Objective and subjective image quality measures of medical screen-film systems

Image quality of Lanex Regular/T Mat G and Lanex Regular/T Mat L screen-film systems was measured objectively by means of noise equivalent quanta (NEQ) and subjectively by means of receiver operating characteristic (ROC) curve analysis to study the link between the two measures. For the objective me...

وصف كامل

التفاصيل البيبلوغرافية
المؤلفون الرئيسيون: Wan Hassan, Wan Muhamad Saridan, Munajat, Yusof, Sahibuddin, Shamsul
التنسيق: مقال
اللغة:English
منشور في: Malaysian Institute of Physics 2004
الموضوعات:
الوصول للمادة أونلاين:http://eprints.utm.my/2520/1/Jurnal_Fizik_Malaysia_25_2004_15-18.pdf
الوصف
الملخص:Image quality of Lanex Regular/T Mat G and Lanex Regular/T Mat L screen-film systems was measured objectively by means of noise equivalent quanta (NEQ) and subjectively by means of receiver operating characteristic (ROC) curve analysis to study the link between the two measures. For the objective measure, uniformly exposed radiographs and radiographs of bar pattern test object were used to determine the NEQ. For the subjective measure, detection of simulated lesions in posteroanterior anthropomorphic chest radiographs by an observer was analysed by the ROC methodology. Both measures show that a low frequency signal shows better on Lanex Regular/T Mat L, whilst a high frequency signal shows better on Lanex Regular/T Mat G.