and left side of the breast. The correlation value of R2 obtained for breast thickness is 0.207 and 0.059 respectively for right and left breast. The denser breast (breast glandularity type D and C) has higher AGD value than less dense breast (type A and B). There is no statistically significant difference of AGD for different mammographic projections.
first_indexed 2024-03-06T15:48:14Z
format Thesis
id usm.eprints-50697
institution Universiti Sains Malaysia
language English
last_indexed 2024-03-06T15:48:14Z
publishDate 2020
record_format dspace
spelling usm.eprints-506972021-11-21T04:43:52Z http://eprints.usm.my/50697/ Study on the factors affecting glandular dose in digital breast tomosynthesis Ab Aziz, Nur Alia Nabihah R Medicine Digital breast tomosynthesis (DBT) is the type of 3 -dimensional mammography that construct three-dimensional images based on the series of 2-dimensional images. DBT is said to has lower radiation dose than the digital mammography. This study focuses on the factors that affecting the average glandular dose (AGD) in DBT. A total of 224 dose data were included in this study, which is 57 samples for RCC, 55 samples for LCC, 57 samples for LMLO, and 55 samples for LMLO projections. All examinations were performed using the same mammographic unit which is Siemens MAMMOMAT Revelation (Siemens Healthineer, German) at Imaging Unit, Advanced Medical and Dental Institute (AMDI), USM, Penang. For the comparison between 2D and 3D mode in DBT, the different is statistically significant where 3D shown a higher mean of AGD than 2D modes. The multiple regression and correlation tests were used to evaluate the correlation between AGD with exposure setting and breast thickness. There is a positive correlation between the AGD with exposure settings and breast thickness. The mAs and kV settings have a linear relationship with AGD with R2 value of 0.562 and 0.605 respectively for right and left side of the breast. The correlation value of R2 obtained for breast thickness is 0.207 and 0.059 respectively for right and left breast. The denser breast (breast glandularity type D and C) has higher AGD value than less dense breast (type A and B). There is no statistically significant difference of AGD for different mammographic projections. 2020-08 Thesis NonPeerReviewed application/pdf en http://eprints.usm.my/50697/1/Nur%20Alia%20Nabihah-24%20pages.pdf Ab Aziz, Nur Alia Nabihah (2020) Study on the factors affecting glandular dose in digital breast tomosynthesis. Masters thesis, Universiti Sains Malaysia.
spellingShingle R Medicine
Ab Aziz, Nur Alia Nabihah
Study on the factors affecting glandular dose in digital breast tomosynthesis
title Study on the factors affecting glandular dose in digital breast tomosynthesis
title_full Study on the factors affecting glandular dose in digital breast tomosynthesis
title_fullStr Study on the factors affecting glandular dose in digital breast tomosynthesis
title_full_unstemmed Study on the factors affecting glandular dose in digital breast tomosynthesis
title_short Study on the factors affecting glandular dose in digital breast tomosynthesis
title_sort study on the factors affecting glandular dose in digital breast tomosynthesis
topic R Medicine
url http://eprints.usm.my/50697/1/Nur%20Alia%20Nabihah-24%20pages.pdf
work_keys_str_mv AT abaziznuralianabihah studyonthefactorsaffectingglandulardoseindigitalbreasttomosynthesis