A comprehensive 3D dosimetric analysis of the vienna ring applicator in high-dose-rate (HDR) brachytherapy for cervical cancer treatment: TLD measurement

Introduction: Intracavitary brachytherapy is the most common approach used for gynecological cancers which happens to be the cancer with highest occurrence among women. Intracavitary-interstitial (IC/IS) brachytherapy is an approach that is developed, and various applicators are utilised in this tre...

Full description

Bibliographic Details
Main Author: Sivaprakasam, Pugalya
Format: Monograph
Language:English
Published: Universiti Sains Malaysia 2024
Subjects:
Online Access:http://eprints.usm.my/61367/1/Pugalyaa%20AP%20Sivaprakasam-E.pdf
_version_ 1824451891123191808
author Sivaprakasam, Pugalya
author_facet Sivaprakasam, Pugalya
author_sort Sivaprakasam, Pugalya
collection USM
description Introduction: Intracavitary brachytherapy is the most common approach used for gynecological cancers which happens to be the cancer with highest occurrence among women. Intracavitary-interstitial (IC/IS) brachytherapy is an approach that is developed, and various applicators are utilised in this treatment method such as the Vienna Ring applicator. However, a comprehensive understanding on the dosimetric behavior of Vienna Ring applicator is required for efficient use in brachytherapy for cervical cancer. Purpose: This study aims to experimentally analyse the dosimetric performance of Vienna Ring applicator in aspects of target coverage, organ at risk (OAR) sparing and ability to deliver the planned dose during treatment. Methods and Materials: A Vienna Ring applicator was assembled in a pelvic phantom with Thermoluminescence Dosimeters (TLD-100) chips were attached to the OARs (bladder and rectum) to act as markers. Computed Tomography (CT) images were obtained for treatment planning. 7Gy was prescribed for irradiation and the treatment was planned for the assessment of dose received by target, OAR and the TLDs at 8 points. The dose measured by the TLD-100 chips during the irradiation and dose calculated by treatment planning system (TPS) on the TLD points were compared. Results: The dose received by target, bladder and rectum are 7.1 Gy, 2.4 Gy and 7.4 Gy respectively as obtained from TPS. Out of 8 TLDs, 7 TLDs have dose deviation less than 20% while 1 TLD (F7) have dose deviation of 31.03% between measured and calculated dose. Conclusion: The dose distribution of Vienna ring applicator to GTV is not homogenous but has high degree of conformity. The Vienna ring applicator was able to spare the bladder and deliver optimum dose to tumour. The Vienna ring applicator can deliver planned dose during treatment delivery due to 90% of the dose deviation was below 20%. Therefore, the applicator can be categorised as a reliable applicator in brachytherapy for cervical cancer.
first_indexed 2025-02-19T02:41:50Z
format Monograph
id usm.eprints-61367
institution Universiti Sains Malaysia
language English
last_indexed 2025-02-19T02:41:50Z
publishDate 2024
publisher Universiti Sains Malaysia
record_format dspace
spelling usm.eprints-613672025-01-06T07:48:50Z http://eprints.usm.my/61367/ A comprehensive 3D dosimetric analysis of the vienna ring applicator in high-dose-rate (HDR) brachytherapy for cervical cancer treatment: TLD measurement Sivaprakasam, Pugalya R Medicine RC71-78.7 Examination. Diagnosis Including raiography Introduction: Intracavitary brachytherapy is the most common approach used for gynecological cancers which happens to be the cancer with highest occurrence among women. Intracavitary-interstitial (IC/IS) brachytherapy is an approach that is developed, and various applicators are utilised in this treatment method such as the Vienna Ring applicator. However, a comprehensive understanding on the dosimetric behavior of Vienna Ring applicator is required for efficient use in brachytherapy for cervical cancer. Purpose: This study aims to experimentally analyse the dosimetric performance of Vienna Ring applicator in aspects of target coverage, organ at risk (OAR) sparing and ability to deliver the planned dose during treatment. Methods and Materials: A Vienna Ring applicator was assembled in a pelvic phantom with Thermoluminescence Dosimeters (TLD-100) chips were attached to the OARs (bladder and rectum) to act as markers. Computed Tomography (CT) images were obtained for treatment planning. 7Gy was prescribed for irradiation and the treatment was planned for the assessment of dose received by target, OAR and the TLDs at 8 points. The dose measured by the TLD-100 chips during the irradiation and dose calculated by treatment planning system (TPS) on the TLD points were compared. Results: The dose received by target, bladder and rectum are 7.1 Gy, 2.4 Gy and 7.4 Gy respectively as obtained from TPS. Out of 8 TLDs, 7 TLDs have dose deviation less than 20% while 1 TLD (F7) have dose deviation of 31.03% between measured and calculated dose. Conclusion: The dose distribution of Vienna ring applicator to GTV is not homogenous but has high degree of conformity. The Vienna ring applicator was able to spare the bladder and deliver optimum dose to tumour. The Vienna ring applicator can deliver planned dose during treatment delivery due to 90% of the dose deviation was below 20%. Therefore, the applicator can be categorised as a reliable applicator in brachytherapy for cervical cancer. Universiti Sains Malaysia 2024 Monograph NonPeerReviewed application/pdf en http://eprints.usm.my/61367/1/Pugalyaa%20AP%20Sivaprakasam-E.pdf Sivaprakasam, Pugalya (2024) A comprehensive 3D dosimetric analysis of the vienna ring applicator in high-dose-rate (HDR) brachytherapy for cervical cancer treatment: TLD measurement. Project Report. Universiti Sains Malaysia. (Submitted)
spellingShingle R Medicine
RC71-78.7 Examination. Diagnosis Including raiography
Sivaprakasam, Pugalya
A comprehensive 3D dosimetric analysis of the vienna ring applicator in high-dose-rate (HDR) brachytherapy for cervical cancer treatment: TLD measurement
title A comprehensive 3D dosimetric analysis of the vienna ring applicator in high-dose-rate (HDR) brachytherapy for cervical cancer treatment: TLD measurement
title_full A comprehensive 3D dosimetric analysis of the vienna ring applicator in high-dose-rate (HDR) brachytherapy for cervical cancer treatment: TLD measurement
title_fullStr A comprehensive 3D dosimetric analysis of the vienna ring applicator in high-dose-rate (HDR) brachytherapy for cervical cancer treatment: TLD measurement
title_full_unstemmed A comprehensive 3D dosimetric analysis of the vienna ring applicator in high-dose-rate (HDR) brachytherapy for cervical cancer treatment: TLD measurement
title_short A comprehensive 3D dosimetric analysis of the vienna ring applicator in high-dose-rate (HDR) brachytherapy for cervical cancer treatment: TLD measurement
title_sort comprehensive 3d dosimetric analysis of the vienna ring applicator in high dose rate hdr brachytherapy for cervical cancer treatment tld measurement
topic R Medicine
RC71-78.7 Examination. Diagnosis Including raiography
url http://eprints.usm.my/61367/1/Pugalyaa%20AP%20Sivaprakasam-E.pdf
work_keys_str_mv AT sivaprakasampugalya acomprehensive3ddosimetricanalysisoftheviennaringapplicatorinhighdoseratehdrbrachytherapyforcervicalcancertreatmenttldmeasurement
AT sivaprakasampugalya comprehensive3ddosimetricanalysisoftheviennaringapplicatorinhighdoseratehdrbrachytherapyforcervicalcancertreatmenttldmeasurement