The effect of the magnetic fields from three different configurations of the MRIgRT systems on the dose deposition from lateral opposing photon beams in a laryngeal geometry – A Monte Carlo study

Objective: To investigate the impact of different magnetic field strengths and orientations on the dose deposition from lateral opposing photon beams irradiating a laryngeal geometry, with particular attention focused on the dose homogeneity of the area around the air cavity inside the larynx. Metho...

Full description

Bibliographic Details
Main Authors: Vivien W.S. Chu, Monica W.K. Kan, Louis K.Y. Lee, Kenneth C.W. Wong, Macy Tong, Anthony T.C. Chan
Format: Article
Language:English
Published: Elsevier 2021-09-01
Series:Radiation Medicine and Protection
Subjects:
Online Access:http://www.sciencedirect.com/science/article/pii/S2666555721000447