Utilização de um monitor de vídeo como fonte de alta tensão para eletroforese capilar

This article describes the use of a conventional CRT monitor as a high voltage power supply for capillary electrophoresis. With this monitor, a 23-kV high voltage with a ripple of 1.32% was observed. The reproducibility of the applied high voltage was evaluated by measuring the standard deviations o...

Full description

Bibliographic Details
Main Authors: José Alberto Fracassi da Silva, Leonardo Frasatto
Format: Article
Language:English
Published: Sociedade Brasileira de Química 2006-12-01
Series:Química Nova
Subjects:
Online Access:http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422006000600038&lng=en&tlng=en