Interface micromotion failure simulation of cementless hip prosthesis

The most commonly reported failure modes of cementless hip stems are loosening and thigh pain; both are attributed to the relative motion at the bone-implant interface due to failure to achieve sufficient primary fixation. Accurate predictions of hip stems’ stability are therefore crucial to the p...

Full description

Bibliographic Details
Main Authors: Abdul Kadir, Mohammed Rafiq, Kamsah, Nazri, Tamin, Mohd. Nasir, Lee, Faisal Abidin
Format: Monograph
Language:English
Published: Faculty of Mechanical Engineering 2007
Subjects:
Online Access:http://eprints.utm.my/5854/1/78057.pdf
_version_ 1825909802746773504
author Abdul Kadir, Mohammed Rafiq
Kamsah, Nazri
Tamin, Mohd. Nasir
Lee, Faisal Abidin
author_facet Abdul Kadir, Mohammed Rafiq
Kamsah, Nazri
Tamin, Mohd. Nasir
Lee, Faisal Abidin
author_sort Abdul Kadir, Mohammed Rafiq
collection ePrints
description The most commonly reported failure modes of cementless hip stems are loosening and thigh pain; both are attributed to the relative motion at the bone-implant interface due to failure to achieve sufficient primary fixation. Accurate predictions of hip stems’ stability are therefore crucial to the pre-clinical analyses of hip arthroplasty. This study uses finite element technique to analyse the effect of muscle forces on the predicted micromotion and therefore stability of cementless femoral components. An in-house experimentally validated micromotion algorithm was used in analyses simulating two of the most common physiological activities – walking and stair-climbing. The results showed that models where muscle loads were included had ten times larger micromotion than those modelled without muscle loads. Ignoring muscle forces in any pre-clinical evaluation of femoral stems are therefore not advisable as it will overestimate the stability of the stem.
first_indexed 2024-03-05T18:07:35Z
format Monograph
id utm.eprints-5854
institution Universiti Teknologi Malaysia - ePrints
language English
last_indexed 2024-03-05T18:07:35Z
publishDate 2007
publisher Faculty of Mechanical Engineering
record_format dspace
spelling utm.eprints-58542017-08-10T04:53:59Z http://eprints.utm.my/5854/ Interface micromotion failure simulation of cementless hip prosthesis Abdul Kadir, Mohammed Rafiq Kamsah, Nazri Tamin, Mohd. Nasir Lee, Faisal Abidin T Technology (General) The most commonly reported failure modes of cementless hip stems are loosening and thigh pain; both are attributed to the relative motion at the bone-implant interface due to failure to achieve sufficient primary fixation. Accurate predictions of hip stems’ stability are therefore crucial to the pre-clinical analyses of hip arthroplasty. This study uses finite element technique to analyse the effect of muscle forces on the predicted micromotion and therefore stability of cementless femoral components. An in-house experimentally validated micromotion algorithm was used in analyses simulating two of the most common physiological activities – walking and stair-climbing. The results showed that models where muscle loads were included had ten times larger micromotion than those modelled without muscle loads. Ignoring muscle forces in any pre-clinical evaluation of femoral stems are therefore not advisable as it will overestimate the stability of the stem. Faculty of Mechanical Engineering 2007 Monograph NonPeerReviewed application/pdf en http://eprints.utm.my/5854/1/78057.pdf Abdul Kadir, Mohammed Rafiq and Kamsah, Nazri and Tamin, Mohd. Nasir and Lee, Faisal Abidin (2007) Interface micromotion failure simulation of cementless hip prosthesis. Project Report. Faculty of Mechanical Engineering, Skudai, Johor. (Unpublished) https://link.springer.com/chapter/10.1007/978-3-540-69139-6_109
spellingShingle T Technology (General)
Abdul Kadir, Mohammed Rafiq
Kamsah, Nazri
Tamin, Mohd. Nasir
Lee, Faisal Abidin
Interface micromotion failure simulation of cementless hip prosthesis
title Interface micromotion failure simulation of cementless hip prosthesis
title_full Interface micromotion failure simulation of cementless hip prosthesis
title_fullStr Interface micromotion failure simulation of cementless hip prosthesis
title_full_unstemmed Interface micromotion failure simulation of cementless hip prosthesis
title_short Interface micromotion failure simulation of cementless hip prosthesis
title_sort interface micromotion failure simulation of cementless hip prosthesis
topic T Technology (General)
url http://eprints.utm.my/5854/1/78057.pdf
work_keys_str_mv AT abdulkadirmohammedrafiq interfacemicromotionfailuresimulationofcementlesshipprosthesis
AT kamsahnazri interfacemicromotionfailuresimulationofcementlesshipprosthesis
AT taminmohdnasir interfacemicromotionfailuresimulationofcementlesshipprosthesis
AT leefaisalabidin interfacemicromotionfailuresimulationofcementlesshipprosthesis