Development and validation of a stability-indicating RP–HPLC method for estimation of atazanavir sulfate in bulk
A stability-indicating reverse phase–high performance liquid chromatography (RP–HPLC) method was developed and validated for the determination of atazanavir sulfate in tablet dosage forms using C18 column Phenomenix (250 mm×4.6 mm, 5 μm) with a mobile phase consisting of 900 mL of HPLC grade methano...
Main Authors: | , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Elsevier
2017-04-01
|
Series: | Journal of Pharmaceutical Analysis |
Subjects: | |