Complementary metal oxide semiconductor electrocardiogram amplifier for low power wearable cardiac screening

Cardiovascular disease is the number one killer disease in Malaysia. Although sudden cardiac arrest is the main cause of death, the Malaysian awareness of towards cardiovascular disease is still low. The trend of health care screening devices in the world is increasingly towards the favor of portabi...

Full description

Bibliographic Details
Main Author: Ow, Tze Weng
Format: Thesis
Language:English
Published: 2016
Subjects:
Online Access:http://eprints.utm.my/78391/1/OwTzeWengMFKE2016.pdf
_version_ 1796862842539868160
author Ow, Tze Weng
author_facet Ow, Tze Weng
author_sort Ow, Tze Weng
collection ePrints
description Cardiovascular disease is the number one killer disease in Malaysia. Although sudden cardiac arrest is the main cause of death, the Malaysian awareness of towards cardiovascular disease is still low. The trend of health care screening devices in the world is increasingly towards the favor of portability and wearability, especially in the most common electrocardiogram (ECG) monitoring system. This is because these wearable screening devices are not restricting the patient’s freedom and daily activities. While the demand of low power and low cost biomedical system on chip (SoC) is increasing in exponential way, the front end ECG amplifiers are still suffering from flicker noise for low frequency cardiac signal acquisition, 50 Hz power line electromagnetic interference, and the large unstable input offsets due to the electrodeskin interface is not attached properly. In this project, a high performance ECG amplifier that suitable for low power wearable cardiac screening is proposed. The amplifier adopts the highly stable folded cascode topology and later being implemented into RC feedback circuit for low frequency DC offset cancellation. By using 0.13 µm CMOS technology from Silterra, the simulation results show that this front end circuit can achieve a very low input referred noise of 1 pV=pHz and high common mode rejection ratio (CMRR) of 174.05 dB. It also gives voltage gain of 75.45 dB with good power supply rejection ratio (PSSR) of 92.12 dB. The total power consumption is only 3 µW and thus suitable to be implemented with further signal processing and classification back end for low power biomedical SoC.
first_indexed 2024-03-05T20:17:43Z
format Thesis
id utm.eprints-78391
institution Universiti Teknologi Malaysia - ePrints
language English
last_indexed 2024-03-05T20:17:43Z
publishDate 2016
record_format dspace
spelling utm.eprints-783912018-08-26T04:56:24Z http://eprints.utm.my/78391/ Complementary metal oxide semiconductor electrocardiogram amplifier for low power wearable cardiac screening Ow, Tze Weng TK Electrical engineering. Electronics Nuclear engineering Cardiovascular disease is the number one killer disease in Malaysia. Although sudden cardiac arrest is the main cause of death, the Malaysian awareness of towards cardiovascular disease is still low. The trend of health care screening devices in the world is increasingly towards the favor of portability and wearability, especially in the most common electrocardiogram (ECG) monitoring system. This is because these wearable screening devices are not restricting the patient’s freedom and daily activities. While the demand of low power and low cost biomedical system on chip (SoC) is increasing in exponential way, the front end ECG amplifiers are still suffering from flicker noise for low frequency cardiac signal acquisition, 50 Hz power line electromagnetic interference, and the large unstable input offsets due to the electrodeskin interface is not attached properly. In this project, a high performance ECG amplifier that suitable for low power wearable cardiac screening is proposed. The amplifier adopts the highly stable folded cascode topology and later being implemented into RC feedback circuit for low frequency DC offset cancellation. By using 0.13 µm CMOS technology from Silterra, the simulation results show that this front end circuit can achieve a very low input referred noise of 1 pV=pHz and high common mode rejection ratio (CMRR) of 174.05 dB. It also gives voltage gain of 75.45 dB with good power supply rejection ratio (PSSR) of 92.12 dB. The total power consumption is only 3 µW and thus suitable to be implemented with further signal processing and classification back end for low power biomedical SoC. 2016-06 Thesis NonPeerReviewed application/pdf en http://eprints.utm.my/78391/1/OwTzeWengMFKE2016.pdf Ow, Tze Weng (2016) Complementary metal oxide semiconductor electrocardiogram amplifier for low power wearable cardiac screening. Masters thesis, Universiti Teknologi Malaysia, Faculty of Electrical Engineering. http://dms.library.utm.my:8080/vital/access/manager/Repository/vital:94014
spellingShingle TK Electrical engineering. Electronics Nuclear engineering
Ow, Tze Weng
Complementary metal oxide semiconductor electrocardiogram amplifier for low power wearable cardiac screening
title Complementary metal oxide semiconductor electrocardiogram amplifier for low power wearable cardiac screening
title_full Complementary metal oxide semiconductor electrocardiogram amplifier for low power wearable cardiac screening
title_fullStr Complementary metal oxide semiconductor electrocardiogram amplifier for low power wearable cardiac screening
title_full_unstemmed Complementary metal oxide semiconductor electrocardiogram amplifier for low power wearable cardiac screening
title_short Complementary metal oxide semiconductor electrocardiogram amplifier for low power wearable cardiac screening
title_sort complementary metal oxide semiconductor electrocardiogram amplifier for low power wearable cardiac screening
topic TK Electrical engineering. Electronics Nuclear engineering
url http://eprints.utm.my/78391/1/OwTzeWengMFKE2016.pdf
work_keys_str_mv AT owtzeweng complementarymetaloxidesemiconductorelectrocardiogramamplifierforlowpowerwearablecardiacscreening